Vitaliy Zhurbenko1, Juan Diego Sánchez Heredia1, Wenjun Wang1, and Jan Henrik Ardenkjær-Larsen1
1Technical University of Denmark, Kgs. Lyngby, Denmark
Synopsis
Imaging of low-abundance nuclei could benefit
from close proximity detecting coils. Close proximity is conveniently achieved
using flexible arrays, which conform to the shape of the subject. In this work,
a design of an 8-element ultra-flexible array for 13C imaging is
presented. Each array element includes a low-noise preamplifier with 13C/1H
active and passive decoupling, as well as high-power pulse protection circuits.
The array is lightweight and easy to handle. The design approach is directly scalable
to larger size arrays.
Introduction
Imaging of 13C isotope requires high
sensitivity detector arrays. One way to increase sensitivity is to implement
close proximity receiver arrays. To achieve close distance to a patient, the
array shape should conform to the shape of the body, which is conveniently
achieved by ultra-flexible coils. Ultra-flexible arrays can adopt to various
patient sizes as well as different body parts 1. In order to keep the high detection performance, which
is particularly important in imaging of low-abundance nuclei, a low-noise
amplification circuitry may be co-integrated with the coil. In order to support
flexibility of the coil and overall handling convenience, this circuitry must
have a compact size, lightweight, and compatible with MR environment, which is
challenging to attain.Methods
To fulfill the requirements outlined above, a compact,
ultra-low noise preamplifier with integrated high-power pulse protection, active
and passive decoupling circuitry is implemented. The size of this preamplifier
circuit is 16mm x 19mm x 5mm. It is based on 3-element
matching and decoupling network 2, 3, which, matches the coil to the optimal noise
impedance of the active device4, Zn,opt,
and, at the same time, provides maximum available impedance 2 Zo at the input terminals of the coil realizing
preamplifier decoupling. The schematic representation of the single array element
is shown in Fig.1.
The coil loop is made of a flexible conducting
material and in this case is based on 1 mm outer diameter coaxial cable.
Only the outer conductor of the coaxial cable is used here. The diameter of the
loop is 8 cm, which is chosen out of imaging considerations1. To avoid interaction with 1H coils of the
scanner, a parallel resonator is integrated in the loop, as shown in
Fig. 1. The resonator consists of lumped LC components and is tuned to block
1H current at 127.8 MHz. The effect of this blocking resonator
is visible when measuring decoupling with a double-loop probe. The results of
the measurements are shown in Fig. 2.
The resulting input impedance of the sample
loaded loop coil is Zc ≈ 1 + j56 Ω
at 32.1 MHz, which is a resonance frequency of 13C nuclei in 3T
GE scanners. This impedance together with parameters of the amplifier, Zn,opt, Za (amplifier input impedance) are used in the design of
the matching and decoupling network. The design is based on equations derived
by Wang2 , which offer four alternative solutions. The
equations can be used to design high impedance Zo as well as low impedance preamplifiers. Out of four
possible solutions, the one shown in Fig. 1 is chosen here for practical
reasons. The circuit is a T-network and consists of a 560 pF series
capacitor, 112 pF parallel capacitor, and 2.7 uH series inductor (ceramic
core 1812CS from Coilcraft). The total noise figure of the preamplifier
including this matching and decoupling network is ≈ 0.6 dB. The
resulting impedance of the preamplifier at the terminals of the coil is shown
in Fig. 3.
As can be seen, the preamplifier is noise matched when
directly connected to the coil and, at the same time, presents high impedance
to the input terminals of the loop. The measured passive decoupling is shown in
Fig. 2.
The circuit provides a local minimum in the double-loop
probe measurements at the 13C resonance frequency. This indicates
that influence of the coil on impedance of other elements in the array will be
minimized.
Results
The designed coil is used as an element in an
8-channel array. The photograph of the array with cover removed is shown in
Fig. 4(a).
The preamp circuit size is a small fraction of the array,
which supports the principle of flexibility. The circuit is connected to the
scanner through a coaxial line and power supply wires.
To minimize noise coupling between the channels, the coils
are critically overlapped. The critical overlapping is found empirically by
measuring SNR as a function of distance between the coils. As one can see from
the measurement results in Fig. 4(b), the critical overlapping appears close
to 2 cm (25% of coil diameter), which is what typically expected from
these types of loop coils.
The array is easy to handle and is compatible with various
parts of the human body. To test imaging properties of the array it was applied
to a human head phantom1. The photograph of the imaging setup in the
scanner is shown in Fig. 4(c). The results of imaging experiments are shown
in Fig.5.Discussion and Conclusion
The
array shows a reasonable coverage of the object to image. As expected, a higher
sensitivity is achieved at the surface, and lower sensitivity further in the
depth, which follows the sensitivity profile of a loop coil. The array is
flexible, lightweight and easy to handle due to flexibility of the implemented
wires and compact preamplifier design with integrated noise matching as well as
passive and active decoupling circuits. The design is easily scalable and is
potentially useful for realizing large arrays preserving flexibility,
sensitivity and patient conformity.Acknowledgements
The authors would like to thank Danish National
Research Foundation (grant DNRF124) for partial support of the activities.References
1. Juan Diego Sanchez-Heredia, Wenjun Wang, Rie B. Olin, Vitaliy Zhurbenko, and Jan Henrik Ardenkjær-Larsen, Enhanced
Low Frequency MRI using Flexible Shape Arrays Made of Standard Wire,
Proceedings of 14th European Conference on Antennas and Propagation. IEEE, 2020, 4 p.
2. Wenjun
Wang, Vitaliy Zhurbenko, Juan Diego Sanchez Heredia, Jan Henrik
Ardenkjær-Larsen. Three-element matching networks for receive-only MRI coil
decoupling. Magn. Reson. in Med., Volume85, Issue1, January 2021, Pages
544-550. (published online: 19 July 2020)
3. Wenjun
Wang, Vitaliy Zhurbenko, Juan Diego Sanchez Heredia, Jan Henrik
Ardenkjær-Larsen. Matching and decoupling networks for receive-only MRI arrays.
Proceedings of 14th European Conference on Antennas and Propagation.
IEEE, 2020, 4 p.
4. LNA
for MRI, elcry1-u datasheet. Available online: elcry.com.