KyungPyo Hong1, Amanda L DiCarlo1, Aggelos K Katsaggelos1,2,3, Florian A Schiffers3, Cynthia K Rigsby1,4, Hassan Haji-Valizadeh5, and Daniel Kim1,6
1Radiology, Northwestern University Feinberg School of Medicine, Chicago, IL, United States, 2Electrical and Computer Engineering, Northwestern University, Evanston, IL, United States, 3Computer Science, Northwestern University, Evanston, IL, United States, 4Medical Imaging, Ann & Robert H. Lurie Children's Hospital of Chicago, Chicago, IL, United States, 5Internal Medicine, Cardiovascular Division, Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, MA, United States, 6Biomedical Engineering, McCormick School of Engineering, Northwestern University, Evanston, IL, United States
Synopsis
While conventional density compensation
function(DCF) performs sufficiently well for filtered backprojection(FBP) and radial
k-space MRI when the Nyquist sampling condition is met and/or evenly-spaced view angles are used, it may perform poorly when sub-sampling
and/or irrational-view angles are used. We propose an optimized DCF for the
aforementioned conditions by calculating the density weights based on geometric
properties of radial k-space sampling in a discrete environment, regardless of
scan conditions such as data sizes and view angles. Compared with standard DCF, the optimized DCF
produces higher signal-to-noise ratio(SNR) in FBP (phantom) and more accurate
flow metrics in 48-fold accelerated, phase-contrast MRI.
Introduction
Density compensation
function (DCF) is an essential component to account for non-uniform data
sampling in filtered backprojection (FBP)1 and radial k-space MRI. The original
DCF was analytically derived as a Jacobian determinant in an analog environment;
it is an “invariant v-shaped” high-pass filter with zero at the center, which
results in reduced signal-to-noise ratio(SNR) and increased streaking artifacts.2 A standard DCF performs sufficiently in
a discrete environment when the Nyquist sampling condition is met and samples
are evenly distributed. The standard DCF, however, may perform sub-optimally in highly-accelerated
radial k-space MRI or when irrational-view angles are used. In this study, we
sought to develop an optimized DCF based on geometric properties of radial k-space
sampling in a discrete environment, regardless of scan conditions (e.g., sub-sampling
factor, view angles), and evaluate its performance with FBP in phantom and compressed
sensing(CS) reconstruction of 48-fold accelerated real-time phase-contrast(PC)
MRI in patients with congenital heart disease(CHD). Methods
Optimized DCF: While each radial spoke samples uniformly-spaced
discrete data points along the radial direction, the spacing along the
circumferential direction between neighboring discrete points varies with the
radius in the polar coordinate system (Figure-1a). We propose a new concept that the spacing along both the circumferential and radial directions
is uniform such that $$$arc\;length\:(\triangle{L})=r\triangle{\theta}=\triangle{r}=1$$$ (Figure-1b), which is also a minimal spacing to avoid overlaps between samples at a given radius. Using
this concept as a criterion, we propose to calculate the degree
of overlap between two neighboring samples at a given radius. Specifically, a
DCF at $$$r$$$ of $$$i^{th}$$$ radial spoke (i.e., $$$DCF_{i}^{r}$$$) can be calculated to be inversely proportional to the sum
of density weights ($$$c_{ij}^{r}$$$) between all neighboring points including itself, as
described in Figure-1c.
Phantom Experiment for
FBP: We scanned a T1MES
phantom3 on a 1.5T MRI system (Aera,Siemens), using a gradient
echo pulse sequence with radial k-space sampling with the following imaging
parameters: field-of-view (FOV)=288x288 mm2, reconstruction matrix=192x192, slice thickness=8 mm, TE/TR=1.5/3.3 msec, receiver-bandwidth=1002 Hz/pixel,
flip angle=15°, 600
radial spokes (i.e., full Nyquist condition), readout duration=2 sec, and dummy
scan (2 sec). We separately acquired two data sets using regular-view angle=0.3° and tiny-golden-view angle=32.034°.4
In-vivo Experiment for CS: 48-fold accelerated, real-time PC data
of 17 pediatric patients with CHD (10 males; mean age=11.3±3.2 years) were
obtained from a previous study5 approved by our institutional review board. MRI scans
included standard clinical 2D PC MRI with Cartesian k-space sampling (i.e.,
clinical PC) and real-time 2D PC MRI with radial k-space sampling (i.e., radial
PC) at up to four locations (aortic valve, pulmonic valve, left-pulmonary
artery, right-pulmonary artery; N=60 vascular planes in total). For the
relevant imaging parameters of clinical PC, please see reference.5 The relevant imaging parameters for 48-fold accelerated, radial
PC were: FOV=288x288 mm2, reconstruction matrix=192x192, in-plane resolution=1.5x1.5 mm2, slice
thickness=6 mm, receiver-bandwidth=745 Hz/pixel, flip angle=12°,
TE/TR=1.8/4.17 msec, prospective ECG-triggering, free-breathing scan time=3.34
sec (835-msec of dummy scan+2.50 sec) leading to 300 continuous radial spokes
per plane, and golden-view angles=111.2461°.4 Velocity-encoding strength was matched
to the clinical PC (150-400 cm/sec). For additional details, please see reference.5
Image Reconstruction: All image reconstructions were separately
performed using GPU-based non-uniform fast Fourier transform (NUFFT)6
with standard and optimized DCF on a GPU workstation (Tesla V100, NVIDIA)
equipped with MATLAB (R2020b,MathWorks). Auto-calibrated gradient delay
correction was performed on all k-space data using the RING method.7
For the phantom data reconstruction, FBP was performed using multi-coil NUFFT. For
the patient data reconstruction, we binned 300 radial spokes to achieve 4 radial
spokes per frame (33.4-msec temporal resolution) and employed a radial CS framework8
using two orthogonal sparsity transforms of temporal total variation (TVt)
and temporal principal component analysis (PCAt) with identical
normalized regularization weight (α)=0.0075 (relative to the maximum signal),
as shown in Figure 2.
