Zu Tao1, Sun Yi2, Wu Dan1, and Zhang Yi1
1Key Laboratory for Biomedical Engineering of Ministry of Education, Department of Biomedical Engineering, College of Biomedical Engineering & Instrument Science, Zhejiang University, Hangzhou, China, 2MR Collaboration, Siemens Healthcare Ltd., Shanghai, China
Synopsis
The clinical use of chemical exchange saturation
transfer (CEST) imaging is limited by its relatively long scan time, because it
typically acquires multiple saturation image frames. Here, a novel auto-calibrated
reconstruction method by joint k-space and image-space parallel imaging (KIPI)
is proposed for faster CEST acquisition. By under-sampling CEST image frames with
variable acceleration factors, KIPI allows an acceleration factor of up to 8-fold
for acquiring source images, yielding a net speed-up of 6-fold in scan time,
and produces image quality close to that of the ground truth.
Introduction
Chemical exchange saturation transfer (CEST) is an
emerging MRI technique that amplifies the detectability of certain
low-concentration biomolecules through their interaction with the abundant water
pool.1,2 However, its
routine clinical use is limited by the relatively long scan time since CEST MRI
acquires multiple image frames.3,4 Recently, a
variably-accelerated SENSE (vSENSE) method was proposed to speed up the CEST acquisition.5,6 Here,
we propose a novel auto-calibrated reconstruction method by joint k-space and
image-space parallel imaging (KIPI) as a further development of the vSENSE
approach.Theory
As an auto-calibrated k-space reconstruction method,
GRAPPA7 estimates the
missing k-space point by a linear combination of the acquired data in the
neighborhood from all channels using the following equation,$$\begin{equation}y=Aw\tag{1} \end{equation}$$where $$$A$$$
represents
the source matrix organized from the auto-calibration signal (ACS) data, $$$y$$$
denotes the target matrix, and $$$w$$$
represents the weights to be
fitted. GRAPPA generates
accurate images robustly when the acceleration factor (AF) is low.8 However, the
GRAPPA weights estimated can be affected by the underlying image contrast. Figure
1 shows images of different contrasts simulated according to Biot-Savart
law9 with eight
receive channels. It is clear that the GRAPPA weights calculated from one
contrast yield significant errors when being applied to the other contrasts (Figs.
1e-1f and 1i-1j).
SENSE10
poses the reconstruction as a linear inverse problem in image space and theoretically yields the optimal solution. However, the accuracy of SENSE
reconstruction depends on the accuracy of the sensitivity maps used, which are
prone to errors in practice.5,6 If the coil
image $$$m_{i}$$$
of channel
$$$i$$$
($$$1\leq i\leq N$$$) is perfectly known, the sensitivity map $$$SE_{i}$$$
can be
obtained by dividing the coil-combined image $$$\rho$$$
into the
image from each channel.$$\begin{equation}SE_{i}=m_{i}/\rho\tag{2}\end{equation}$$
Thus, for retroactive SENSE reconstruction
with AF=2 (without loss of generality), there is $$\begin{equation}\begin{split}\begin{aligned}s_{N*1}&=m_{N*1}^{1}+m_{N*1}^{2}\\&=\rho^{1}SE_{N*1}^{1}+\rho^{2}SE_{N*1}^{2}\\&=\left[\begin{array}{ll}SE_{N*1}^{1}&SE_{N*1}^{2}\end{array}\right]\left[\begin{array}{l}\rho^{1}\\\rho^{2}\end{array}\right]\end{aligned}\end{split} \tag{3}\end{equation}$$ where $$$s_{N*1}$$$
represents the folded coil image
vector, and the superscript represents the spatial position. The least squares
solution to Eq. [3] is$$\begin{equation}\begin{split}\begin{aligned}\left[\begin{array}{c}\hat{\rho}^{1}\\\widehat{\rho}^{2}\end{array}\right]&=\left(SEE^{H}SEE\right)^{-1}SEE^{H}s_{N*1}\\&=\underbrace{\left(SEE^{H}SEE\right)^{-1}SEE^{H}SEE}_{I}\left[\begin{array}{l}\rho^{1}\\\rho^{2}\end{array}\right]=\left[\begin{array}{l}\rho^{1}\\\rho^{2}\end{array}\right]\end{aligned}\end{split} \tag{4}\end{equation}$$ where $$$SEE=\left[\begin{array}{ll}SE_{N*1}^{1}&\left.SE_{N*1}^{2}\right]\end{array}\right.$$$
and $$$I$$$
is the
identity matrix.
However,
the coil image $$$m_{i}$$$
is unknown
for SENSE imaging with AF > 1 in practice. The KIPI method calculates $$$m_{i}$$$
with GRAPPA reconstruction
from a low-AF calibration frame. Then, the sensitivity maps can be calculated
with Eq. [2], and be applied to the other frames with high-AF SENSE
reconstruction. Notably, the derivation above (Eqs. [2-4]) illustrates that the
SENSE solution using the sensitivity maps derived from the calibration frame itself
is identical to its GRAPPA solution.
