Fróði Gregersen1,2,3,4, Cihan Göksu2,5, Gregor Schaefers6,7, Rong Xue4,8,9, Axel Thielscher1,2, and Lars Hanson1,2
1Section for Magnetic Resonance, DTU Health Tech, Technical University of Denmark, Kgs Lyngby, Denmark, 2Danish Research Centre for Magnetic Resonance, Centre for Functional and Diagnostic Imaging and Research, Copenhagen University Hospital, Amager and Hvidovre, Denmark, 3Sino-Danish Center for Education and Research, Aarhus, Denmark, 4University of Chinese Academic of Sciences, Beijing, China, 5High-Field Magnetic Resonance Center, Max-Planck-Institute for Biological Cybernetic, Tübingen, Germany, 6MRI-STaR-Magnetic Resonance Institute for Safety, Technology and Research GmbH, Gelsenkirchen, Germany, 7MR:comp GmbH, MR Safety Testing Laboratory, Gelsenkirchen, Germany, 8State Key Laboratory of Brain and Cognitive Science, Beijing MRI Center for Brain Research, Institute of Biophysics, Chinese Academy of Sciences, Beijing, China, 9Beijing Institute for Brain Disorders, Beijing, China
Synopsis
Combining transcranial electrical stimulation (TES) with MRI offers
various interesting research opportunities, but also introduces safety
concerns. Coupling between the RF field and highly conductive TES leads can
lead to skin burns. These safety issues are usually mitigated with the use of
safety resistors and controlled lead paths that reduce the power absorbed by
the leads. However, these methods introduce practical limitations for combined
TES/MRI experiments, such as limited stimulation currents and cable stray
fields corrupting MR current density imaging. We overcome these limitations by
using low-conductivity silicone-rubber as TES leads. Simulations and
temperature measurements are used for safety assessment.
Introduction
In recent years, there has been a
growing interest in combined MRI and transcranial electric stimulation (TES)
experiments such as TES/fMRI studies for neuroscientific research and MR
current density imaging (MRCDI) to map the injected currents with MRI. Extra safety
measures have to be taken when introducing high-conductivity lead wires into
the MR environment1. For specific resonance conditions,
coupling between the RF field of the scanner and the current injection setup
can result in burns of the subject’s scalp. For the TES device most commonly
used in combination with MRI (DC-STIMULATOR MR, neuroCare Group GmbH, München,
Germany) the safety issues are addressed by using 5 kΩ safety resistors to
shorten the highly conductive path of the leads near the head. With 30 V supply,
the stimulation current is limited to under 3 mA. Additionally, the
twisted leads have to exit the head coil as seen in fig. 1a. This restricts
lead configuration, which is a problem for MRCDI experiments as stray magnetic
fields from the leads are detrimental when measuring fields from currents
inside the brain2.
The aim of this work was to design new lead wires with
flexible lead configurations and higher maximum currents while focusing on MR safety.
Additionally, we wanted the leads to be safe at both 3T and 7T, whereas the currently
available systems are not approved for 7T.Methods
To avoid the possibility of high-amplitude standing waves we
constructed new leads from carbon-doped silicone-rubber with a conductivity of
29.4 S/m (ELASTOSIL® R 570/60 RUSS, Wacker, Munich, Germany). The
silicone-rubber is routinely used in the MR environment as surface electrodes
for TES. We constructed two electrode types, namely circular and
center-surround as seen in fig. 1b-c. The leads are 90 cm long and their
cross-sectional area is 10 mm2. The resistance of each lead wire is
2 kΩ ± 200 Ω.
Finite-difference time-domain (FDTD) simulations were
performed in Sim4Life (ZMT, Zurich, Switzerland) for safety evaluations of the
electrode and leads. The body model Duke from the IT’IS foundation was used in
the simulations3. We
first investigated the relationship between the conductivity of the leads and
the severity of the resonant ‘antenna effect’, where standing waves caused by
the RF field are formed on leads with specific lengths. These simulations were
performed at 298 MHz (proton Larmor frequency at 7T). The setup is shown
in fig. 2a with the high-pass birdcage head coil used for 298 MHz
simulations (7T volume T/R, Nova Medical, Wilmington, MA). To find the resonance
length, the lead length was varied from 0 cm to 100 cm with
10 cm increments including 25 cm and 75 cm. Simulations were
also performed with the worst-case length (25 cm) with varying lead
conductivity. The severity of the ‘antenna effect’ was assessed by evaluating
the power loss on the electrodes.
