Petros Martirosian1, Rolf Pohmann2, Martin Schwartz1,3, Thomas Kuestner4, Manuel Kolb4, Ahmed Othman4, Cecilia Zhang4, Klaus Scheffler2,5, Konstantin Nikolaou4, Fritz Schick1, and Ferdinand Seith4
1Section on Experimental Radiology, University of Tübingen, Tübingen, Germany, 2Max Planck Institute for Biological Cybernetics, Tübingen, Germany, 3Institute of Signal Processing and System Theory, University of Stuttgart, Stuttgart, Germany, 4Department of Diagnostic and Interventional Radiology, University of Tübingen, Tübingen, Germany, 5Department of Biomedical Magnetic Resonance, University of Tübingen, Tübingen, Germany
Synopsis
Pseudo-continuous-arterial-spin-labeling (PCASL) has been
successfully applied in the lung providing high quality perfusion images. The pulmonary blood flow and the
respiratory system interact closely: the intrathoracic pressure has impact on
the venous return. Therefore, in this work, we evaluate the effects of
intrathoracic pressure on lung perfusion by using PCASL imaging in
end-expiratory and end-inspiratory breath-hold. PCASL imaging is able to
detect changes of parenchymal lung perfusion caused by alterations of the
intrathoracic pressure. Perfusion signal measured under end-inspiratory condition were noticeably reduced as compared to
end-expiratory breath-hold. This correlated significantly with measured blood
flow volume through the pulmonary trunk.
Introduction and Purpose
Measurement of pulmonary perfusion
and visualization of its spatial distribution can be of clinical significance
in diseases affecting the pulmonary vessels, the pulmonary interstitium or in
bronchial carcinoma.1-3 Although the lungs receive the entire
cardiac output, perfusion imaging using MRI is still a challenging task due to
the low proton density, the breathing movements and the high pulsatility of the
pulmonary circulation. Recently, an ECG-triggered pseudo-continuous ASL (PCASL)
approach using balanced steady-state free-precession (SSFP) imaging was applied
for perfusion measurements of the lung and provides high image
quality and reliable quantitative perfusion maps even under free-breathing
conditions.4 However, the pulmonary blood flow and the respiratory
system influence each other: pulmonary vascular resistance, for example, is correlated
to the lung volume and the intrathoracic pressure has impact on the venous
return.5 In this work, we are aiming to evaluate the effects of
intrathoracic pressure on lung perfusion by using PCASL-bSSFP imaging in end-expiratory
and end-inspiratory breath-hold.Methods
Four healthy volunteers (age 28.8±3.6 years, male) were examined on a 1.5T MR scanner (Avantofit,
Siemens Healthcare, Erlangen, Germany). Three respiration schemes were
performed: 1) expiratory breath-hold; 2) inspiratory breath-hold; 3) and
free-breathing. For each respiration scheme, the lung perfusion and the blood
flow of the pulmonary trunk (TP) were measured.
The perfusion measurements were performed using the
ECG-triggered PCASL-bSSFP sequence by labeling the
pulmonary trunk during the systolic cardiac period (Figure 1). Labeling parameters were: duration, 400 ms;
post-labeling delay, 1000 ms; flip angle, 25°; repetition delay, 3 sec. Coronal
single-slice images were acquired in diastole of the next cardiac
cycle using following parameters: TR, 2.1 ms; TE, 0.9
ms; flip angle, 70°; slice thickness, 20 mm; in-plane resolution, 2.5×2.5 mm2;
matrix size, 144×192; readout bandwidth, 1260 Hz/pixel. In breath-hold
examinations, one label-control image pair was acquired and twelve
label-control image pairs were measured under free-breathing condition. A
proton-density weighted bSSFP image was also acquired at the start of each
scan. The scan time of breath-hold and free-breathing measurements was approx. 13
sec and 2 min, respectively.
Blood flow measurements of TP was performed using
standard phase-contrast imaging. For expiratory and inspiratory breath-hold
examinations prospective triggering was used and retrospective gating was
applied in free-breathing flow measurements.
