Kuan-Hung Cho1, Po-Hsun He1, Hsuan-Han Chiang1, Ming-Jye Chen1, Ezequiel Farrher2, Nadim Jon Shah2,3, Chang-Hoon Choi2, and Li-Wei Kuo1,4
1Institute of Biomedical Engineering and Nanomedicine, National Health Research Institutes, Miaoli County, Taiwan, 2Institute of Neuroscience and Medicine – 4, Forschungszentrum Juelich, Juelich, Germany, 3Department of Neurology, Faculty of Medicine, RWTH Aachen University, Aachen, Germany, 4Institute of Medical Device and Imaging, National Taiwan University College of Medicine, Taipei, Taiwan
Synopsis
For mouse brain imaging,
determining the size of a loop coil is a trade-off between SNR and imaging
coverage. Here, we propose a multi-purpose coil arrangement incorporating a
small loop coil and a position-adjustable add-on saddle coil to enhance the imaging
coverage of anatomical images and improve the SNR in the regions distant to the
surface coil. The result shows the imaging coverage can be adjustable for
targeting different brain regions and demonstrated this design concept in mouse
brain imaging experiments.
Introduction
In
pre-clinical MR experiments, a single loop coil has been widely used for
imaging the rodent brains due to its simple design and high signal-to-noise
ratio (SNR) close to the coil. The imaging coverage and sensitivity of the loop
coil are largely dependent on the size of the loop, but determining this is
always a trade-off. Some applications1 focus more on the cortical surface of the brain
(mostly close to the surface coil) whereas the others2 concentrate on the deeper brain region. In the
latter case, for example, the size of the loop coil needs to be enlarged to fulfill
the targeted imaging coverage resulting in the degradation in SNR. In order to
address this issue, we propose a simple but multi-purpose coil approach. A
small enough loop coil covering the limited cortical area with very high SNR
can be incorporated with a saddle coil. The add-on saddle coil is capable of
adjusting its position to improve
the imaging coverage and the SNR in the deeper brain regions, such as
hypothalamus. Moreover, it is placed perpendicularly to the small loop coil so
that it can be driven in quadrature and gain the benefits from the quadrature
mode3, 4. The function of the small loop coil is to maintain
sufficiently high SNR in the regions close to the surface coil, while that of the
add-on coil is to extend the imaging coverage in the regions distant to the small
loop coil and to improve SNR in both surface and deeper regions.Methods
MR experiments were
performed on a pre-clinical MRI platform incorporating a high-strength gradient
coil system dedicated for small animal imaging use on a human-size 3 T magnet5. The small loop coil for mouse brains (15 mm × 8 mm)
and the saddle coil (length of 20 mm, diameter of 15 mm, and angular aperture
of 75o) were built, respectively, as shown in Figure 1a and 1b. To be
operated in a quadrature mode and to minimize any coupling, both coils were placed
orthogonally (Figure 1c and 1d). The aforementioned small surface coil was also
used as a reference for the comparison. A fast spin echo sequence was used to
acquire the T2WI with the following parameters: in-plane resolution of 104 μm ×
104 μm, slice thickness of 0.8 mm, TR of 4000 ms, TE of 68 ms and 2 averages. The
SNR values on various regions-of-interests (ROIs) along the imaging plane
perpendicular to the surface coil were calculated.Results
Figure 2 shows the T2WIs of mouse brain acquired using one of
the configurations of the proposed coil assembly. At first glance, the imaging
depth of the proposed hybrid coil configuration is larger than that of the small-size
surface coil. Figure 3a shows the selected ROIs (ROI#1-4: close to the
surface coil; ROI#5-8: middle distance to the surface coil; ROI#9-12: farthest
from the surface coil) overlaid on the representative T2WIs corresponding to different
coil configurations, respectively. The SNR of all selected ROIs is shown in Figure 3b. It is observed that the SNR of the proposed combined
coil is superior to that of the small surface coil among all ROIs.Discussion&Conclusion
In this study, we have
successfully demonstrated that the proposed hybrid coil could significantly
improve the SNR and imaging coverage in the target brain region compared with
the conventional coil. By minimizing the coupling effect, the SNR in the
regions close to the small-size surface coil is still sufficiently high. This
proposed hybrid coil configuration could benefit the mouse brain MRI studies in
those applications associated with deep brain regions. Although we have only demonstrated
one configuration, the add-on coil can be positioned in any locations
vertically along the small loop coil and weight more on the surface cortical
area or far deeper brain region. In order to cover the deeper region, we normally
need to use either lossy volume coil or much larger surface coil. Both degrade
the quality of MR image which is not desirable. In conclusion, we proposed a multi-purpose
coil combining a small surface coil with a position-adjustable saddle coil,
which provides sufficiently high SNR in the brain regions close to the surface
coil and improves overall SNR especially in the brain regions distant to the
surface coil, respectively. Further examination is needed to evaluate the combined
coil performance at various positions, particularly, at the far deeper regions
of the mouse brain, such as hypothalamus and brainstem. Acknowledgements
This work was supported by the Ministry of Science and Technology, Taiwan [MOST-108-2911-I-400-502, MOST-109-2314-B-400-023, MOST-109-2321-B-400-006, and MOST-109-2221-E-400-001-MY2], and the National Health Research Institutes, Taiwan [NHRI-BN-109-PP-06 and NHRI-BN-109-SP-11].References
- Wu, K.-J.; Yu, S.-J.; Chiang, C.-W.; Lee, Y.-W.; Yen, B. L.;
Tseng, P.-C.; Hsu, C.-S.; Kuo, L.-W.; Wang, Y., Neuroprotective Action of Human
Wharton's Jelly-Derived Mesenchymal Stromal Cell Transplants in a Rodent Model
of Stroke. Cell Transplantation 2018, 27, (11), 1603-1612.
- Lizarbe, B.; Benítez, A.; Sánchez-Montañés, M.;
Lago-Fernández, L. F.; Garcia-Martin, M. L.; López-Larrubia, P.; Cerdán, S.,
Imaging hypothalamic activity using diffusion weighted magnetic resonance
imaging in the mouse and human brain. NeuroImage
64, 448-457.
- Ha, Y.; Choi, C. H.; Shah, N. J., Development and
Implementation of a PIN-Diode Controlled, Quadrature-Enhanced, Double-Tuned RF
Coil for Sodium MRI. IEEE Transactions on
Medical Imaging 2018, 37, (7),
1626-1631.
- Chen, C. N.; Hoult, D. I.; Sank, V. J., Quadrature detection
coils—A further √2 improvement in sensitivity. Journal of Magnetic Resonance (1969) 1983, 54, (2), 324-327.
- Cho,
K.-H.; Huang, S.-M.; Choi, C.-H.; Chen, M.-J.; Chiang, H.-H.; Buschbeck, R. P.;
Farrher, E.; Shah, N. J.; Garipov, R.; Chang, C.-P.; Chang, H.; Kuo, L.-W.,
Development, integration and use of an ultra-high-strength gradient system on a
human-size 3 T magnet for small animal MRI. PLOS
ONE 2019, 14, (6), e0217916.