Verónica Aramendía-Vidaurreta1,2, Pedro M. Gordaliza3,4, Marta Vidorreta5, Rebeca Echeverría-Chasco1,2, Gorka Bastarrika1,2, Arrate Muñoz-Barrutia3,4, and María A. Fernández-Seara1,2
1Radiology, Clínica Universidad de Navarra, Pamplona, Spain, 2IdiSNA, Instituto de Investigación Sanitaria de Navarra, Pamplona, Spain, 3Universidad Carlos III de Madrid, Madrid, Spain, 4Instituto de Investigación Sanitaria Gregorio Marañón, Madrid, Spain, 5Siemens Healthineers, Madrid, Spain
Synopsis
This work investigated the
impact of three different breathing strategies, named breathhold (BH), synchronized-breathing (SB) and free-breathing (FB), together with motion detection and
correction algorithms in myocardial arterial spin labeling (ASL) images. Results indicate the superiority of FB combined with pairwise
registration, which showed higher accuracy (in synthetic images) and higher intrasession
reproducibility together with lower variability across subjects (in in vivo
images). BH and SB after motion detection provided similar results, but their
practical application is more complicated as it demands the subject's
collaboration to follow the respiratory pattern in SB or perform the apneas in
BH.
INTRODUCTION
Arterial spin labeling (ASL) is
a promising non-contrast alternative to first-pass imaging for the
quantification of myocardial blood flow (MBF). Nevertheless, the presence of
cardiac and respiratory motion hinders its application. Motion effects can be minimized
with ECG-gating and breathing strategies, such as breathhold (BH)1, synchronized-breathing (SB)2,3 or free-breathing (FB) combined with image registration4,5.
For the non-rigid registration
of ASL images, pairwise5,6 or
groupwise7 approaches are available. The pairwise method
registers first a reference label-control pair and subsequently all other label
and control images independently to their corresponding reference, to minimize
the effects of label-control intensity differences. In contrast, the groupwise approach
registers simultaneously the entire dataset (baseline, label and control images).
This work aimed to investigate
the impact of different breathing strategies together with motion detection and
correction algorithms in myocardial ASL images and to assess reproducibility of
the perfusion measurements.METHODS
Protocol:
12 healthy subjects were
subjected to a cardiac MRI exam on a 3T Skyra with an 18-channel array coil.
The scanning protocol is presented in Table1. The ECG-gated FAIR-ASL sequence included four
presaturation pulses, a hyperbolic-secant adiabatic inversion pulse, inversion
time (TI) of 1s and bSSFP readout. A baseline image (M0) was also acquired. The
ASL sequence was run under different breathing strategies (BH, SB and FB),
twice in each condition, to assess intrasession reproducibility. Respiratory
motion was expected to be minimized in the BH and SB strategies, in which,
unlike FB, there is an effort to freeze motion due to respiration.
Motion
detection:
A voxel within the anterior
myocardial segment was manually selected in an image acquired at expiration. The
coefficient of variation (CVvoxel), computed as the ratio of the
standard deviation (SD) to the mean intensity across images, was calculated at
this location.
Sequences were classified as
having low (CVvoxel<=10%) or moderate (CVvoxel>10%)
motion.
In the group of low motion,
no outliers were discarded. In the group of moderate motion, outliers were
identified as those whose intensity at this voxel was deviated by more than 2SD
below the mean.
Motion
correction:
The
performance of aforementioned registration approaches was tested in synthetic
and in vivo ASL images. Two synthetic datasets were created based on the
appearance of in vivo images, simulating signal with and without perfusion. Motion
between label-control image pairs was introduced in all directions (Figure 2A). Both
sets were corrupted with Gaussian noise. In synthetic data, the
algorithms’ performance was assessed by comparison of mean intensity values at
the myocardium (before and after registration), which were used to compute the accuracy
error, as the absolute difference between measured and reference intensity divided
by the reference intensity. Registrations were run with Elastix8.
In
vivo data analysis:
After motion detection and
correction, eight in vivo perfusion datasets were analyzed (Figure 1). Control (C)
and label (L) images were pairwise subtracted and averaged. Myocardial regions
of interest were manually drawn from the average perfusion-weighted image (in SB
and FB datasets) and from each perfusion-weighted image (in BH dataset). Outliers
in the perfusion-weighted image series were identified as ±2SD from the mean.
MBF (ml/g/min) was estimated as:
$$$\frac{60 \cdot \lambda \cdot (C - L)}{2 \cdot M_{0} \cdot TI \cdot exp^{\frac{TI}{T1_{blood}}}}$$$ $$$\lambda$$$=1ml/g; T1blood=1.664s. Data
assessment
was done by comparison of the MBF variability found across subjects and
intrasession reproducibility.
