Johannes Petzold1, Bernd Ittermann1, and Frank Seifert1
1Physikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany
Synopsis
This
simulation study compares two concepts ensuring IEC-compliant SAR values for
parallel transmission (pTx) in respect to their robustness against subject- and
position-changes: 1) direct SAR assessment (SL) and 2) an amplitude limit for
all pTx channels (AL). For the example case of a 3T pTx body coil, SL resulted
in ten-fold the allowed SAR in one model for the same excitation vector that obeys
IEC limits in another model. For AL, less than twice the SAR was found in this
case. AL allows for a lower safety factor, therefore, resulting in a higher mean(B1+) compared to SL.
Introduction
Implant
safety revitalized the interest in parallel transmission (pTx) at lower field
strengths.1 Simulation studies at 7T show that
a slight shift in position or rotation of the head can lead to a tripled local specific
absorption rate (SAR) compared to the position used for radio frequency (RF)
optimization.2 In this simulation study, two safety
concepts for a 3T pTx body coil are compared with respect to the robustness of
their results, if the patient’s position, posture, or body mass index (BMI)
deviate (slightly) from the simulated scenario during the safety assessment.Methods
Electric
fields were extracted from electromagnetic FDTD simulations of an 8-channel 3T
(128 MHz) body coil with voxel models Duke (three
y-positions)
and Ella (five BMI/posture variants)
3 in Sim4Life
4. Ten-gram averaged $$$Q$$$-matrices
5 were calculated from the point SAR
matrixes of the 2mm isotropic voxel grid. Additionally, one whole-head and one
whole-body $$$Q$$$-matrix were generated. Normalized
matrices were created by dividing each averaged
$$$Q$$$-matrix by its corresponding IEC normal-mode SAR limit
6.
Safety Concepts
Two pTx safety
concepts were compared with respect to their ability to interpolate between
positions, extrapolate to new positions, robustness against model changes and
conservativeness: direct SAR-Limit (SL) and Amplitude-Limit (AL). SL is the
standard approach where all excitation vectors are permitted that satisfy the IEC limits. AL
7 permits all excitation vectors
$$$u$$$ with components
$$$u_k$$$ not exceeding a universal amplitude limit
$$$\alpha$$$.
AL is more restrictive than SL, therefore, as excitations with uneven element
voltages are penalized.
SAR-limiting
criteria during the safety assessment (“anchor simulations“) were constructed
as follows:
- Amplitude Limit:
An upper bound for the normalized
$$$\widehat{\text{SAR}}$$$ associated with each
$$$\hat{Q}$$$ was derived with Hölder’s inequality8:$$\widehat{\text{SAR}}=u^\dagger\hat{Q}u\leq(\max_k|u_k|)^2\sum_{ij}|\hat{Q}_{ij}|.$$By defining$$\alpha\equiv\min_l\sqrt{\frac{1}{\sum_{ij}|\hat{Q}_{ij}^{(l)}|}},$$where the minimum runs over all
$$$\hat{Q}$$$-matrices,
$$$\widehat{\text{SAR}}\leq1$$$ becomes equivalent to
$$$\max_k|u_k|\leq\alpha$$$.
- Direct SAR Limit:
An excitation vector
$$$u$$$ is safe, if
$$$\max_lu^\dagger\hat{Q}_lu\leq1$$$.
For the
subsequent “target simulations” with slightly different patient position,
posture, etc., the anchor-simulation-limits were fully exploited.
Worst-case analysis- Direct SAR Limit:
Virtual observation points (VOPs)9 with SAR overshoot
$$$\varepsilon=1\%$$$ were calculated for each simulation using all $$$\hat{Q}$$$.
A Nelder-Mead optimization was carried out with
10000 random initial vectors and these VOPs to find the 10 excitation vectors with
the highest target simulation VOP-SAR. In a second step, these 10 vectors
$$$u_s$$$ were scaled to satisfy
$$$\max_au_s^\dagger\hat{Q}_au_s=1$$$ for all anchor
$$$\hat{Q}_a$$$-matrices
to eliminate VOP-compression artefacts in
$$$\widehat{\widetilde{SAR}}=\max_{u_s,\hat{Q}_t}u_s^\dagger\hat{Q}_tu_s$$$
of all target
$$$\hat{Q}_t$$$.
- Amplitude Limit:
The target simulation’s (amplitude limit
$$$\alpha_t$$$)
theoretical maximum normalized
SAR for anchor
simulation limit $$$\alpha_a$$$
is$$\widehat{\widetilde{SAR}}=(\frac{\alpha_t}{\alpha_a})^2.$$
Sweep analysisThe target
simulation’s maximum normalized SAR is calculated for 10000 random excitation
vectors scaled to satisfy the anchor simulation’s safety limit.
Results
Unknown patient model
All
simulations were grouped by the z-distance between coil center and model’s heart
(Figs. 1A and 2C). For model-family Ella, simulations were only available in
the gray interval. The maximum amplitude limit
$$$\alpha$$$ of all simulations in a group deviates by less
than ~20% (Fig. 2A). Worst case analyses in each group for i) all combinations
of one anchor and one target ("$$$i\,\rightarrow\,j$$$")
and ii) all combinations of one target and all remaining simulations as anchor
("$$$(\text{all}-j)\,\rightarrow\,j$$$")
(Figs. 1B and 2B) show up to 10 times the allowed SAR values for (SL,$$$\,i\,\rightarrow\,j$$$) and up to 2 times for (AL,$$$\,i\,\rightarrow\,j$$$).
