Limin Zhou1, Yiming Wang1, and Ananth Madhuranthakam1,2
1Department of Radiology, University of Texas Southwestern Medical Center, Dallas, TX, United States, 2Advanced Imaging Research Center, University of Texas Southwestern Medical Center, Dallas, TX, United States
Synopsis
Pseudo-continuous arterial spin labeling
(pCASL) has been the recommended ASL method for brain, and one of the two
methods for kidneys. However, the reliability of labeling efficiency of pCASL is
a major concern, especially in renal imaging due to the increased B0
inhomogeneities. In this study, we implemented and evaluated the optimized unbalanced
pCASL gradient scheme generated by numerical simulation with both perfusion
phantom and 4 healthy volunteers. The results showed that this optimized
unbalanced pCASL gradient scheme was more robust to off-resonance than the
corresponding balanced pCASL gradient scheme.
Introduction
Recent consensuses have identified pseudo-continuous
arterial spin labeling (pCASL) as the labeling method of choice for ASL in both
brain (1) and kidneys (2). However, one of the major concerns of pCASL is the reliability
of labeling efficiency, especially in renal application. The increased off-resonance
at the labeling plane, which often occurs near the lungs, decreases labeling
efficiency and sometimes can lead to complete loss of perfusion signal (3). Recent
studies have shown improved robustness of B0 inhomogeneities in
brain by pCASL with unbalanced labeling scheme (4). Therefore, the goal of this
study was to demonstrate the increased robustness of pCASL in renal applications
using an optimized unbalanced gradient scheme, evaluated in both perfusion
phantom and healthy volunteers.Methods
To optimize the labeling efficiency, numerical simulations of
Bloch equations were performed to evaluate different unbalanced gradient schemes
for both tortuous flow and off-resonance as previously described (4, 5). The robustness
of pCASL sequence with the optimized unbalanced gradient scheme was tested both
on a previously designed 3D printed perfusion phantom (6) and 4 healthy
volunteers’ kidneys with IRB approval on a 3T MRI scanner (Ingenia,
Philips Healthcare).
Image Acquisition and Analysis: The off-resonance
effects were approximated by adding extra phase offsets between RF pulses
during the labeling. All pCASL images were reconstructed in MATLAB
including k-space filtering and complex k-space subtraction.
For perfusion phantom: pCASL images were acquired in a single axial plane using single
shot TSE with the following parameters: TR/TE
= 6000/16 ms, FOV = 150x150 mm2, matrix = 64x63, acquired resolution
= 2.34x2.38x3 mm3, echo spacing = 3.2 ms, ETL = 63, label duration =
1.8 s, post-label delay = 1.8 s, labeling RF interval = 1.2 ms, 1 repetition, 4
background suppression pulses and inflow saturation pulses. To compare the
robustness of balanced and unbalanced gradient scheme at different off-resonance, 2D pCASL
with 41 dynamics were performed for each phase offset ranging from -500 Hz to 500
Hz at 25 Hz increments. An unbalanced gradient scheme was also run with
Gmax/Gave ratio of 10 and 7,
with the Gave of 0.5 mT/m. The total acquisition time for each 2D
pCASL with 41 dynamics was 8:58 minutes.
For healthy volunteers:
All 2D pCASL images were acquired in a single coronal plane using single
shot TSE scans with the following parameters: TR/TE = 6500/42 ms, FOV = 223x375
mm2, matrix = 88x150, acquired resolution = 2.53x2.50x10 mm3,
reconstructed resolution = 0.59x0.59x10 mm3, echo spacing = 4.7 ms,
ETL = 83, label duration = 1.5 s, post-label delay = 1.5 s, labeling RF
interval = 1.2 ms, Gmax = 3.5 mT/m, Gave = 0.5 mT/m, 4 background suppression
pulses and inflow saturation pulses. To compare the robustness of unbalanced
and balanced gradient schemes to different off-resonance, 2D pCASL images were
acquired with 4 signal averages at each phase offset ranging from -300 Hz to 300
Hz at 50 Hz increments (total 13 dynamics). The total acquisition time for 2D
pCASL with 13 dynamics and 4 averages was 11:38 minutes. An additional set of
images without extra off resonance was also acquired using the unbalanced and
balanced gradient schemes with 16 signal averages at 3:24 minutes each. A separate
M0 was acquired in 7 seconds with the same acquisition parameters for renal blood flow
quantification.Results
The numerical simulations demonstrated
high labeling efficiency with the unbalanced pCASL gradient scheme with Gmax/Gave
ratio between 5 to 7 and Gave greater than 0.4 mT/m (Fig. 1) across broad
off-resonance (0-420 Hz) and blood velocity (0-123 cm/s). In perfusion phantom,
the unbalanced pCASL with gradient ratio of 10 (Fig. 2c) had lower labeling
efficiency than the unbalanced pCASL with gradient ratio of 7 (Fig. 2d), while
the control signal remained intact. This demonstrated that unbalance pCASL with
Gmax/Gave ratio of 7 will have higher signal difference between control and
label images and are more robust to off-resonance. This behavior is also
observed in normal volunteer’s kidneys, showing increased robustness to
off-resonance with unbalanced pCASL (Fig. 3-5). Figure 4 interrogates the
signal intensity and since the Gave in unbalanced pCASL control images is 0, it
is more robust to off resonance giving higher signal intensity than balanced
pCASL whose Gave in control is not equal to zero. Discussion and Conclusion
Numerical simulations suggest that high labeling efficiency
can be achieved using unbalanced pCASL gradient scheme with Gmax/Gave ratio
between 5 to 7 and Gave greather than 0.4 mT/m across a broad range of
off-resonance and blood velocities. This optimized unbalanced pCASL gradient
scheme (Gave = 0.5 mT/m, Gmax/Gave = 7) improves the robustness of pCASL
labeling to off-resonance effects in both perfusion phantom and healthy
volunteers’ kidneys. Acknowledgements
This work was partly supported by the NIH/NCI
grant U01CA207091.References
1. Alsop, D.C.,
et al., MRM, 2015. 73(1): p. 102-16.
2. Nery, F.,
et al., MAGMA, 2020. 33: p. 141-161.
3. Echeverria-Chasco,
R., et al., MRM, 2020. 85 (3): p. 1507-1521.
4. Zhao, L.,
et al., MRM, 2017. 78(4): p. 1342-1351.
5. Maccotta
L., et al., NMR Biomed, 1997;10(4–5):216–221.
6. Greer,
J.S., et al., ISMRM 2017:3805.