Statistical
analysis:
Friedman
test, within-subject coefficient of variation (wsCV) and Bland-Altman plots.RESULTS AND DISCUSSION
After motion detection, the percentage
of detected outliers (mean±SD: 30 ±10% in FB and 16±16% in SB)
was significantly greater in the FB than SB datasets (p=0.03,Wilcoxon-signed-rank).
Therefore, detected outliers were only discarded in the SB strategy.
Figure 2(B-C) shows the results
obtained in the myocardium after the registration of synthetic images. In the
set with no perfusion, the intensity across images presents low variability. In
the set with perfusion, an alternating label-control zigzag pattern is
observed. The average perfusion-weighted signal was lower than the reference
value and the accuracy error in the perfusion-weighted images (in bold) was
higher in the groupwise approach.
Figure 3 shows boxplots of MBF measurements in vivo. The Friedman test revealed significant differences (p=5.65·10-8)
across datasets. Post-hoc comparisons after Hochber correction showed that
these differences lied between the original FB and all other (p=0.05 for all
comparisons), but the original SB dataset. This can be explained by the
overestimation in MBF values and large variability across subjects found in the
original datasets due to the fact that, in the presence of motion, perfusion
values are contaminated with the intensity signal of the blood pool.
Figure 4 shows Bland-Altman plots and
the wsCV of MBF measurements in vivo. BH,
SBDD and FBP yielded the best reproducibility with wsCV
of 16%, 16% and 14%, respectively. Bland-Altman showed a lack of agreement in
the original SB and FB perfusion measurements. After
motion detection or correction, limits of agreement were narrower, indicating
more confident measurements.CONCLUSION
Synthetic and experimental results agreed in the
superiority of FB after pairwise registration, which showed higher accuracy (synthetic
images) and higher intrasession reproducibility and lower variability across
subjects (in vivo images). BH and SB after motion detection provided similar
results, but their practical application is more complicated
because it requires the subject’s collaboration.Acknowledgements
This work has been supported
by Asociación de Amigos de la Universidad
de Navarra and Banco Santander.References
1. Z.
Zun, E. C. Wong, and K. S. Nayak. Assessment of myocardial blood flow (MBF) in
humans using arterial spin labeling (ASL): feasibility and noise analysis. Magn.
Reson. Med. vol. 62, no. 4, pp. 975–983, 2009, doi: 10.1002/mrm.22088.
2. B. P. Poncelet et al. Measurement of human myocardial
perfusion by double-gated flow alternating inversion recovery EPI. Magn.
Reson. Med. vol. 41, no. 3, pp. 510–519, 1999, doi:
10.1002/(SICI)1522-2594(199903)41:3<510::AID-MRM13>3.0.CO;2-G.
3. V. Aramendía-Vidaurreta, R. Echeverría-Chasco, M. Vidorreta,
G. Bastarrika, and M. A. Fernández-Seara. Quantification of Myocardial
Perfusion With Vasodilation Using Arterial Spin Labeling at 1.5T. J. Magn.
Reson. Imaging, 2020, doi: 10.1002/jmri.27396.
4. D. J. J. Wang, X. Bi, B. B. Avants, T. Meng, S. Zuehlsdorff,
and J. A. Detre. Estimation of perfusion and arterial transit time in
myocardium using free-breathing myocardial arterial spin labeling with
navigator-echo. Magn. Reson. Med., vol. 64, no. 5, pp. 1289–1295, 2010,
doi: 10.1002/mrm.22630.
5. V. Aramendía-Vidaurreta, A. García-Osés, M. Vidorreta, G.
Bastarrika, and M. A. Fernández-Seara. Optimal repetition time for free
breathing myocardial arterial spin labeling. NMR Biomed., no. January,
p. e4077, 2019, doi: 10.1002/nbm.4077.
6. A. Javed, T. R. Jao, and K. S. Nayak. Motion correction
facilitates the automation of cardiac ASL perfusion imaging. 2015. doi:
10.1186/1532-429X-17-S1-P51.
7. W. Huizinga et al., “PCA-based groupwise image
registration for quantitative MRI,” Med. Image Anal., vol. 29, pp.
65–78, 2016, doi: 10.1016/j.media.2015.12.004.
8. S. Klein, M. Staring, K. Murphy, M. A. Viergever, and J. P. W.
Pluim, “Elastix: A toolbox for intensity-based medical image registration,” IEEE
Trans. Med. Imaging, vol. 29, no. 1, pp. 196–205, 2010, doi:
10.1109/TMI.2009.2035616.