Valerie Klein1,2, Mathias Davids1,2,3, Donald Straney2, Livia Vendramini2, Lothar R. Schad1, Maaike van den Boomen2,3,4, Christopher Nguyen2,3,4, Lawrence L. Wald2,3,5, and Bastien Guerin2,3
1Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany, 2A. A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, MA, United States, 3Harvard Medical School, Boston, MA, United States, 4Cardiovascular Research Center, Cardiology Division, Massachusetts General Hospital, Charlestown, MA, United States, 5Harvard-MIT Division of Health Sciences and Technology, Cambridge, MA, United States
Synopsis
We developed a magnetic stimulator to measure cardiac stimulation (CS)
thresholds of time-varying magnetic fields in pigs. The stimulator consists of
a high-power capacitor bank (C=110 µF) discharging into a pancake coil. We will
use the measured thresholds to validate a previously developed CS simulation
framework that combines electromagnetic field simulations with
electrophysiological models of excitable cardiac fibers. We generate whole-body
porcine voxel models from multi-bed MRI Dixon and CINE acquisitions to
reproduce the anatomy and posture of the animals in our simulations. Carefully
validated CS simulations may eventually be useful in setting cardiac safety
limits for MRI gradient applications.
Purpose
Time-varying magnetic fields such as MRI
gradient fields induce E-fields in the body that can potentially stimulate the
heart[1,2]. We have recently developed a simulation framework for the prediction of
magnetic cardiac stimulation (CS) thresholds[3-5]. The framework combines E-field simulations in detailed body models with
electrophysiological modeling of excitable cardiac fibers. Initial simulations
successfully replicated CS thresholds of previous canine experiments performed
in 1992[2,6], but a more detailed validation is needed whereby the exact experimental
setup (anatomy, posture, coil position, etc.) is correctly represented in our
model[3,4]. Here, we report on our ongoing effort to validate our CS simulations in
pigs, the primary animal model for the human cardiovascular system[7]. Methods
Magnetic stimulator: Figure 1A shows the
magnetic stimulator, which is essentially a large capacitor bank discharging
into a coil. The capacitors (Umax=18 kV, C=110 µF, HVP, Martinsried,
Munich, Germany) are charged by a TDK-Lambda (Tokyo, Japan) FLX-HV power supply.
Once the high-power relay (Ross Engineering, Campbell, CA) is closed, the
charge disperses into the inductive coil at a resonance
frequency of ~1 kHz, and with a decay rate determined by the coil’s AC
resistance (~200
mΩ,
Fig. 1B). The coil is a flat pancake
design (Dout=40 cm, Din=2.5 cm, 38 turns, Fig. 1C) made of 2.3-mm copper
wire covered with epoxy and has an inductance of L≈230 µH.
Because
of the high current/voltages, safety circuits in the pulse generator use
additional relays and high-energy resistors for emergency discharge of the
capacitors. As an additional precaution, the system is operated by a remote
control. The discharge is triggered by the combination of a manual button press
(arming the system) and an optical pulse from a laptop monitoring ECG (triggering during
diastole). The coil is connected to the pulse generator with a 40
kV-rated 8AWG cable, and a set of GES (Munich, Germany) SB135/HB135 connectors
and attached
below the table at heart center (Fig. 1C). The amplitude of the discharge
pulses will be iteratively increased until the CS threshold is reached, which
will be identified and recorded as an ectopic heartbeat on the ECG and SpO2
trace.
CS
threshold simulations: This study is conducted with approval
and under advice of the MGH Institutional Animal Care and Use Committee. The
pigs were pre-anesthetized with Telazol and Atropine, intubated, ventilated with
O2 (mixed with 1-3% isoflurane), maintained under anesthesia with isoflurane (1-3%),
and placed in left lateral recumbency. We acquired fat-water separated Dixon volumes (500x325x344mm3, 2mm isotropic resolution, TR/TE1/TE2=5/1.23/2.46ms) as
well as CINE acquisitions (1.3x1.3x2.5mm3, 41 slices, 18 frames, TE/TR=4.9/35.4ms, double
oblique along the heart short axis) of the beating heart on a 3T
Prisma scanner (Siemens Healthineers, Erlangen, Germany) for seven pigs. Dixon volumes were
acquired in 5 bed positions separated by 200 mm and stitched together in
postprocessing. We created porcine body models by assigning electrical
conductivity values[8]
based on the local fat fraction, ranging from
pure fat (0.057 S/m) to pure muscle (0.355 S/m). We segmented the lungs (σ=0.11
S/m), assigned the outermost voxel layer as skin (σ=0.17 S/m), and set all
voxels with low signal intensity as low-conductive tissue (σ=3.5 mS/m) (Fig.
2A). The heart was manually segmented from the CINE data (diastole frame, Fig.
2B), and cardiac Purkinje and
ventricular muscle fibers were added using rule-based modeling
algorithms[9,10]. We
then simulated E-fields using Sim4Life’s low-frequency solver (Zurich MedTech,
Switzerland), and projected the result onto the fibers and integrated to obtain
electrical potentials. Finally, we
modulated the potential along the fibers over time by the B-field waveform’s
temporal rate of change (dB/dt) created by the capacitor discharge and evaluated
the electrophysiological fiber models to predict CS threshold values.Results
We compared different coil designs for the E-field
induced in the porcine heart and found the flat pancake coil design to be the
most effective (Fig. 3).
