Matthew Tarasek1, Tom Foo1, Mark Vermilyea1, Desmond Yeo1, Isabelle Jansen1, Eric Budesheim 1, and Keith Park1
1GE Global Research, Niskayuna, NY, United States
Synopsis
In this work we evaluate the MRI acoustic changes brought
about by modifying the RF shield structure in a head-only (MAGNUS) 3T imaging
system. Specifically, we use a standardized acoustic measurement protocol to compare
sound pressure levels (SPL) in a fiberglass patient-bore structure to a similar
newly developed carbon fiber reinforced polymer (CFRP) structure. Results
indicate that a modification of the structural properties of the patient-bore
structure properties can provide SPL reduction for MR imaging applications, and
is likely due to the different modulus of elasticity between the two materials.
Purpose
Sound pressure waves created from the gradient coil moving in
a main magnetic field create audible acoustic noise in MRI, which can pose a
significant safety risk to the patient and operator. Previous work has focused
on sound pressure level (SPL) reduction through various software [1-5] and
hardware [6-8] approaches. Here we present data suggesting that the patient-bore
structure, a standard component of most clinical MRI scanners, may be useful as
an acoustic barrier for acoustic noise reduction. We provide an acoustic
evaluation and comparison between a standard fiberglass reinforced polymer
(FRP) patient-bore structure and a newly developed carbon fiber reinforced
polymer (CFRP) weave patient-bore structure.Methods
The patient-bore stepped cylindrical structure can be seen in
Figure 1 (Left). It has a stepped profile that accommodates the 42-diameter
gradient bore, and flairs to meet the 90-cm warm bore of the 3T magnet. As a
requirement to eliminate interaction between the transmit/receive RF coil and
the gradient coils, a stainless-steel mesh was mounted on the outside of the
patient-bore structure to serve as the system RF shield when the structure was
constructed from FRP. With the patient-bore structure constructed from CFRP,
the stainless-steel mesh was not necessary and was not used. This was because
CFRP has high electrical conductivity at high-frequency and can thus serve as an
RF shield and also provide structural strength to the patient-bore structure.
However, the focus of this work was in reduction of acoustic noise.
Acoustic measurements were made using (i) the
standard FRP patient-bore structure and (ii) the CFRP structure of the same dimensions as for the
FRP version. The acoustic measurement configuration inside the scanner bore is visualized
in Figure 1 (Right) and was repeated for both test cases. Multiple scan
sequences representing clinically relevant imaging protocols were tested,
following the maximum clinical acoustic noise (MCAN) protocol defined by [9].
The gradient waveforms were measured using an oscilloscope to ensure identical
duty cycles, gradient amplitudes, and slew rates for the corresponding
sequences in both test arrangements. Acoustic measurements were made using a
Model G4, Type 2250 handheld analyzer and Type 4189 microphone (Brüel &
Kjær, Nærum, Denmark). The microphone was placed at the imaging isocenter,
perpendicular to the patient axis, 80mm above the patient table. The
A-weighted, equivalent continuous sound level was recorded in decibels (LAeq,
dBA), along with the unweighted peak sound level (Lpeak, dB). SPL
test results for 11 clinical sequences or MRI pulse sequence variations are
summarized in Figures 2 and 3.Results and Discussion
The
CFRP patient-bore structure provided an average of 3.6 dBA, and a maximum 8 dBA
reduction in LAeq relative to the FRP patient-bore structure.
Similar results were seen for peak sound SPL measurements, Lpeak ,
with an average of 2.3 dB, and a maximum
7 dB reduction (Figures 2 and 3). The variation of the measured SPL between
different pulse sequences is expected since different pulse sequences have
different waveform frequency content. We note that every sequence was played
identically for each test case, so there was no difference in waveform
frequency content for the same pulse sequence in different test
arrangements. The results indicate that the use of CFRP for the patient-bore
structure is a viable option for MRI scanner hardware noise derating. The
reduction in SPL was likely due to the different modulus of elasticity between
the materials (240 GPa for CFRP versus 90 GPa for FRP). The findings indicate
that material selection for the patient-bore structure and any other structures
within the magnet warm bore is an important consideration in SPL management for
high-performance MRI scanners, especially at higher field strengths and/or even
higher performance gradient coils [10].Acknowledgements
Eric Fiveland and Charles Seeley at GE Global Research.References
[1] Idiyatullin D, Corum C, Park J-Y,
Garwood M. Fast and quiet MRI using a swept radiofrequency. J Magn Reson. 2006;
181(2):342–349. [PubMed: 16782371]
[2] Hennel F. Fast spin echo and fast gradient
echo MRI with low acoustic noise. J Magn Reson Imaging. 2001; 13(6):960–6.
[PubMed: 11382960]
[3] Segbers M, Rizzo Sierra CV, Duifhuis H,
Hoogduin JM. Shaping and timing gradient pulses to reduce MRI acoustic noise.
Magn Reson Med. 2010; 64(2):546–53. [PubMed: 20665798]
[4] Heismann B, Ott M, Grodzki D.
Sequence-based acoustic noise reduction of clinical MRI scans. Magn Reson Med.
2014; 73(3):1104–9. [PubMed: 24889327]
[5] Tan ET, Hardy CJ, Shu Y, In MH, Guidon A,
Huston J 3rd, Bernstein MA, Foo TKF. Reduced acoustic noise in diffusion tensor
imaging on a compact MRI system. Magn Reson Med 2018; 79: 2902-11. [PMID:
28971512]
[6] Mansfield P, Chapman BL, Bowtell R, Glover
P, Coxon R, Harvey PR. Active acoustic screening: reduction of noise in
gradient coils by Lorentz force balancing. Magn Reson Med. 1995; 33(2):276–281.
[PubMed: 7707921]
[7] Bowtell RW, Mansfield P. Quiet transverse
gradient coils: Lorentz force balanced designs using geometrical similitude.
Magn Reson Med. 1995; 34(3):494–497. [PubMed: 7500892]
[8] Edelstein WA, Hedeen RA, Mallozzi RP,
El-Hamamsy SA, Ackermann RA, Havens TJ. Making MRI quieter. Magn Reson Imaging.
2002 Feb; 20(2):155–63. [PubMed: 12034336]
[9] ACOUSTIC NOISE MEASUREMENT PROCEDURE FOR
DIAGNOSTIC MAGNETIC RESONANCE IMAGING DEVICES. NEMA-MS 4. National Electrical
Manufacturers Association. 2010
[10] Foo TKF, Tan ET, Vermilyea ME, et al.
Highly efficient head-only magnetic field insert gradient coil for achieving
simultaneous high gradient amplitude and slew rate at 3.0T (MAGNUS) for brain
microstructure imaging. Magn Reson Med. 2020;83:2356–2369.