Mohamad Abdi1, Daniel S Weller1,2, and Frederick H Epstein1,3
1Biomedical Engineering, University of Virginia, Charlottesville, VA, United States, 2Electrical and Computer Engineering, University of Virginia, Charlottesville, VA, United States, 3Radiology, University of Virginia, Charlottesville, VA, United States
Synopsis
Conventionally in MRI, respiratory
motion leads to shifts of tissue position in the image domain that correspond
to linear phase errors in the k-space domain. For DENSE, in addition to
position shifts, respiratory motion is displacement-encoded in the stimulated
echo, leading to a constant phase error in the k-space domain. We show that in
segmented DENSE acquisitions, motion compensation can be applied using per-segment
linear and constant phase corrections. As constant phase corrections using image-based
navigators are challenging, we show that deep leaning is potentially an effective
solution using simulated training data.
Purpose
Conventionally in MRI, respiratory motion leads
to shifts of tissue position in the image domain that correspond to linear
phase errors in k-space. For segmented
acquisitions, each k-space segment corresponds to a unique breathing-induced
position shift, and a corresponding linear phase correction can be applied to
each segment of k-space to perform motion compensation1. Image-based navigators are one method to
estimate the per-segment position shift and corresponding linear phase
correction2. For
displacement encoding with stimulated echoes (DENSE), here we show that in
addition to the position shift, respiratory motion (assumed to be rigid) leads
to a phase shift in the image domain, which corresponds to a constant phase
error in k-space. Thus, to perform
motion compensation for segmented DENSE, per-segment linear and constant phase
corrections must be computed and applied.
Because DENSE typically employs both in-plane displacement-encoding
gradients and through-plane dephasing gradients3, constant phase
errors may arise from both in-plane and through-plane respiratory motion, which
presents challenges to an iNAV-based approach. Using simulations, we show the feasibility of deep learning (DL) to
compute the linear and constant phase corrections and to perform motion
correction.Theory
The displacement-encoded stimulated echo in
DENSE is described in Equation 1: $$$M_{DE}=\
M(x)\ e^{-j2\pi k_e \Delta x_c}$$$ where the
signal phase is proportional to the product of the displacement-encoding
frequency, ke, and the myocardial displacement, Δxc4.
The effect of displacement due to respiratory motion (ΔxR) leads to
Equation 2:
$$$\
M_{DE}^R=\ M\left(x-\Delta x_R\right)e^{-j2\pi k_e\Delta x_c}\ e^{-j2\pi
k_e\Delta x_R}$$$ which accounts
for both the shifted position of the magnetization as well as the phase shift
resulting from displacement encoding of ΔxR. The position and phase shifts
due to respiratory motion in the image domain lead to linear and constant phase
errors, respectively, in k-space. For a segmented DENSE acquisition, the
respiratory-corrupted signal, S, can be written as in Equation 3:
$$$S(x)=\sum_{i=1}^{L}{F^HU_iF\ M_{DE}^{R_i}}$$$ , where F is the Fourier transform, L is the number of segments, Ui is the
sampling function of the ith segment, and H is the Hermitian
transpose operator. Equation 4:
$$$S_{mc}(x)=\sum_{i=1}^{L}{F^H\left[e^{j2\pi\
k\Delta x_{R_i}}e^{j{2\pi\phi}_i}\ \left(U_iFM_{DE}^{R_i}\right)\right]}$$$
shows how per-segment linear and constant phase
corrections can be applied in k-space to correct respiratory-induced artifacts of
the displacement-encoded stimulated echo, where k is the independent spatial
frequency variable, ΔxR is the
slope of the linear phase error,
φi is the
constant phase error, and Smc is the motion-compensated image. This
idea is demonstrated in Figure 1, where we used a computational deforming-heart
phantom and Bloch equation simulations of the stimulated echo. The position of
the phantom was shifted over time using a sine wave to simulate respiratory
motion during imaging. Blurring and signal cancellation artifacts due to
motion-induced position and phase shifts are shown in Figure 1, as are
motion-compensated images computed using the motion correction of Equation 4. We
postulated that DL would provide effective motion compensation for respiratory-corrupted
DENSE, and we pursued this idea using computer simulations.Methods
A
Bloch-equation-based DENSE simulator was used to simulate free-breathing
acquisitions of a computational deforming heart phantom. Respiratory motion was
simulated using a sine wave. Training data were generated by varying the
respiratory period, motion magnitude and initial phase in the ranges 15-25 cycles/min,
0-15 mm, and 0-0.5 cycles, respectively. A spiral cine DENSE acquisition with the
following parameters was simulated: in-plane spatial resolution of 2.5×2.5 mm2,
balanced 3-point displacement encoding with displacement encoding frequency of
0.06 cyc/mm, 4 spiral interleaves per image with 1 interleaf acquired per
heartbeat. We used 21,600, 4,800 and 2,400 simulated images for training,
testing and validation, respectively.
An encoder-type convolutional neural network (CNN) with two convolutional layers was implemented to
estimate the linear and constant phase corrections. Max-pooling and dropout are followed after each convolutional layer and a fully connected layer is used as final layer. A
diagram of the CNN and its training is illustrated in Figure 2. After training,
per-segment linear and constant phase corrections were estimated using the CNN
for the images in the test set. The corrections were implemented using Equation
4. The normalized difference (ND) defined in Equation 5:
$$$ND\
=\frac{\left|\phi_{estimated}\right|-\left|\phi_{truth}\right|}{\phi_{truth}^{max}-\phi_{truth}^{min}\
\ }$$$ was used
to quantify the accuracy of estimations compared to the ground-truth values.Results
Figure 3 shows examples of Bloch-equation-simulated
motion-corrupted DENSE images and their corresponding motion-compensated images
(compensated using Equation 4 with CNN-derived phase correction values), as
well as the ground-truth images. Two examples are shown: one from an early cardiac
frame where the effect of the constant phase error is insignificant and another
from a later cardiac frame where both artifact types are significant. The plots
of ND show that the CNN computed phase correction values within approximately
6% and 2% of the true values.Discussion
We developed a new motion model to describe the
effects of respiratory motion in DENSE imaging. The model incorporates the respiratory-motion-induced
phase shift of the stimulated echo, in addition to the position shift which is
common to all MRI. Further, we showed
the potential of DL as a method for performing motion compensation.
Further work will augment the training data with vivo DENSE images and evaluate
the method for in vivo DENSE imaging.Acknowledgements
This work was supported by R01HL147104.References
1. P.G. Batchelor et al. Magn
Reson Med. 2005 Nov;54(5):1273-80. doi: 10.1002/mrm.20656.
2. M. Henningsson et al, Magn Reson Med . 2012
Feb;67(2):437-45. doi: 10.1002/mrm.23027.
3. Zhong et al. Magn Reson Med. 2006
Nov;56(5):1126-31. doi: 10.1002/mrm.21058.
4. D. Kim et al. Radiology.
2004 Mar;230(3):862-71. doi: 10.1148/radiol.2303021213. Epub 2004 Jan 22.