Jack Allen1,2, George Mathew1, Miriam Conway1, Sophie Jenkins1, David Firmin1,2, Jennifer Keegan1, Sonya V. Babu-Narayan1, and Peter Gatehouse1
1Cardiovascular Biomedical Research Unit, Royal Brompton and Harefield NHS Trust, London, United Kingdom, 2National Heart and Lung Institute, Imperial College London, London, United Kingdom
Synopsis
3D Late Gadolinium-Enhanced (LGE) imaging is used to assess
scarring in patients with Atrial Fibrillation (AF). Acquiring image data during
every cardiac cycle allows a reasonable total scan duration but exacerbates ghosting artefacts caused by variable heart rates, such as those of patients with
AF. Dynamic Inversion Time (TI) methods improve image quality by modifying the
TI for each cardiac cycle. We present the initial stage of a patient study to validate a recently-proposed blood-focused dynamic TI algorithm. No
improvement was found in comparison to the original algorithm. Future comparisons will include more patients with high R-wave interval variability.
Introduction
Inversion-recovery 3D LGE imaging is a useful tool to map atrial
scar distribution in patients with Atrial Fibrillation (AF). However, arrhythmia
during the 3D scans in this patient population causes ghosting and poor
myocardial nulling. Alternate-cycle inversion-recovery imaging [1, 2] reduces the sensitivity to heart rate variability but requires
a longer scan duration unless an adaptation is made, which could potentially
introduce other compromises.
Dynamic TI algorithms [3, 4] improve
3D LGE image quality when acquiring in every cardiac cycle, by updating TI on a
beat-by-beat basis. The original algorithm [3] improved myocardial
nulling by adjusting TI in “real time” in the current cycle of the
sequence according to the previous cardiac cycle duration. While myocardial nulling is
important, phase-encode ghosting
from variations in the brighter blood signal remains troublesome in 3D imaging.
Reductions in blood ghosting, by targeting consistent blood
signal across the TIs while maintaining low myocardial signal, were shown by simulations
and phantom studies for a blood-focused dynamic TI algorithm [4].
Figure 1 compares the method of [4] with [3] in
terms of blood Mz.
Here we extend the work of [4] to a preliminary
patient study.Methods
We compared the performance of the blood-focused algorithm
[4] with the original algorithm [3]. The blood-focused
algorithm uses the full trigger interval and sequence timing history (including
partial-saturation by acquisition) to model the blood and myocardium
longitudinal magnetisation (Mz) evolution in real time during the sequence
acquisition. It uses blood T1, myocardium T1 and an estimation of the mean RR interval
to calculate the required TI in the current cycle to consistently achieve an
unvarying target blood Mz, which is the blood Mz at the myocardium null time
for the given mean interval.
3D LGE inversion-recovery scans were performed after 0.1mmol/kg
gadolinium-based-contrast-agent (gadobutrol) administration in 13 patients with
congenital heart disease, under local ethics approval. Each patient was scanned
with two 3D LGE acquisitions, first the original dynamic TI algorithm [3]
followed by the blood-focused algorithm [4] with otherwise
identical imaging parameters, using single R-wave gating and the CLAWS phase-encode
strategy in free respiration [5] on an Avanto 1.5T scanner (Siemens
Healthineers, Erlangen Germany): segmented 3D spoiled gradient-echo, 32-36
slices, 1.5x1.5x4mm; reconstructed to 64-72 slices, 0.7x0.7x2mm. TI scout scans
were performed before each 3D LGE acquisition, to estimate myocardium T1 for
each algorithm. Blood T1 was measured prior to the blood-focused algorithm,
using a single-slice short-axis MOLLI acquisition. All target blood Mz values
were calculated as if the mean RR-interval was 1000ms.
To
assess changes in blood ghosting artefact, signal was measured from regions positioned
to detect phase-encode ghosting from blood in the heart (i.e. in-line with the
heart in the right-left primary phase-encoding direction, Fig. 2) using the
central 11 slices of the 3D imaging. As reference, mean intraventricular blood
signal was measured for the same slices. To account for blood signal changes
between the two 3D LGE scans, ghosting artefact mean was measured relative to
blood signal mean.Results
Figure 2 shows the centre slices for patient 5 as an example of image quality and signal intensity.
Measured ghost and blood signal means, with ghost-to-blood
ratios, are shown in Fig. 3. No significant changes between the two algorithms (two
tailed Student t-test of 13 pairs) were detected in mean ghosting (p>0.4) or
ghost-to-blood ratio (p>0.5). The intraventricular mean blood signal was 72+/-31
for the original algorithm and 70+/-21 for blood-focused algorithm (p>0.7).
Figure 4 shows the ghost-to-blood ratio versus the R-wave
interval variability for the 13 pairs of 3D LGE scans, where large R-R interval standard deviation represents high heart rate
variability. The highest R-R interval variability as a percentage of the corresponding
mean R-R interval was observed for patients 3 and 8, at 23%. There was no
significant change in mean RR interval between the two 3D LGE scans (p>0.7). Discussion
No significant difference was observed between the two
algorithms. For each subject, variations in blood signal between the 3D LGE
scans (arising from the target blood Mz selection mixed with gadolinium
“wash-out”) may have influenced the blood ghosting-to-signal ratio. Clinical
imaging required the original algorithm to be run first although
randomised order was preferable.
Future studies are needed to increase the cohort size as differences in respiratory motion and associated residual PE ghosting between
the two 3D scans were likely a major confounder in the ghosting assessment,
even though ROIs were placed to avoid respiratory ghosting of the anterior
chest wall. Further, more patients with AF are necessary.
Randomised scan acquisition order could reduce bias from
blood T1 changes. The blood-based algorithm depends on an accurate Mz model,
which depends on RF flip angle accuracy, making it important for RF flip angle
mapping to be performed to assess the B1+ variation within the heart. Myocardial ghosting possibly confounded the blood ghosting measurement, which could be reduced using a patient-specific mean interval for each target
blood Mz calculation.Conclusions
No improvement was found in this initial stage of an in
vivo validation of the blood-based dynamic TI algorithm [4]
for improved image quality in 3D LGE imaging. Assessment in a larger group is
required, including more arrhythmia patients.Acknowledgements
This work was supported by the British Heart Foundation (PG/17/81/33345).References
1. Kino, A., et al. (2009), American Journal of Roentgenology. 2009;193: W381-W388.
2. Kido, T., et al. (2014), European Journal of Radiology, volume 83 Issue 12 2014. Pages 2159-21663
3. Keegan, J., et al. (2015), Magn. Reson. Med., 73: 646-654.
4. Allen, J., et al. (2020), ISMRM. #2059
5. Jhooti, P., et al. (2010), Magn. Reson. Med., 64: 1015-1026.