Alix Plumley1, Philip Schmid1, and Emre Kopanoglu1
1Cardiff University Brain Research Imaging Centre, Cardiff, United Kingdom
Synopsis
Subject motion in parallel-transmit (pTx) causes
channels’ electric field interference patterns to change, influencing SAR distributions.
This can cause safety limits to be exceeded when SAR-constrained pulses are
designed for one position. Here, we consider effects of pTx coil dimensions on SAR sensitivity to motion by simulating 6 differently-sized coil models, and evaluating SAR at 19
displaced positions. Our results agree with those previously reported for
the similar-sized coil, but SAR sensitivity was generally lower for larger coils, and higher for smaller coils, with maximum
motion-induced local-SAR increase of 3.8-fold and 1.6-fold for the
smallest and largest coil models, respectively.
Introduction
Parallel transmission (pTx) can overcome B1+ inhomogeneity in 7T
head imaging1-3, however specific absorption rate (SAR) hotspots and
associated tissue heating can exceed safety limits when head motion occurs due
to constructive interference between channels4-6. Motion-induced increases
in peak local-SAR of over 200% have been reported for multi-spoke pTx pulses, along
with unpredictable changes in hotspot location4.
Interference patterns depend on the coil as well as the load, therefore
SAR depends on coil geometry7-8. Here,
we investigate whether pTx coil element dimensions affect SAR motion-sensitivity.
For this, we compare motion-induced SAR changes using simulations with 6 different
coils.Methods
6 coil models (A: smallest, to F: largest) consisted of 8-channel loop
arrays; the height, width (of loops), and radius (of array) of which varied ±25%,
±25%, and +25% respectively, compared to the base model C (figure 1). For each
coil model, 1 central, and 19 off-centre positions were simulated using the
Ella model9 in Sim4Life
(ZMT, Zurich, Switzerland), totalling 120 datasets. SAR was reported as most
sensitive to axial displacements4, so we focus on these. Simulation parameters
followed those in [4]. Positions were defined with respect to the array origin
across coil models.
Pulses were designed to uniformly excite an axial slice in
quadrature and pTx (1/2/3/5-spokes) modes (1-spoke is RF-shimming). Pulses were
designed with an adaptation of [10-11] using B1-distributions at the central
position for each coil. Pulses were not SAR-constrained, but RF power was
penalized (along with magnitude excitation error) in the cost function (with Tikhonov
regularization = 0.1). Separate pTx pulses were designed for 6 slices (figure
1), yielding 25 pulses per coil.
Motion-sensitivity was quantified as SAR at each off-centre
position normalised by that at the corresponding central position (SARcentre)
for each coil respectively. Whole-head (gSAR) and peak 10-gram averaged
local-SAR (psSAR) were evaluated at all positions using voxelwise and 10-g
average Q-matrices12 respectively. Tissue volumes exposed to higher
SAR than psSARcentre are also investigated (named ‘high-SAR tissue’
here). To exclude pulse design effects, eigenvalue-based SAR (eigSAR; assumes worst-case
channel interferences) was also calculated using the 10-g Q-matrices.Results and Discussion
Coil C has similar dimensions to that in [4] and exhibited similar
SAR-sensitivity to motion as expected (RF-shimming: 2.7 vs 2.4-fold psSAR
increase). For the larger coils, psSAR sensitivity was lower; maximum increase for coil F was 1.3-fold, whereas for the
smallest coil (A) we observed a much higher local-SAR increase of
3.8-fold. Similarly, coil C’s eigSAR increased up to 43%, matching previous
findings4, however eigSAR was more motion-sensitive among smaller
coil models (especially to posterior shifts) where it increased by 72%.
For pTx pulses, gSAR increased by up to 38% and 49% for RF-shim
and multi-spoke pulses, respectively (both coil A). psSAR for RF-shimming was
the most motion-sensitive metric, and was especially sensitive among smaller
coils. psSAR increased by 3.8-fold, 1.7-fold, and
1.3-fold in the worst RF-shim case for coils A, D, and F, respectively. Figure 2 shows RF-shim SAR-sensitivity
for a representative slice (slices 2-5 showed similar patterns, while 1 and 6
did not - discussed later).