In both cases, the
$$$(\text{all}-j)\,\rightarrow\,j$$$ approach is safer with SAR-factors
$$$<3$$$ (SL) and
$$$<1.5$$$ (AL).
Mean($$$B_1^+$$$) in the image plane was optimized by a Nelder-Mead
algorithm for both concepts and all simulations and compared to the CP-mode
value (Fig. 2D). While SL outperforms the CP mode
($$$\text{mean}(B_{1,\text{SL}}^+)/\text{mean}(B_{1,\text{CP}}^+)\approx\,1.1$$$)
,
AL falls behind
($$$\text{mean}(B_{1,\text{AL}}^+)/\text{mean}(B_{1,\text{CP}}^+)\approx\,0.75$$$).
Applying a safety
factor
$$$\sqrt{\frac{1}{\widehat{\widetilde{\text{SAR}}}}}$$$,
however, obtained from
the maximum of the respective ($$$\,i\,\rightarrow\,j$$$)
cases of both concepts
(Fig. 2B) reverses the picture with
$$$\sqrt{\frac{1}{\widehat{\widetilde{\text{SAR}}}_{\text{SL}}}}\frac{\text{mean}(B_{1,\text{SL}}^+)}{\text{mean}(B_{1,\text{CP}}^+)}\approx\,0.4$$$,
now, while
$$$\sqrt{\frac{1}{\widehat{\widetilde{\text{SAR}}}_{\text{AL}}}}\frac{\text{mean}(B_{1,\text{AL}}^+)}{\text{mean}(B_{1,\text{CP}}^+)}\approx\,0.7$$$,
see Fig. 2E. Note that no safety factors were applied to CP, here, as it serves as a fixed reference.
Assumption of correct patient model
Three
simulations of Duke with steps of $$$\Delta$$$z=50mm were compared (Figs. 3 and 4). Using
the outer two as anchors to infer the target in the middle (interpolation) with SL resulted in a
local normalized-SAR maximum of 1.1 in the target (Fig. 5A). Extrapolation from one anchor to the neighboring
target resulted in a
maximum normalized SAR of 1.8 (Fig. 5B), while the theoretical maximum for
AL was 1.25 (Fig. 4B).Discussion
SL with positional
interpolation is SAR-wise safe while offering the best performance if the
correct patient’s model was used during safety assessment.
A possible inter-patient
SAR overshoot of up to a factor of 10, however, necessitates
a high safety factor for SL as it is not feasible to digitize patients on the fly and
only
~40 human voxel models10 are currently available as anchors.
While AL’s
$$$\text{mean}(B_1^+)$$$ is
~30% lower compared to SL for a single model, the AL’s
inter-patient SAR overshoot is much lower (<2) due
to its phase-invariance. AL is therefore outperforming SL by
~30% in
$$$\text{mean}(B_1^+)$$$ when the safety margin is factored in.Conclusion
Due to
its better robustness against model-variations, AL requires lower safety
margins and therefore provides higher $$$\text{mean}(B_1^+)$$$ than SL.Acknowledgements
This project
(17IND01 MIMAS) has received funding from the EMPIR programme co-financed by
the Participating States and from the European Union's Horizon 2020 research
and innovation programme.References
- Winter L, Seifert F, Zilberti L, Murbach M, Ittermann B.
MRI‐Related Heating of Implants and Devices: A Review. J. Magn. Reson. Imaging 2020:jmri.27194 doi:
10.1002/jmri.27194.
- Kopanoglu E, Deniz CM, Erturk MA, Wise RG. Specific absorption rate implications of within-scan patient head
motion for ultra-high field MRI. Magn. Reson. Med. 2020;84:2724–2738 doi: 10.1002/mrm.28276.
- Gosselin M-C, Neufeld E, Moser H, et al. Development
of a new generation of high-resolution anatomical models for medical device
evaluation: the Virtual Population 3.0. Phys. Med. Biol. 2014;59:5287–5303 doi:
10.1088/0031-9155/59/18/5287.
- Sim4Life 5.2, ZMT Zurich MedTech, Zürich,
Switzerland
- Bardati F, Borrani A, Gerardino A,
Lovisolo GA. SAR optimization in a phased array radiofrequency hyperthermia
system. IEEE Trans. Biomed. Eng. 1995;42:1201–1207 doi: 10.1109/10.476127.
- IEC 60601-2-33 Ed. 3.2: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnosis. International Electrotechnical Commission, 2015.
- Seifert F, Cassara A, Weidemann G,
Ittermann B. Reliable and robust RF safety assessment of transmit array coils
at ultrahigh fields. In: Proc. Intl. Soc. Mag. Reson. Med. 22. ; 2014. p. 4891.
- Hölder O. Ueber einen Mittelwerthssatz. Nachrichten Von Königl Ges.
Wiss. Georg-Augusts-Univ. Zu Gött. 1889:38–47.
- Eichfelder G, Gebhardt M. Local
specific absorption rate control for parallel transmission by virtual
observation points. Magn. Reson. Med. 2011;66:1468–1476 doi: 10.1002/mrm.22927.
- Makarov SN, Noetscher GM, Yanamadala
J, et al. Virtual Human Models for Electromagnetic Studies and Their
Applications. IEEE Rev. Biomed. Eng. 2017;10:95–121 doi:
10.1109/RBME.2017.2722420.