Jonathan Weine1, Robbert J. H. van Gorkum1, Christian T. Stoeck1, Valery Vishnevskiy1, Thomas Joyce1, and Sebastian Kozerke1
1Institute for Biomedical Engineering, University and ETH Zurich, Zürich, Switzerland
Synopsis
Cardiac DTI provides invaluable information about the state
of myocardial microstructure. Motion and systematic signal variations of the
imaging process influence the tensor inference. Image registration prior to tensor
fitting with an LSQ estimator is the common data processing approach. The
feasibility of training a neural network with simulated data modelling tensors
and slice misalignment due to free breathing for inference of diffusion tensors
from free-breathing in vivo data is investigated. Evaluation on simulated test
data demonstrates feasibility of the training process. Application to in vivo
data shows promising results of the CNN especially at myocardial borders.
Introduction
Cardiac diffusion tensor imaging (cDTI) has shown to provide
invaluable information about the state of the myocardial microstructure in both
healthy and diseased conditions1-5. Although it is a promising
technique, challenges with respect to motion sensitivity remain to be
addressed. Cardiac motion-induced signal loss during data acquisition in spin-echo
sequences can be alleviated by using second order motion-compensated diffusion gradient
waveforms6-8. Patient-friendly free-breathing acquisition strategies
often result in spatial misalignment of the data, which require the use of non-rigid
registration prior to parameter inference. Since image contrast varies between
diffusion-weighted images (DWI), registration can be challenging. To address
this, the utilization of neural networks (NN) for registering DWIs is actively investigated9-11.
NNs have also been proposed for DTI parameter estimation on registered data to address
the ill-posed nature of such inverse problems12,13. Besides the
noise sensitivity, systematic signal variation due to residual spatial mismatch
combined with partial volume effects could cause instability in parameter
inference. As this predominantly affects the border regions of the left
ventricle (LV), a manually drawn LV mask usually excludes these areas1,5.
Instead of reducing the spatial mismatch prior to parameter
inference by image registration, thereby risking spatial averaging effects, the
varying position can be interpreted as an additional probing dimension. To exploit
this information, this work examines the feasibility of a convolutional neural network
(CNN) based end-to-end parameter inference on SE-M012 data without an explicit
registration step.Methods
A residual CNN (Fig. 1) was trained on simulated artificial
data to infer the tensor-values from data according to the 2-shell b-value scheme used by von Deuster et al.
14
Data was generated such that prior knowledge on physiological mean diffusivity (MD)
and fractional anisotropy (FA) values obtained with M012 spin-echo cDTI is
implicitly captured
15. Additionally, non-rigid motion between
diffusion weightings, obtained from in vivo data registration was incorporated
into simulation. All modules for data-generation as well as network training
and statistical analysis were implemented in Python 3.6.9 using TensorFlow 2.2.1,
NumPy 1.18, SciPy 1.4.1.
Data
generation- Rejection sample eigenvalue triples within
such that MD and FA are uniformly distributed over
the defined intervals.
- Generate tensor maps that represent
a transmurally varying helix-angle (HA) and random patches of absolute E2 sheetlet
angle (E2A) inside an in vivo LV mask.
- Randomly select a subset of
reference points inside the LV mask and scale the selected tensors with the
sampled eigenvalue triples from step 1.
- Interpolate the reference tensors in
log-Euclidean space16 with a radial basis function kernel and a
circular distance measure to obtain locally smooth tensor maps.
-
Randomly generate a lesion inside
the LV, repeat steps 1 to 4 with suitable parameters and combine the maps.
- Evaluate the imaging forward model of
the single-shot M012-SE-EPI sequence including multiple
coils and motion between each acquisition and introduce diffusion contrast by
evaluating the bilinear form according to $$$S(b_i)=S_0e^{-b_iD_{ij}b_j}$$$ with
$$$\langle b_i,b_i\rangle=b$$$.
Network
training and testing
1500
diffusion datasets including 12 b-vectors
($$$b_i$$$
) and 12 averages per diffusion
weighting were simulated using 129 LV masks. As diffusion tensor ground truth
for training, the simulation was performed without motion and noise addition
followed by inference as linear least squares problem (LLSQ) with a Trust
Region Reflective algorithm. Figure 2 illustrates a simulated example dataset. As
loss function, the sum of mean squared errors (MSE) of the inferred tensor
entries
, MD and FA was used.
To evaluate the performance of the network the mean absolute
error (MAE) for MD, FA and eigenvalues of the tensors inferred on a simulated
test dataset with unused LV-masks were calculated. As lower bound and upper
bound of error the LLSQ inferences on noisy but perfectly aligned data and
noisy unregistered data were evaluated.
As no ground truth for in vivo data was available, a
qualitative inspection of the network inference on data acquired in a previous
study was performed.
Results
The distributions of inferred
tensor metrics according to the described estimation for lower bound (LB),
upper bound (UB) and the trained CNN compared to the simulated ground truth are
shown in Fig 3. The normalized histogram show, that the CNN overall infers less
outliers.
Figure 4 shows absolute
error maps of MD and FA for a subset of simulated test examples. The animated
maps show outlier-errors at the border of the mask even for the LB estimation,
which results from the mask being larger than the extend of the simulated LV.
Therefore, an outlier rejection of absolute errors being 20 times larger than
the median (resulting of ~2% of rejected values) is applied.