Figure 3 shows coils’ worst-cases across all pTx (including
RF-shim) pulses. psSAR sensitivity and high-SAR tissue was lowest for the
taller coil models, where psSAR increased by a maximum of 96% and 59% for coils
D and F. As the number of spokes in pulses increases, SAR-sensitivity
generally reduced and became more similar across coil models (figure 4.i),
however the three largest coil models’ worst-cases were multi-spoke pulses. Worst-cases per slice show that coils D-F yielded less high-SAR
tissue than smaller coils throughout the
central slices (figure 4.ii).
For quadrature pulses, gSAR was relatively stable, increasing by
less than 5% (coil A). psSAR was most sensitive for coil C where it almost doubled,
while psSAR for coils A and E increased by 18% and 5%, respectively (figure 5).
Unlike pTx, in all quadrature worst-cases, the local-hotspot shifted from the
front to back of the head. Even without motion, coil A experienced
intense local-hotspots, meaning psSAR change was low when the hotspot
shifted but retained similar intensity. For large coils, fields
were smoother throughout the head, leading to less intense hotspots even after
motion. Coil C did not exhibit either behaviour, hence the largest relative
increase.
SARcentre (not shown) was higher when pulses were
designed using slices 1 and 6. This is less concerning than motion-sensitivity
of SAR, as SARcentre can be constrained during pulse design. For
coil B, these slices fell beyond the loops’ vertical extent, and motion-sensitivity
was notably lower than other slices (though psSARcentre was around
threefold). For 14% of pulses, worst-case psSAR occurred at intermediate positions
(ie. not the extremes). These pulses were for larger coils and/or slices 5 and
6, where psSAR sensitivity was lower.Conclusions
SAR for smaller pTx coils was generally more motion-sensitive than that for larger coils (wider or taller loops, and/or larger
array radius). Maximum observed local-SAR increases due to motion were 3.8-fold and 1.6-fold for the smallest and largest coil models respectively. This is independent of pulse design, since eigenvalue-based SAR
(which depends only on transmit fields) followed this pattern; however
pTx pulses with small coils elicit the most concerning local-SAR increases.Acknowledgements
No acknowledgement found.References
[1] J. Vaughan, et
al., 7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head
images. Mag Res Med,64 (1), pp. 24–30, 2001.
[2] U. Katscher and
P. Bornert, Parallel RF transmission in MRI. NMR Biomed,19 (3), pp.
393–400, 2006.
[3] C. M. Deniz,
Parallel transmission for ultrahigh field MRI. Topics in Mag Res Imag,28 (3),
pp. 159–171, 2019.
[4] E. Kopanoglu, et
al., Specific absorption rate implications of within-scan patient head motion
for ultra-high field MRI. Mag Res Med, 2020.
[5] S. Wolf, et al.,
SAR simulations for high-field MRI: How much detail, effort, and accuracy is
needed? Mag Res Med, 69 (4), pp. 1157–1168, 2013.
[6] N. Schon, et
al., Impact of respiration on B1+ field and SAR distribution at 7T using a
novel EM simulation setup. Proc. Intl. Soc. Mag. Reson. Med., 28, 2020.
[7] U. Katscher, et
al., Basic considerations on the impact of the coil array on the performance of
transmit sense. Mag Res Mat Phys Biol Med, 18 (2), pp. 81–88, 2005.
[8] C. M. Deniz, et
al., Radiofrequency energy deposition and radiofrequency power requirements in
parallel transmission with increasing distance from the coil to the sample. Mag
Res Med, 75 (1), pp. 423–432, 2016.
[9] M-C. Gosselin, et al., Development of a new generation of
high-resolution anatomical models for medical device evaluation: The virtual
population 3.0. Phys Med Bio, 59 (18), pp. 5287–5303, 2014.
[10] W. Grissom, et al. Spatial domain method for the design of RF pulses in
multicoil parallel excitation. Magnetic Resonance in Medicine, 56,
620–629, 2006.
[11] ISMRM RF Pulse-design Challenge.
http://challenge.ismrm.org/. 2016.
[12] I. Graesslin, et al., A specific absorption rate prediction concept
for parallel transmission MR. Mag Res Med, 68 (5), pp. 1664–1674, 2012.