We also simulated the realistic lead configurations as shown
in fig. 2b-e. This was both done at 298 MHz and 128 MHz (proton
Larmor frequency at 3T). A generic body coil was used for simulations at 128 MHz
(shown in fig. 2d-e). The specific absorption rate (SAR) was obtained from the
FDTD simulations. Following the international guidelines IEC 60601-2-334 we evaluated the influence the
electrodes and leads have on SAR by comparing head SAR and 1 g local head SAR
for a reference simulation with simulations including electrodes and leads. We
also performed temperature measurements with high SAR sequences to ensure
compliance with international guidelines4.Results and discussions
The worst-case ‘antenna effect’ was found for ¼ RF wavelength
corresponding to 25 cm in air for 298 MHz simulations (fig. 3a). This
is in agreement with previous experimental data5. The antenna effect is often believed
to occur at integer multiples of ½ RF wavelength, but this depends on the
boundary condition of the leads5. In fig 3b for varying conductivities
at the worst-case length, it is clear that the antenna effect is eliminated for low-conductive materials, such as
silicone-rubber.
The electrodes/leads have some influence on the distribution
of local SAR levels (fig. 3c), but the peak 1 g local head SAR is always
in the same location as for the reference simulations. The results for each
electrode type and lead configuration as shown in fig. 2b-e is presented in
table 1. Rm and R1g are the ratios of head SAR and peak
1 g local head SAR for a given simulation to reference simulation. Only
minimal SAR changes are observed both before and after normalization of the B1
fields.
Results from temperature measurements are shown in table 2
with one example shown in fig. 3d. The numbers on the electrodes in fig. 3d
indicate the positions given in table 2. More heating was observed for the
center-surround electrodes, although there was no difference when the leads
were attached. No heating above the guidelines of 39 °C was observed4.Conclusion
We have overcome the limitations of
the commercial TES/MRI equipment by using low-conductive silicone-rubber. The
elimination of the antenna effect allows for more flexible lead configurations
and the lower overall impedance increases the maximum allowed stimulation
current. Additionally, we have shown that the setup can safely be used at both
3T and 7T.Acknowledgements
This study was supported by the Lundbeck
Foundation (grants R313-2019-622 and R244-2017-196 to AT), the Chinese National Major
Scientific Equipment R&D Project (grant ZDYZ2010-2), and a PhD stipend of
the Sino-Danish Center for Education and Research to FG. The authors thank Zuo
Zhentao, Hasan Hüseyin Eroğlu, Vincent Boer, and Esben Thade Pedersen for kind technical help.References
1. Panych LP, Madore B. The physics of MRI safety. J Magn
Reson Imaging. 2018;47(1):28-43. doi:10.1002/jmri.25761
2. Göksu C,
Scheffler K, Siebner HR, Thielscher A, Hanson LG. The stray magnetic fields in
Magnetic Resonance Current Density Imaging (MRCDI). Phys Medica.
2019;59(February):142-150. doi:10.1016/j.ejmp.2019.02.022
3. Gosselin
MC, Neufeld E, Moser H, et al. Development of a new generation of
high-resolution anatomical models for medical device evaluation: The Virtual
Population 3.0. Phys Med Biol. 2014;59(18):5287-5303.
doi:10.1088/0031-9155/59/18/5287
4. IEC.
Particular requirements for the basic safety and essential performance of
magnetic resonance equipment for medical diagnosis. IEC 60601-2-33. 2010.
5. Balasubramanian
M, Wells WM, Ives JR, Britz P, Mulkern R V., Orbach DB. RF Heating of Gold Cup
and Conductive Plastic Electrodes during Simultaneous EEG and MRI. Neurodiagn
J. 2017;57(1):69-83. doi:10.1080/21646821.2017.1256722