PCASL images were
registered prior to further evaluation. Image registration were conducted by
LAP6,7 using an in-house developed MATLAB (The MathWorks, Natick, MA)
script. Perfusion signal was
quantified in manually drawn regions of interest in perfusion-weighted
(Control-Label) images. ROIs were carefully placed in the peripheral parenchyma
of the right and left lung to avoid contribution of macroscopic vessels. The TP
blood flow was quantified using Argus software on the scanner (Siemens Healthcare, Erlangen, Germany). The
relationship between parenchymal perfusion and TP blood flow volume was
calculated using Pearson’s correlation coefficient.Results
PCASL perfusion measurements of the lung and TP blood
flow measurements of healthy volunteers were successfully performed under all
three respiratory conditions (Table 1). Perfusion signal values (mean±std) averaged over all
subjects were 39.1±9.3 and 19.9±1.4 in expiratory and inspiratory breathing
conditions, respectively. Corresponding mean values for TP blood flow volume
were 90±7 ml/hb and 59±12 ml/hb. The results show a significant reduction of
parenchymal perfusion signal as well as TP blood flow volume measured in
inspiration compared to expiration. The relative changes in perfusion signal
and TP blood flow volume for expiratory and inspiratory conditions are highly
correlated (Pearson, 0.93). In contrast, the perfusion signal measured in free
breathing is almost identical to the values in expiration.
Figure 2 shows PCASL-bSSFP perfusion-weighted images of
three healthy volunteers measured under expiratory, inspiratory and
free-breathing conditions. Perfusion images acquired in expiratory breath-hold
and free-breathing show high signal in lung parenchyma. Both measurements
reveal no significant difference in perfusion signal. Free-breathing perfusion images
show higher SNR due to averaging of multiple measurements.Discussion and Conclusion
The
results of this preliminary study demonstrate that PCASL-bSSFP sequence is able
to detect changes of parenchymal lung perfusion caused by changes of the
intrathoracic pressure. Perfusion signal measured under end-inspiratory condition was noticeably reduced as compared to
end-expiratory breath-hold. This correlated significantly with the measured
blood flow volume through the pulmonary trunk. In contrast, the parenchymal
perfusion signal as well as TP blood flow volume measured under free-breathing
condition are almost identical to those obtained under expiratory condition.
Nevertheless, care should be taken when analyzing PCASL data acquired under
free-breathing. Especially in patients with shortness of breath, large changes
in lung volume can lead to a reduction in the measured perfusion signal.Acknowledgements
No acknowledgement found.References
1. Ohno Y, et al. Dynamic perfusion MRI: Capability for
evaluation of disease severity and progression of pulmonary arterial hypertension
in patients with connective tissue disease. J Magn Reson Imaging 2008;28(4):887-899.
2. Tsuchiya N, et al. Decrease of pulmonary blood flow
detected by phase contrast MRI is correlated with a decrease in lung volume and
increase of lung fibrosis area determined by computed tomography in
interstitial lung disease. Eur J Radiol 2016;85:1581-1585.
3. Sauter AW, et al. Multifunctional profiling of
non-small cell lung cancer using 18F-FDG PET/CT and volume perfusion CT. J Nucl
Med 2012;53:521-529.
4. Seith F, et al. Imaging pulmonary blood flow using
pseudocontinuous arterial spin labeling (PCASL) with balanced steady-state
free-precession (bSSFP) readout at 1.5T. J Magn Reson Imaging 2020;52:1767–1782.
5. Pengo MF, et al. Cardiorespiratory interaction with
continuous positive airway pressure. J Thorac Dis
2018;10(Suppl 1):57-70.
6.
Gilliam C, Küstner T, Blu T. 3D
motion flow estimation using local all-pass filters. IEEE International
Symposium on Biomedical Imaging (ISBI); 2016. p 282-285.
7.
Küstner T, Schwartz M, Gatidis
S, Schwenzer NF, Schmidt H, Yang B, Gilliam C, Blu T, Schick F. Local All-Pass
Filter (LAP): Efficient Optical Flow-Based Image Registration. Proc ISMRM
Workshop on Motion Correction in MRI and MRS; 2017.