Xiaowei Zou1, Venkata V Chebrolu2, and Nakul Gupta3
1Siemens Medical Solutions USA Inc., Houston, TX, United States, 2Siemens Medical Solutions USA Inc., Rochester, MN, United States, 3Department of Radiology, Houston Methodist Research Institute, Houston, TX, United States
Synopsis
Fat-suppressed turbo-spin-echo (TSE) imaging is very
important to visualize knee pathology such as cartilage defects in clinical routine.
Ultra-high-field 7T magnetic resonance imaging (MRI) provides higher
signal-to-noise-ratio (SNR) and contrast-to-noise-ratio (CNR) than 3T and 1.5T
MRI, enabling better visualization of fine anatomical structures and
physiological effects. However, imaging at 7T has higher receive and transmit
non-uniformity that may degrade its clinical value. Recently, a UNIform
COmbined RecoNstruction (UNICORN) algorithm was proposed for reducing the receive
non-uniformity. The purpose of this preliminary
work is to quantitatively and qualitatively evaluate and optimize UNICORN
performance for 7T clinical TSE with fat suppression.
INTRODUCTION:
Ultra-high-field 7T magnetic
resonance imaging (MRI) provides higher signal-to-noise-ratio (SNR) and
contrast-to-noise-ratio (CNR) than 3T and 1.5T MRI, enabling better visualization
of fine anatomical structures and physiological effects. Imaging at 7T,
however, has higher receive and transmit non-uniformity that may degrade its
clinical value. Recently, a UNIform COmbined RecoNstruction (UNICORN) algorithm
was proposed for reducing the receive non-uniformity1. In that study, three
radiologists qualitatively evaluated UNICORN performance on non-fat suppressed
7T turbo-spin-echo (TSE) knee MR images for SNR, uniformity, contrast, and
overall image quality. The three radiologists preferred UNICORN with high
inter-rater-agreement for uniformity, contrast, and overall image quality. The inter-rater
agreement regarding SNR was moderate, possibly because the non-fat suppressed 7T
TSE images had high enough SNR. Because fat suppression is very important in
clinical knee imaging to enhance the visibility of pathology such as cartilage
defects, the purpose of this preliminary work is to quantitatively and qualitatively
evaluate and optimize UNICORN performance for 7T clinical TSE with fat suppression.METHODS:
Data Acquisition:
This work was conducted as part
of an institutional review board approved clinical trial performed on a MAGNETOM
Terra scanner (Siemens Healthcare, Erlangen, Germany) using a single-channel
transmit, 28-channel phased-array receive knee coil (QED, Quality
Electrodynamics, Mayfield Village, OH). One subject’s knee was imaged in a
sagittal orientation using a two-dimensional TSE sequence with a refocusing
flip angle (FA) of 126o. The scan was repeated with a refocusing FA of
163o, keeping all other scan parameters identical. The acquisition
parameters included: field-of-view (FOV)
= 160 x 144 mm2, in-plane resolution = 0.4 x 0.4 mm2,
slice thickness = 2 mm, number of slices = 47, echo time (TE) = 31 ms, repetition
time (TR) = 6000 ms, spectral fat suppression ON, generalized auto-calibrating
partially parallel acquisitions (GRAPPA) acceleration factor = 2. In addition,
the transmit-field (B1) filter was turned ON to reduce the B1-induced
non-uniformity2. Scan time for each acquisition was 4:50 min.
Image Reconstruction:
The raw measurement data was retrospectively reconstructed
using a prototype implementation of the UNICORN algorithm with eight different
parameter settings denoted as RRx1 to RRx8 (Table 1). Each setting
reconstructed three image series per acquisition: no normalization (original), with
UNICORN normalization (UNICORN), and with both UNICORN and B1 normalization
(UNICORN+B1). There were 48 total reconstructed image series.
Quantitative Image Analysis:
Quantitative image analysis was performed in MATLAB (MathWorks,
Natick, MA). A board-certified body radiologist manually drew five regions of
interest (ROIs) on the original images: one large region of cartilage (number
of voxels = 478), one large region of skeletal muscle (number of voxels = 9031),
two adjacent small regions of cartilage (number of voxels = 207) and synovial fluid
(number of voxels = 121), and one large region of background for noise estimation
(number of voxels = 814). The SNR of the two large muscle and cartilage ROIs was
defined as the mean voxel intensity of each ROI divided by the mean voxel
intensity of the noise ROI. A metric of uniformity, UNI, was defined as the mean
voxel intensity of the two large ROIs divided by the standard deviation of the voxel
intensities within the same ROI. CNR was defined as the difference between the
mean voxel intensities of the small cartilage and fluid ROIs divided by the
mean voxel intensity of the noise ROI.
Qualitative Clinical Evaluation:
The radiologist, who was blinded to the quantitative
ROI analysis, also rated each image series on a 0-3 scale (0 = worst, 3 = best)
in terms of SNR, uniformity, contrast.RESULTS:
As shown in Figure 1, the UNICORN+B1 images of RRx4 (ScaleMode
= 2, CenterMode = 2, and SensitivityEnhancement
= 0) had the highest UNI of the large cartilage and muscle ROIs, while the
UNICORN+B1 images of RRx5 (ScaleMode = 0, CenterMode = 0, and SensitivityEnhancement = 1) had the
highest SNR of these two ROIs as well as the highest CNR. Exemplary images of
the original and UNICORN+B1 images of RRx1 to RRx8 are displayed in Figure 2. Clinical
evaluation showed that all the settings had very good SNR (all ratings = 3) and
contrast (all ratings = 3). Compared with the original images, the UNICORN+B1
images of all the settings had the same or better uniformity (Fig. 1),
consistent with the findings from the quantitative analysis. DISCUSSION:
This preliminary work
demonstrates that UNICORN plus B1 filtering is a promising reconstruction tool
to improve image quality of clinical fat-suppressed TSE of the knee at 7T. The
limitations of this work are small sample size and lack of inter-rater
validation. Future work will include more subjects and more ratersCONCLUSION:
UNICORN is a promising method to improve uniformity and
contrast of 7T clinical fat-suppressed TSE images of the human knee. Acknowledgements
This work is supported by Methodist - Siemens Collaborative Research Funding. References
1. Chebrolu VV, Kollasch PD, Deshpande V, Grinstead J, Howe
BM, Frick MA, Fagan AJ, Benner T, Heidemann RM, Felmlee JP, Amrami KK. Uniform
combined reconstruction of multichannel 7T knee MRI receive coil data without
the use of a reference scan. J Magn Reson Imaging. 2019 Nov;50(5):1534-1544.
doi: 10.1002/jmri.26691. Epub 2019 Feb 19. PMID: 30779475.
2. Jellus, V., and B. Kiefer. "Optimization of the
homomorphic filter for bias field correction." In Proc Intl Soc Magn Reson
Med., 2005, Vol. 13, p.2247.