Overall,
KIPI turns the original vSENSE approach into an auto-calibrated parallel
imaging method by incorporating both GRAPPA and SENSE reconstruction. During
data sampling, KIPI chooses an image frame as the calibration frame with low AF
with the ACS data, and undersamples the other image frames using variably high
AF without ACS data. During reconstruction, KIPI inherits the robustness of
auto-calibrated GRAPPA imaging, and avoids the drawback of its susceptibility
to underlying contrasts by calculating sensitivity maps. Furthermore, KIPI can
adopt the artifact suppression algorithm used in vSENSE.5,6 The KIPI approach
is summarized in Figure 2.
Methods
Human experiments were conducted on a 3T Siemens
Prisma MRI system using a 64-channel-receive head coil. A SPACE-CEST sequence11 was run using
the following readout parameters: FOV=212×212×201mm3,
resolution=2.8×2.8×2.8mm3, turbo factor=140, and GRAPPA factor=2×2 with
ACS acquired. Seven CEST saturation offsets for APT-weighted (APTw) imaging
were executed, including (S0), ±3, ±3.5, and ±4-ppm. And a dual-echo
3D GRE sequence was used for B0 field mapping.
For
retrospective reconstruction, the +3.5-ppm frame was selected as the
calibration frame with AF=2×2 to generate the sensitivity maps and correction
maps, and the remaining 6 frames had AF=2×4 (in the phase-encoding and partition-encoding
directions, respectively). Note that only the calibration frame retained ACS
data.Results
Figures 1g-h show KIPI allowed a calibration frame with a different
image contrast to be used, generating reconstruction errors substantially
smaller than those from GRAPPA (Figs. 1k-l vs. 1i-j). On the contrary,
the accuracy of GRAPPA was comprised when reusing the kernel weights across
different image contrasts (Figs. 1e-f).
The
conventional GRAPPA (AF=2×2) scan (Fig. 3a) was considered ground truth
here. KIPI (Fig. 3c) generated source images that better agreed with the
ground-truth results than those reconstructed from GRAPPA (Fig. 3b) using
the same data (AF=2×4) with RNMSE of 0.012 versus 0.032, respectively. Besides,
unfolding artifacts were evident in the GRAPPA images (Figs. 3b, d; red
arrows).
Figure 4 displays
the APTw images calculated from the source images of Fig. 3 after B0
correction and image alignment. Substantial artifacts can be seen in the APTw
images reconstructed by GRAPPA (Fig. 4b) while
the KIPI method generated APTw maps (Fig. 4c) that were consistent with
the ground truth (Fig. 4a).Conclusion
KIPI is a novel auto-calibrated parallel imaging
method that incorporates the advantages of k-space and image-space
reconstruction methods. KIPI allowed an acceleration factor of up to 8-fold for
acquiring CEST source images, yielding a net speed-up of 6-fold in scan time, and
essentially without compromising the image quality. KIPI should facilitate the
translation of CEST imaging in clinical routines.Acknowledgements
NSFC grant number: 81971605, 61801421.References
1.Ward KM, Aletras AH, Balaban RS. A new class of contrast agents
for MRI based on proton chemical exchange dependent saturation transfer (CEST).
J. Magn. Reson. 2000;143(1):79-87.
2.van Zijl
PCM, Yadav NN. Chemical Exchange Saturation Transfer (CEST): What is in a Name
and What Isn't? Magn Reson Med 2011;65(4):927-948.
3.Zhou JY, Lal B, Wilson DA, et al. Amide proton transfer (APT)
contrast for imaging of brain tumors. Magn Reson Med 2003;50(6):1120-1126.
4.Zhou J, Blakeley JO, Hua J, et al. Practical data acquisition
method for human brain tumor amide proton transfer (APT) imaging. Magn Reson Med 2008;60(4):842-849.
5.Zhang Y, Heo H-Y, Lee D-H, et al. Chemical Exchange Saturation
Transfer (CEST) Imaging With Fast Variably-accelerated Sensitivity Encoding
(vSENSE). Magn Reson Med 2017;77(6):2225-2238.
6.Zhang Y, Heo H-Y, Jiang S, et al. Fast 3D chemical exchange
saturation transfer imaging with variably-accelerated sensitivity encoding
(vSENSE). Magn Reson Med 2019;82(6):2046-2061.
7.Griswold MA, Jakob PM, Heidemann RM, et al. Generalized
Autocalibrating Partially Parallel Acquisitions (GRAPPA). Magn Reson Med 2002;47(6):1202-1210.
8.Blaimer M, Breuer F, Mueller M, et al. SMASH, SENSE, PILS,
GRAPPA: how to choose the optimal method. Top Magn Reson Imaging 2004;15(4):223-236.
9.Guerquin-Kern M, Lejeune L, Pruessmann KP, et al. Realistic
Analytical Phantoms for Parallel Magnetic Resonance Imaging. IEEE Trans Med Imaging 2012;31(3):626-636.
10.Pruessmann
KP, Weiger M, Scheidegger MB, et al. SENSE: Sensitivity encoding for fast MRI.
Magn Reson Med 1999;42(5):952-962.
11.Zhang Y, Yong X, Liu
R, et al. Whole-brain chemical exchange saturation transfer imaging with
optimized turbo spin echo readout. Magn Reson Med 2020;84(3):1161-1172.