This yields MAEs for MD and FA of tensors inferred
on the simulated test data of $$$(|\Delta MD|_{LB}:2.2\cdot10^{-5} \frac{mm^2}{s}, |\Delta FA|_{LB}:1.8\cdot10^{-2})$$$ for the LB estimation, $$$(|\Delta MD|_{UB}:18\cdot10^{-5} \frac{mm^2}{s}, |\Delta FA|_{UB}:10\cdot10^{-2})$$$ for the UB estimation and $$$(|\Delta MD|_{ResNet}:8\cdot10^{-5} \frac{mm^2}{s}, |\Delta FA|_{ResNet}:5\cdot10^{-2})$$$ for the CNN.
Figure 5 shows the MD values of
inferred tensors for one in-vivo dataset.Discussion
The results
demonstrate the feasibility of training a CNN on simulated data to increase
robustness of tensor inference for cDTI by implicitly including systematic and
random signal variations as a data prior. While the inference on in vivo data shows
promising results, further evaluation on data including lesions is necessary.Acknowledgements
Swiss National Science Foundation Grant PZ00P2_174144
References
1. Nielles-Vallespin S, Scott A, Ferreira P, Khalique Z,
Pennell D, Firmin D. Cardiac Diffusion: Technique and Practical Applications. J
Magn Reson Imaging. Published online 2019. doi:10.1002/jmri.26912
2. Gotschy A, von Deuster C, van Gorkum RJH, et al. Characterizing cardiac involvement
in amyloidosis using cardiovascular magnetic resonance diffusion tensor
imaging. J Cardiovasc Magn Reson. 2019;21(1):56.
doi:10.1186/s12968-019-0563-2
3. von Deuster C, Sammut E, Asner L, et al. Studying Dynamic Myofiber Aggregate
Reorientation in Dilated Cardiomyopathy Using In Vivo Magnetic Resonance
Diffusion Tensor Imaging. Circ Cardiovasc Imaging. 2016;9(10).
doi:10.1161/CIRCIMAGING.116.005018
4. Zohya K, F. FP, D. SA, et al. Diffusion Tensor Cardiovascular
Magnetic Resonance in Cardiac Amyloidosis. Circ Cardiovasc Imaging.
2020;13(5):e009901. doi:10.1161/CIRCIMAGING.119.009901
5. Das
A, Chowdhary A, Kelly C, et al. Insight Into Myocardial Microstructure of
Athletes and Hypertrophic Cardiomyopathy Patients Using Diffusion Tensor
Imaging. J Magn Reson Imaging. 2021;53(1):73-82.
doi:https://doi.org/10.1002/jmri.27257
6. Stoeck CT, von Deuster C, Genet M, Atkinson D, Kozerke
S. Second-order motion-compensated spin echo diffusion tensor imaging of the
human heart. Magn Reson Med. 2016;75(4):1669-1676. doi:10.1002/mrm.25784
7.
Welsh CL, Dibella EVR, Hsu EW. Higher-Order
Motion-Compensation for In Vivo Cardiac Diffusion Tensor Imaging in Rats.
2015;34(9):1843-1853. doi:10.1109/tmi.2015.2411571
8. Lasic S, Szczepankiewicz F, Dall’Armellina E, et al.
Motion-compensated b-tensor encoding for in vivo cardiac diffusion-weighted
imaging. NMR Biomed. 2020;33(2):e4213. doi:10.1002/nbm.4213
9. Ferreira PF, Martin RR, Scott AD, et al. Automating in
vivo cardiac diffusion tensor postprocessing with deep learning–based
segmentation. Magn Reson Med. 2020;84(5):2801-2814. doi:10.1002/mrm.28294
10. Deng Z, Wang L, Wu Q, et al. Investigation
of in Vivo Human Cardiac Diffusion Tensor Imaging Using Unsupervised Dense
Encoder-Fusion-Decoder Network. IEEE Access. Published online 2020:1.
doi:10.1109/ACCESS.2020.3040330
11. Nguyen CT, Christodoulou AG,
Coll-Font J, et al. Free-breathing diffusion tensor MRI of the whole
left ventricle using second-order motion compensation and multitasking
respiratory motion correction. Magn Reson Med. 2020;n/a(n/a).
doi:https://doi.org/10.1002/mrm.28611
12. Aliotta
E, Nourzadeh H, Sanders J, Muller D, Ennis DB. Highly accelerated, model-free
diffusion tensor MRI reconstruction using neural networks. Med Phys.
2019;46(4):1581-1591. doi:10.1002/mp.13400
13. Aliotta
E, Nourzadeh H, Patel SH. Extracting diffusion tensor fractional anisotropy and
mean diffusivity from 3-direction DWI scans using deep learning. Magn Reson
Med. 2021;85(2):845-854. doi:https://doi.org/10.1002/mrm.28470
14. von
Deuster C, Stoeck CT, Genet M, Atkinson D, Kozerke S. Spin echo versus
stimulated echo diffusion tensor imaging of the in vivo human heart. Magn
Reson Med. 2016;76(3):862-872. doi:https://doi.org/10.1002/mrm.25998
15. Das
A, Kelly C, Teh I, Stoeck CT, Kozerke S, Brown LAE, Saunderson CED, Craven T,
Swoboda PP, Levelt E, Greenwood JP, Plein S, Schneider JE, Dallarmellina. Longitudinal
changes in diffusion tensor imaging parameters following acute ST elevation
myocardial infarction. Proceedings of the
17th Annual Meeting of the European Association of Cardiovascular
Imaging. 2019; P459.
16. Arsigny
V, Fillard P, Pennec X, Ayache N. Log-Euclidean metrics for fast and simple
calculus on diffusion tensors. Magn Reson Med. 2006;56(2):411-421.
doi:10.1002/mrm.20965