Chen Lin1, LeRoy H Stecker1, Brittany L Benson1, Craig A Hildestad2, Shengzhen Tao1, and Robert A Pooley1
1Radiology, Mayo Clinic, Jacksonville, FL, United States, 2Mayo Clinic, Jacksonville, FL, United States
Synopsis
A novel
acoustic noise shield constructed of clear acrylic has been developed and
evaluated. The results suggest that: 1) it is effective for MR acoustic noise
reduction, 2) PET image quality exceeds ACR accreditation testing standards
with the proposed acoustic shield in use, showing its compatibility with PET
imaging, 3) there is a small impact on PET SUV which can be compensated with
attenuation correction, and 4) it does not cause claustrophobia.
INTRODUCTION
High levels
of acoustic noise during MRI are a major source of discomfort as well as a risk
for injury. Therefore, adequate hearing protection is required for MR exams
including PET-MR. While there are many ways to reduce acoustic noise exposure
during MRI, each has its own limitations. Foe example, large headphones cannot be used
with smaller head coils and earplugs do not fit well in some patients, derating
the attenuation by up to 50%. Placing acoustic damping material between the
head coil and the scanner bore has been shown to be an effective option for noise
reduction in pediatric MRI [1]. However, such acoustic hood was not intended
for PET-MR and not suitable for claustrophobic and adult patients. We have
therefore designed and evaluated a new type of acoustic noise shield
constructed with clear acrylic and optimized for PET-MR and for
an improved patient experience.METHODS
Clear
acrylic was chosen as the material for this PET-MR acoustic shield because of
its optical transparency, relatively high density of 1.18g/cm3 for good
acoustic attenuation, relatively low mass attenuation coefficient of 9.324x10-2cm2
/g for 511keV photons, good mechanical properties for fabrication and commercial
availability in the desired size and shape.
As
shown in figure 1, the PET-MR acoustic shield consists of ¼ inch thick acrylic
arc of 20.4inch long and 20inch diameter, with a ¼ inch thick back panel joined
into one rigid piece. It weighs 4.73kg. The shield rests on the scanner table when
placed over an RF coil, and it is compatible with different types of head and
neck coils including GE 8ch HR Brain Array, Nova Medical 32ch Head
Coil and GE 19ch Head and Neck Unit.
The
evaluation of MR acoustic noise reduction and PET signal attenuation with this
acoustic shield was performed on a clinical PET-MR scanner (SIGNA PET/MR, GE
Healthcare, Wauwastosa, WI) which has a 60 cm bore and 44mT/m peak amplitude
and 200 T/m/s peak slew rate gradients. Equivalent continuous sound pressure level
(LAeq) during MR scanning and the ambient background noise were measured with
and without the acoustic noise shield using a sound level meter (Model 2250L, Bruel&Kjaer, Nærum, Dennmark) and an MR safe microphone. The microphone was positioned
adjacent to a spherical phantom or a volunteer’s ear inside an 8ch head coil as shown in
figure 2. Continuous measurement of sound pressure level (SPL) in the initial 30 seconds after
pre-scan was recorded and averaged for frequently used clinical MR sequences
listed in figure 3.
PET
images of an ACR PET phantom were also acquired with and without the acoustic
shield to evaluate its effect on PET image quality. The acquisition, reconstruction
and analysis were performed according to the ACR PET accreditation program [2]. RESULTS
The
measured SPL for clinical MR sequences with and without the shield
is summarized in figure 3. There is minimal difference in background noise with
and without the shield. In volunteer testing, the difference in SPL with and without shield varied from +1.2dB to -17.2dB depending on the pulse
sequence. For a hypothetical MR exam protocol that includes all sequences in figure
3, the reduction in time-averaged SPL with the shield are 10.2dB and 11.5dB for two volunteers respectively.
The
PET images acquired with and without the shield both exceeded ACR PET image
quality criteria. The measured SUV of 4 different cylinders as well as the
background in the ACR PET phantom with and without the acoustic shield are
listed in figure 4. There is a 6.3–15.4% decrease in SUV without correcting for
PET attenuation due to the acoustic shield.
The
acoustic shield has been used in clinical MR and PET-MR exams of more than 150
patients. There has been no report of claustrophobia from volunteers and
patients due to the use of this acoustic shield.DISCUSSION
The
reduction of acoustic noise with the shield varies with MR sequence. This is
presumably due to the difference in frequency spectrum of the gradient sound
produced by different pulse sequences. There is a slight difference in noise
reduction measured from two volunteers. This is likely due to the difference in
body habitus as the body also acts as a sound barrier partially blocking the
open end of the acoustic shield.
The
PET images acquired with the proposed acoustic shield in place exceed all image
quality criteria for ACR PET accreditation, demonstrating that the proposed
shield design is compatible with simultaneous PET imaging, despite an observed
SUV reduction of 6.5%. In theory, a correction can be made in PET image
reconstruction to compensate for the attenuation by acoustic shield. However,
this is currently not implemented due to the lack of flexibility in software. A
reduction of PET sensitivity can be compensated by slightly longer PET acquisition
time to allow the same number of counts. Such adjustment should not be
difficult as PET-MR bedtime is usually limited by MR acquisition time.
The
acoustic shield did not cause any claustrophobia in our test and evaluation.
This is because the thin clear acrylic material is almost completely
transparent. It did not make volunteers and patients feel like they were being further
confined in a small space.Acknowledgements
No acknowledgement found.References
- Nordell
A, et al. The acoustic hood: a patient-independent device improving acoustic
noise protection during neonatal magnetic resonance imaging. Acta Paediatr.
2009 Aug;98(8):1278-83
- ACR
Nuclear Medicine and PET Accreditation Phantom Testing (https://accreditationsupport.acr.org/support/solutions/articles/11000062800-phantom-testing-pet)