Patricia Pais-Roldan1, Seong Dae Yun1, and Jon N Shah1
1Forschungszentrum Juelich, Juelich, Germany
Synopsis
Due
to their rapid acquisition and high contrast-to-noise-ratio, MRI
sequences based on gradient-echo (GE) schemes, e.g., gradient-echo
echo-planar-imaging (GE-EPI), are the most commonly used method in
human functional MRI. However, their enhanced sensitivity to veins
restricts their use in laminar fMRI due to poor signal localization,
i.e., venous bias. Here, we investigated the spatial specificity of
high-resolution GE signals, pre-processed with ten different
approaches. Removal of motion parameters, physiological signals and
non-GM tissue contributions, as well as regression of the
pre-processed phase image from the magnitude image, significantly
increased the spatial specificity of GE-fMRI in resting-state and
task paradigms.
Introduction
With
the advent of sub-millimetre resolution functional magnetic resonance
imaging (fMRI), the function of the cerebral cortex is increasingly
being assessed at different depths. Although several studies have
investigated human laminar responses to particular tasks (e.g.,
[1-3]);
whole-brain resting-state fMRI (rs-fMRI) remains challenging due to
the limited field of view afforded by most high-resolution sequences.
Gradient-echo (GE) fMRI schemes offer the possibility of achieving
very high BOLD signal sensitivity, which facilitates investigations
of the cortical thickness in the whole brain. However, GE sequences
are sensitive to magnetic field inhomogeneities and introduce bias
towards the superficial layers of the cortex due to the influence of
ascending and pial veins. Task-fMRI obtained with GE sequences can be
partially corrected by applying existing methods [4-9];
in contrast, no specific method has been optimized to analyse laminar
rs-fMRI. Here, we applied ten different pre-processing pipelines to
whole-brain, sub-millimetre GE fMRI (0.63×0.63×0.63mm) to study
their effect on the spatial localization of GE-signals.Methods
One
rs-fMRI scan (~10 min) and one task-fMRI scan (~8 min) were obtained
from 13 volunteers using a Siemens Magnetom Terra 7T scanner with a
1-channel Tx / 32-channel Rx Nova Medical head coil. The task
consisted of finger tapping following the protocol: 21s-ON, 21s-OFF x
12. One volunteer performed an additional task consisting of touching
the thumb with the index finger, i.e., adding somato-sensation. fMRI
was obtained with GE EPIK (GE EPI with keyhole) [10].
TR/TE = 3500/22 ms, FA = 85°, partial Fourier = 5/8, 3-fold
in-plane/3-fold inter-plane (multi-band) acceleration, matrix = 336 ×
336 × 123 slices, voxel size = 0.63 × 0.63 × 0.63 mm3.
Pre-processing was performed with ten different pipelines (Fig.
1a).
The most complete pipeline included slice-timing correction,
realignment, regression of motion parameters, regression of the pulse
and respiration (RETROCOR [11]),
regression of the mean signal of the cerebrospinal fluid (CSF) and
white matter (WM), and regression of the pre-processed phase image
[12].
Smoothing was also added to one pipeline for comparison. The effects
of pre-processing were evaluated in the resting-state data in terms
of mean correlation with noise, power of low-frequency fluctuations
in the grey matter (GM) and signal homogeneity (a joined measure of
temporal correlation and coherence in line crossing the cortical
ribbon) (Fig.
2).
The specificity of the corrected evoked responses was assessed by
computing line-activation profiles and the t-statistic GM/CSF ratio
(Fig.
3).Results
High-resolution
GE-data which was only subjected to realignment resulted in high
levels of noise contamination (Fig.
1b).
The addition of a band-pass filter, but not motion regression,
significantly decreased the influence of non-neuronal signals. In
contrast, voxel-wise regression of the phase signal, previously
pre-processed in the same way as the magnitude image (here,
slice-timing correction, realignment and filtering) resulted in a
significant decrease in signal contamination (pre-processing pipeline
#6 vs. #2 in Fig.
1b).
Application of the phase-based regression at later stages did not
significantly reduce noise. Regression of the physiological
recordings and added CSF/WM regression achieved maximum signal
cleaning. More complete pipelines also resulted in higher signal
power in the cortical GM (Fig.
2b).
The addition of smoothing increased the power within the GM,
presumably due to averaging with CSF voxels, but led to the highest
signal homogeneity (blurring) (see #9 vs. #10 in Fig.
2c).