Image Analysis: For phantom data, we measured SNR. For in-vivo data, we quantified the forward-flow volume, backward-flow volume,
and peak-velocity per location by manually drawing regions-of-interest in
Matlab. For radial PC, we analyzed the first full-heartbeat in all subjects.
Statistical
Analysis: We conducted linear-regression and
Bland-Altman analyses on forward-flow volume, backward-flow volume, and peak-velocity
to assess the degree of association and agreement, respectively, between
clinical and radial PC datasets. p<0.05
was considered statistically significant. Results
Figure
3a compares the mean DCF of 600 radial spokes between standard vs. optimized DCF.
As shown in Figure 3b, the resulting FBP images of the phantom using optimized
DCF produced 11 to 18% higher SNR than standard DCF for both regular and
irrational-view angles. Figure 4 shows representative magnitude (zero-filled NUFFT
and CS reconstruction) and phase-difference images in a pulmonary-valve plane, and
the corresponding plots of mean velocity. Radial PC with optimized DCF produced
higher dynamic range of mean velocity (i.e., less temporal blurring) than that
with standard DCF. As
shown in Figure 5, compared
with clinical PC as reference, the levels of association and agreement in
forward-flow volume, backward-flow volume, and peak-velocity were better for optimized
DCF than standard DCF.Discussions
This
study demonstrates that optimized DCF produces higher SNR in FBP(phantom) and
more accurate peak-velocity, forward-flow volume, and backward-flow volume in
48-fold accelerated radial PC MRI. Future
work includes extending the optimized DCF for 3D radial scans.Acknowledgements
This work was supported in part by the following
grants: National Institutes
of Health (R01HL116895, R01HL138578, R21EB024315, R21AG055954, R01HL151079) and
American Heart Association (19IPLOI34760317). References
1. Shepp
LA, Logan BF. The Fourier reconstruction of a head section. IEEE Transactions
on Nuclear Science 1974;21(3):21-43.
2. Crawford
CR. CT filtration aliasing artifacts. IEEE Trans Med Imaging 1991;10(1):99-102.
3. Captur
G, Gatehouse P, Keenan KE, Heslinga FG, Bruehl R, Prothmann M, Graves MJ, Eames
RJ, Torlasco C, Benedetti G, Donovan J, Ittermann B, Boubertakh R, Bathgate A,
Royet C, Pang W, Nezafat R, Salerno M, Kellman P, Moon JC. A medical device-grade
T1 and ECV phantom for global T1 mapping quality assurance-the T1 Mapping and
ECV Standardization in cardiovascular magnetic resonance (T1MES) program. J
Cardiovasc Magn Reson 2016;18(1):58.
4. Winkelmann
S, Schaeffter T, Koehler T, Eggers H, Doessel O. An optimal radial profile
order based on the Golden Ratio for time-resolved MRI. IEEE Trans Med Imaging
2007;26(1):68-76.
5. Haji-Valizadeh
H, Feng L, Ma LE, Shen D, Block KT, Robinson JD, Markl M, Rigsby CK, Kim D.
Highly accelerated, real-time phase-contrast MRI using radial k-space sampling
and GROG-GRASP reconstruction: a feasibility study in pediatric patients with
congenital heart disease. NMR Biomed 2020;33(5):e4240.
6. Knoll
F, Schwarzl A, Diwoky C, Sodickson DK. gpuNUFFT-an open source GPU library for
3D regridding with direct Matlab interface. In: Proceedings of the 22rd Annual
Meeting of ISMRM, Melbourne, Australia 2014. Abstract No. 4297.
7. Rosenzweig
S, Holme HCM, Uecker M. Simple auto-calibrated gradient delay estimation from
few spokes using Radial Intersections (RING). Magn Reson Med
2019;81(3):1898-1906.
8. Feng L, Grimm R, Block KT, Chandarana
H, Kim S, Xu J, Axel L, Sodickson DK, Otazo R. Golden-angle radial sparse
parallel MRI: combination of compressed sensing, parallel imaging, and golden-angle
radial sampling for fast and flexible dynamic volumetric MRI. Magn Reson Med 2014;72(3):707-717.