Øystein Bech Gadmar1, Anne-Hilde Farstad2, Berit Elstad2, Piotr Sowa2, and Wibeke Nordhøy1
1Diagnostic Physics, Oslo University Hospital, Oslo, Norway, 2Radiology, Oslo University Hospital, Oslo, Norway
Synopsis
A MatLab-based inversion
recovery sequence simulator/calculator was developed with the purpose
of determining and testing optimal parameters for 3D IR acquisitions
with the purpose of detecting Multiple Sclerosis lesions in brains.
Single inversion FLAIR and dual inversion DIR sequences were studied
including a “True-T2” DIR sequence removing the undesired T1
weighting inherent in IR sequences to improve lesion-WM contrast. A
T2 preparation phase further helps facilitate T1 suppression.
Optimized IR sequences were
tested on healthy volunteers and some MS patients, on 1.5 T and 3.0 T
MR scanners. Good lesion contrast efficiency with high SNR was found
for True-T2 DIR.
Introduction
A simulation
tool for inversion recovery MR sequences was developed and used to
improve brain lesion detection in test persons including MS patients.
These contrast optimized sequences were tested on MR scanners with promising results.Methods
A Matlab-based inversion
recovery (IR) sequence simulator was developed to determine sequence
parameters optimizing signal-to-noise (SNR) and contrast-to-noise
(CNR) efficiency (SNR/t and CNR/t, resp.) for white matter lesions
(WML) versus background signal. 3D TSE fluid attenuated IR (FLAIR)
and double inversion (DIR) sequences (Siemens SPACE, GE CUBE readout)
were optimized with CNR/t as the main target. Optimized sequences
were tested on 1.5 T and 3 T MRI scanners on 2 healthy volunteers
with visible WML commonly associated with normal aging and on 5
multiple sclerosis (MS) patients.
DIR is traditionally used for
simultaneous suppression of cerebrospinal fluid (CSF) and WM. IR
produces T1 weighting that counteracts the desired T2 contrast. In
this study we used the idea described by Madhuranthakam et al1
to suppress the T1 contrast between WML and WM instead of WM,
improving T2 lesion contrast. This sequence is denoted “True-T2”
instead of the standard WM-nulled DIR. The further influence of a T2
preparation method reducing tissue T1 weighting was also studied2.Results
The
“IR-Calc”
Matlab
application calculates inversion and effective echo times given
sequence parameter settings including TR, inversion efficiency,
readout time and T2 preparation time along with relaxation times
based on literature values at desired field strengths. These
calculations are illustrated and corroborated by longitudinal
magnetization evolution simulation. Actual sequence settings on
different scanners may differ somewhat from the simulated values due
to different vendor implementations; this was taken into account in
the scanner tests.
Figure
1 shows a simulated FLAIR sequence using settings typical of vendor
implementations optimized for efficient scanning at the cost of some
image quality. The negative impact of short TR on the resulting T2
contrast (fig. 1b) is somewhat offset by the use of T2 preparation,
effectively making the inversion recovery of the tissue signals more
of a saturation recovery. Nevertheless, T1 weighting between WM and
WML leads to the need for a long TE to achieve optimal lesion
contrast, impacting both SNR and CNR negatively.
In
figure 2 a simulated True-T2 DIR sequence is shown. Even though
double inversion inherently leads to less signal intensity at
readout, the removal of WM-WML T1 weighting in this sequence enables
a shorter optimal TE of 86 ms, resulting in at least as good SNR and
CNR as a FLAIR sequence with the same acquisition time. Note that the
TR in this simulation is longer than in the previous one, which is
also beneficial for the image quality even accounting for total
acquisition time. In the simulations, this is reflected by an
increase in contrast efficiency (CNR/t) as well as the CNR itself.
Figure
3 shows DIR images acquired on a 1.5 T MR scanner of a healthy
volunteer displaying non-pathological WM lesions. The need for
moderate/short TE in the SNR starved standard WM-nulled sequence
leads to nearly no lesion-to-GM contrast particularly at 1.5 T. The
True-T2
DIR image
shows high contrast between lesion and all other tissues.
Figure
4 shows MS patient images acquired on a 1.5 T MR scanner. The FLAIR
images display good SNR and lesion contrast, but the True-T2 DIR
images have superior contrast particularly vs GM. The WM signal in
the latter images looks noisier but this is likely due to the partial
WM suppression accentuating low-contrast anatomical differences.Discussion & Conclusion
Lesions
in WM appear more conspicuous in WM-nulled than in FLAIR images due
to the suppressed WM signal. However, due to poor SNR and CNR at 1.5
T, WM-nulled DIR adds no significant information compared to FLAIR.
In True-T2 DIR images, however, lesions are better contrasted both to
WM and GM, and are thus more detectable in juxtacortical/GM lesions
than in FLAIR images. SNR is also much improved from standard DIR and
comparable to FLAIR SNR. At 1.5 T, GM-associated lesions are often
poorly detectable in both FLAIR and WM-nulled images; these lesions
are somewhat better delineated by WM-nulled DIR at 3 T. With True-T2
and T2-preparated DIR however, SNR/CNR efficiency both in WM and
GM-associated lesions are superior compared to the other IR sequences
although optimized FLAIR with T2 preparation at 3 T also performs
well in this respect.
In conclusion, using True-T2
DIR sequences with T2 preparation and not too short TR wherever
available looks like a promising approach to MS lesion detection.Acknowledgements
No acknowledgement found.References
-
Madhuranthakam AJ et al. Optimized
Double Inversion Recovery for Reduction of T1
Weighting in Fluid-Attenuated Inversion Recovery. Mag Res Med
2012; 67(1):81–88
- Visser F et al. High-Resolution
Magnetization-Prepared 3D-FLAIR Imaging at 7.0 Tesla.
Mag Res Med 2010; 64:194–202.