Signal individuality, e.g., low similarity between neighbour voxels,
was maximized by regressing out motion, physiological and CSF/WM
tissue signals (see #5 and #9 in Fig.
2c).
Signals evoked in the pre-central gyrus by two different tasks were
best distinguished, based on laminar profiles, with pre-processing
pipelines #4, 5, 8 and 9, i.e., those including more complete
regression steps which identified an activation peak in deep and
intermediate depths in response to a motor task and motor + sensory
task, respectively (Fig.
3a).
The application of smoothing eradicated the definition of deeper
peaks, leaving only the typical GE-bias in superficial layers, near
the CSF. Group analysis demonstrated the significant effect the
phase-regression step had on the localization of evoked signals in
the GM, i.e., a higher average t-statistic in the GM with respect to
CSF (Fig.
3c).Discussion
Our
results demonstrate that, when subjected to certain pre-processing
steps, high-resolution GE-fMRI can be used to track laminar
responses. In general, we observed higher signal independence when
more pre-processing steps were applied to the data; however, the
smoothing step deteriorated the voxel heterogeneity. The use of
phase-regression, a method introduced by Menon [12],
produced a significant improvement in terms of signal cleaning,
especially when applied in little-pre-processed data; however, its
addition to already-highly-pre-processed data did not significantly
modify signal individuality or contamination, although it did improve
the localization of evoked responses in the GM. Interestingly, the
correction of partial volumes by removing CSF/WM did not
significantly improve the specificity of resting-state signals, i.e.
did not alter the homogeneity levels, possibly due to lower partial
volume effects in high-resolution images. The fMRI sequence employed
here is known to have a robust performance against field distortions,
therefore, common EPI sequences may require additional field
correction steps [13].Conclusion
Laminar-fMRI,
especially resting-state applications requiring whole-brain coverage,
could benefit from an optimal pre-processing (e.g. #8-#9), to exploit the benefits of GE-sequences.Acknowledgements
We thank Elke Bechholz for technical support, Claire Rick for proof-reading and the fMRI volunteers for their excellent cooperation.References
1. Kashyap,
S., et al., Resolving
laminar activation in human V1 using ultra-high spatial resolution
fMRI at 7T.
Sci Rep, 2018. 8(1):
p. 17063.
2. Huber,
L., et al., High-Resolution
CBV-fMRI Allows Mapping of Laminar Activity and Connectivity of
Cortical Input and Output in Human M1.
Neuron, 2017. 96(6):
p. 1253-1263 e7.
3. Polimeni,
J.R., et al., Laminar
analysis of 7T BOLD using an imposed spatial activation pattern in
human V1.
Neuroimage, 2010. 52(4):
p. 1334-46.
4. Havlicek,
M. and K. Uludag, A
dynamical model of the laminar BOLD response.
Neuroimage, 2020. 204:
p. 116209.
5. Kay,
K., et al., A
temporal decomposition method for identifying venous effects in
task-based fMRI.
Nat Methods, 2020. 17(10):
p. 1033-1039.
6. Fracasso,
A., et al., Laminar
imaging of positive and negative BOLD in human visual cortex at 7T.
Neuroimage, 2018. 164:
p. 100-111.
7. Kashyap,
S., et al., Impact
of acquisition and analysis strategies on cortical depth-dependent
fMRI.
Neuroimage, 2018. 168:
p. 332-344.
8. Heinzle,
J., et al., A
hemodynamic model for layered BOLD signals.
Neuroimage, 2016. 125:
p. 556-570.
9. Markuerkiaga,
I., M. Barth, and D.G. Norris, A
cortical vascular model for examining the specificity of the laminar
BOLD signal.
Neuroimage, 2016. 132:
p. 491-498.
10. Yun,
S., P. Pais-Roldán, and N.J. Shah. Detection
of Cortical Depth-dependent Functional Activation using Whole-brain,
Half-millimetre Resolution EPIK at 7T.
in ISMRM.
2020. Paris, France.
11. Glover,
G.H., T.Q. Li, and D. Ress, Image-based
method for retrospective correction of physiological motion effects
in fMRI: RETROICOR.
Magn Reson Med, 2000. 44(1):
p. 162-7.
12. Menon,
R.S., Postacquisition
suppression of large-vessel BOLD signals in high-resolution fMRI.
Magn Reson Med, 2002. 47(1):
p. 1-9.
13. Schallmo,
M.P., Weldon, K.B., Sponheim, S.R., Olman, C.A., Assessing
methods for geometric distortion compensation in 7T gradient echo
fMRI data.
bioRxiv, 2020.