Ivan Maximov1,2, Oliver Geier3, Elias Kellner4, Helle Pfeiffer3, Valerij G Kiselev4, and Marco Reisert4
1Western Norway University of Applied Sciences, Bergen, Norway, 2NORMENT, University of Oslo, Oslo, Norway, 3Oslo University Hospital, Oslo, Norway, 4University Medical Center Freiburg, Freiburg, Germany
Synopsis
Myelin water imaging (MWI) is a useful tool to probe and provide a
quantitative measure of myelin content in the human brain in vivo.
The most common MRI technique for MWI is based on multi-echo T2
measurements allowing one to estimate different T2 contributions into
the signal decay. However, the conventional non-negative least
squares algorithm is computationally very challenging and vulnerable
to image artefacts. In the present work we developed an optimised
framework for MWI enabling improved pipeline and fast metric
estimations using Bayesian regression.
Introduction
An accurate assessment of myelin content poses a challenge to in vivo neurological MRI. There are a few approaches allowing one
quantitatively or qualitatively to evaluate the amount of myelin
water in the brain, but most of them are based on indirect myelin
measures1,2. It is still a challenge to quantify myelin
directly and the developed methods are limited for in vivo
clinical applications3 due to their time consuming implementation. One of the most frequently used approaches is based
on a multi-echo T2 (MET2) measurement modelling for at least two water
pools with different T2 relaxation times, namely, myelin water
trapped between the myelin bilayers and intra- and extra-axonal
water2. A major problem in MET2 is that the fitting
algorithms used for the estimation of the model parameters are highly
vulnerable to noise and imaging artefacts4. In the present
work, we suggest an optimised MET2 framework providing an increase in
precision and accuracy of MWI assessments. This is accomplished by
using noise5 and Gibbs-ringing6 correction
steps and Bayesian estimator7 allowing one significantly
to accelerate the computations.Methods
A scheme of the
optimised framework is presented in Fig. 1 and consists of two major
parts. First data preparation and pre-processing including at least
three basic steps: noise correction based on removal of noise-only
principle components5, Gibbs-ringing correction for all
images6, and all volume normalisation, including a
Gaussian smoothing with small kernel.
Second the
estimation of the myelin water fraction by means of a Bayesian
estimation.
Other steps such as a bias field correction, frequency drift,
gradient non-linearities correction etc. can be included as well.
This is based on the
idea of modelling the signal decomposition by the assumed parameter
distribution7. The MET2 signal decay is simulated for many
parameters using the following representation:
S = v1exp(-t/TM)
+ v2exp(-t/TA)
+ v3exp(-t/TCSF),
where ∑vi=1are
the fractions of myelin water (M), intra- and extra-axonal water (A),
and cerebrospinal fluid (CSF), respectively. The simulated signal is
distorted by a non-central χ2
noise. The Bayesian estimation is defined as a maximisation of a
posteriori probability p as
x(f) = argmax
p(x|f),
where x is the parameter set of
relaxation model, and f is the signal features. The
simulations establish the training set allowing us to find an
optimised set of polynomial regressors solving the optimisation
problem. For a comparison, we used a standard fitting approach based
on non-negative least squares2,4,5.
After signed informed consent prior to the participation, we measured 32
years old healthy male volunteer. Imaging was performed on a 3T Prisma scanner (Siemens Healthcare, Erlangen, Germany). In vivo
MET2 data consist of 32 echos with the TE step equals
to 15.6 ms. Image resolution is 2 mm3.
Repetition time is 5.06
s.Results
In Fig. 2 we present
the correlation results of model training using simulations with the
Gaussian distributions G (mean, std) of all parameters:
v1
= G (10, 30); v2
= G (70, 30); v3
= G (20, 30); TM
= G (15, 30)ms; TA
= G (200, 100)ms; TCSF
= G (800, 200)ms.
Typical values used for
the water fractions and relaxation times in
the simulations were found in the
literature1,2. Notably, a creation of training sets and computing of the
Bayesian estimator is performed on a laptop within minutes. As a
result, the following estimation of the myelin water fraction and the
related relaxation times for in vivo data
takes a few seconds. Fig. 3 presents the images of myelin related
maps estimated by the Bayesian regression and non-negative least
squares4.Discussion and conclusion
The presented framework with a Bayesian regression
approach is very fast and simple method for myelin water imaging. The
developed optimised framework keeps a quite wide range of options for
the myelin model representation by choosing different signal behaviour,
noise level, and parameter distributions. The computational
superiority and simplicity of the implementation allows one to use
the Bayesian regression and proposed pipeline as a helpful tool in
myelin model development and clinical research.Acknowledgements
This
work was funded by the Research Council of Norway (249795).References
1. MacKay, Laule,
Brain Plast. 2 (2016) 71-91
2. MacKay, Laule,
Vavasour et al., Magn Reson Imaging 25 (2006) 515-525
3.
Weiger, Froidevaux, Baadsvik et al., Neuroimage 217 (2020) 116888
4.
Doucette, Kames, Rauscher, Z. Med. Phys. 10.1016/j.zemedi.2020.04.001
5.
Does, Olesen, Harkins et al., Magn Reson Med 81 (2019) 3503-3514
6.
Kellner, Dhital, Kiselev, Reisert, Magn Reson Med 76 (2016) 1574-1581
7.
Reisert, Kellner, Dhital, Henning, Kiselev, Neuroimage 147 (2017)
964-975