Eline Huizing1, Richte C.L. Schuurmann2, Frank F.J. Simonis3, Çağdaş Ünlü1, Henri G.D. Leuvenink4, and Jean-Paul P.M. de Vries2
1Department of Surgery, Northwest Clinics, Alkmaar, Netherlands, 2Department of Surgery, University Medical Center Groningen, Groningen, Netherlands, 3University of Twente, Enschede, Netherlands, 4Surgical Research Laboratory, University Medical Center Groningen, Groningen, Netherlands
Synopsis
Optimal
treatment parameters of balloon angioplasty for revascularization such as inflation
time and pressure are subject to debate. Limited data exists because accurate
assessment of vascular function is challenging and therefore the influence procedure
parameters have on arterial wall damage is largely unknown. This study shows that
portable MRI at 0.125x0.125x2 mm³ resolution can be used for assessment of diameter
and wall thickness during balloon angioplasty in ex vivo porcine arteries. Even
after deflation and removal of the balloon the arterial diameter remained
enlarged; this was coupled to a decrease in wall thickness.
Introduction
Revascularization
remains the cornerstone for limb salvage in patients with severe lower limb
ischemia1. In recent years, the endovascular approach has
become the first choice for revascularization2. Traditionally, endovascular procedures were
performed using a plain balloon, but many specific devices are being developed
to facilitate the procedure and aim for better outcomes3. Also, optimal treatment options such as
inflation time and inflation pressure are subject of debate. There is, however, limited data that
identifies the optimal endovascular strategy and technique4. Further, the influence that different procedure
parameters have on arterial wall damage is still unknown, and accurate assessment
of vascular function is challenging. This study describes the feasibility of a
portable low-field MRI scanner for accurate real-time assessment of artery
function (diameter change) during balloon angioplasty in ex-vivo arteries with
realistic blood flow.Method
A
setup was built with realistic flow (70 ml/min) of porcine blood through a
porcine cadaver artery flowing at physiological temperatures through ex vivo
bench-top model. A 0.4T portable MRI scanner with 1 cm diameter bore (Pure
Devices, Rimpar, Germany) was used for imaging. 2D MR scans were performed in
transversal direction and longitudinal direction using a spoiled gradient echo
sequence, with parameters as described in Table 1.Results
Scans were
acquired before balloon insertion, after balloon inflation, after balloon
deflation and finally after balloon removal; the resulting images are displayed
in Figure 1.
The
acquired resolution of 0.125x0.125 mm² of the transversal scans is sufficient
to accurately image the artery with an arterial wall thickness of 1.2 mm before
inflation. The lumen diameter of the artery was 3.2 mm, which increased to 5.2
mm after balloon inflation. Because the balloon was filled with water, it can
be clearly seen in the scans. After deflation of the balloon, the arterial lumen
diameter remained increased compared to initial measurements (4.2 mm) and the
wall thickness was decreased to 0.8 mm.Discussion
The
resolution of the transversal and longitudinal low-field tabletop MRI images
were sufficient to detect increase of the artery lumen diameter and decrease of
the artery wall thickness. The boundaries between the balloon, lumen and vessel
wall were clearly distinguishable when the vessel was placed inside the scanner.
When the
artery is surrounded by Dulbecco's Modified Eagle Medium (DMEM) instead of air,
arterial function will be preserved. Subsequently, vessel injury can be
assessed after balloon angioplasty using macroscopic and microscopic
histopathology. However, since signal intensity of DMEM approaches that of the
arterial wall, wall diameter assessment becomes more challenging. A solution
would be to suppress the signal from DMEM by adding a high concentration of T2-shortening
agent. The effect of these additives on the functioning of the artery has yet
to be investigated.
Currently,
2D analyses were performed with high spatial resolution and relatively low scan
times, but 3D sequences are also optional, making isotropic resolution reachable.
Future experiments will investigate the possibility of measuring luminal flow rate
using the tabletop MRI.Conclusion
It
is feasible to image the arterial wall of ex vivo porcine arteries using a
portable low-field MRI scanner during several steps in a catheter inflation
procedure. This enables measuring diameter and wall thickness change, together
with vascular function this is essential for defining the optimal balloon
angioplasty strategy.Acknowledgements
No acknowledgement found.References
1. Conte
MS, Bradbury AW, Kolh P, et al. Global vascular guidelines on the
management of chronic limb-threatening ischemia. J Vasc Surg. 2019;69(6S):3S-125S.e40.
2. Aboyans
V, Ricco J-B, Bartelink M-LEL, et al. Editor’s Choice - 2017 ESC Guidelines
on the Diagnosis and Treatment of Peripheral
Arterial Diseases, in collaboration with the European Society for
Vascular Surgery (ESVS). Eur J Vasc Endovasc Surg Off J
Eur Soc Vasc Surg. 2018;55(3):305-368.
3. Zeller T, Rastan A,
Macharzina R, et al. Novel Approaches to the Management of Advanced Peripheral
Artery Disease: Perspectives on
Drug-Coated Balloons, Drug-Eluting Stents, and Bioresorbable Scaffolds. Curr
Cardiol Rep. 2015;17(9):624.
4. Farber A, Eberhardt RT.
The Current State of Critical Limb Ischemia: A Systematic Review. JAMA Surg.
2016;151(11):1070-1077.