Figure 4A shows a
comparison between the measured and simulated B-field profiles of this coil. Figure
4B shows the measured and simulated voltage, current, and dB/dt waveforms resulting
from the capacitor discharge at U=9 kV (Bpeak≈2.0
T, dB/dtpeak≈17500 T/s at coil center). This corresponds to Bpeak≈3.9
T and dB/dtpeak≈35000 T/s at peak charging voltage (Umax=18
kV).
Figure
5 compares CS thresholds simulated in the porcine models with the hardware
limits of the stimulator. The average threshold 95th percentile
E-field magnitude in the heart of the seven porcine models is E95,thresh=232±17
V/m (compared to E95,peak=338±22 V/m at hardware limit). The average
simulation threshold B-field amplitude at coil center is Bthresh=3.4±0.3
T (dB/dtthresh=24200±2300 T/s), which is within the stimulator’s
performance limits. Increasing the capacitance to C=220 µF would lower the
resonance frequency and decrease the (simulated) thresholds to significantly
below the hardware limits (Bthresh=3.3±0.3 T, dB/dtthresh=17500±1700
T/s).Discussion
CS requires very high magnetic field energies[11], and we previously
found that a typical TMS stimulator or MRI gradient would not be powerful
enough to guarantee CS[12]. We therefore
developed a high-power magnetic stimulator similar to that used in the canine
experiments of Mouchawar et al.[2] to validate our CS simulations in pigs. These
experiments will ultimately allow us to validate our CS simulations, which
could eventually become useful in setting appropriate cardiac safety limits for
MRI applications.Acknowledgements
This
study is in part supported by the ISMRM Research Exchange Grant and the German
Academic Exchange Service (DAAD) as well as NIH award number R01 EB028250.References
[1] Reilly J P,
Principles of nerve and heart excitation by time-varying magnetic fields. Ann.
N. Y. Acad. Sci., 1992. 649: 96-117.
[2] Mouchawar G A, Bourland J D,
Nyenhuis J A, Geddes L A, Foster K S, Jones J T, and Graber G P, Closed-chest
cardiac stimulation with a pulsed magnetic field. Med. Biol. Eng. Comput.,
1992. 30: 162-168.
[3] Klein V, Davids M, Schad L R, Wald L
L, and Guérin B, Investigating cardiac stimulation limits of MRI gradient coils
using electromagnetic and electrophysiological simulations in human and canine
body models. Magn. Reson. Med., 2021. 85: 1047–1061.
[4] Klein V, Davids M, Schad L R, Wald L
L, and Guérin B. Simulation of electromagnetic cardiac stimulation: Validation
in dogs and application to human threshold limits for MRI gradient coils. In: Proceedings of the 28th Annual Meeting of
the ISMRM. 2020. p. 1125.
[5] Klein V, Davids M, Schad L R, Wald L
L, and Guérin B. Simulation of MRI gradient-induced cardiac stimulation: Are
the current safety regulations too conservative? In: Proceedings of the 27th Annual Meeting of the ISMRM. 2019.
Montréal, Canada; p. 729.
[6] Nyenhuis J A, Bourland J D, Schaefer
D J, Foster K S, Schoelein W E, Mouchawar G A, Elabbady T Z, Geddes L A, and
Riehl M E. Measurement of cardiac stimulation thresholds for pulsed z-gradient
fields in a 1.5-T magnet. In: Proceedings
of the 11th Annual Meeting of the SMRM. 1992. Berlin, Germany; p. 586.
[7] Dixon J A and Spinale F G, Large
animal models of heart failure. Circ., 2009. 2(3): 262-271.
[8] Hasgall P A, Di Gennaro F,
Baumgartner C, Gosselin M C, Payne D, Klingenböck A, and Kuster N, IT’IS
Database for thermal and electromagnetic parameters of biological tissues
Version 4.0. 2018.
[9] Ijiri T, Ashihara T, Yamaguchi T,
Takayama K, Igarashi T, Shimada T, Namba T, Haraguchi R, and Nakazawa K, A
procedural method for modeling the Purkinje fibers of the heart. J. Physiol. Sci., 2008. 58(7): 481-486.
[10] Bayer J D, Blake R C, Plank G, and
Trayanova N A, A novel rule-based algorithm for assigning myocardial fiber
orientation to computational heart models.
Ann. Biomed. Eng., 2012. 40(10):
2243-2254.
[11] Schaefer D J, Bourland J D, and
Nyenhuis J A, Review of patient safety in time-varying gradient fields. J.
Magn. Reson. Imaging, 2000. 12: 20-29.
[12] Klein V, Davids M,
Nguyen C, Schad L R, Wald L L, and Guérin B. Feasibility of using transcranial
magnetic stimulation devices to study magnetically induced cardiac stimulation
in pigs. In: Proceedings of the 28th
Annual Meeting of the ISMRM. 2020. p. 111.