Jason Meyers1, David Prutchi1, and Ramez Shehada2
1Impulse Dynamics (USA) Inc., Marlton, NJ, United States, 2Medical Technology Laboratories, La Mirada, CA, United States
Synopsis
The MR environment poses a tissue heating hazard to patients
with cardiac IPGs (implantable pulse generators), such as pacemakers and defibrillators,
due to RF currents circulating in the loop formed by the IPG, a transvenous lead,
and the tissue. Heating will be limited
by the impedance of this loop at the MR frequency (63.87 MHz for 1.5T). The IPG
input impedance (a portion of this loop) was measured in six MR-Conditional IPG’s.
All have a comparable, low impedance, suggesting
that device manufacturers are not intentionally adding impedance and that some
interchangeability may be possible without changing
RF-induced heating.
Introduction and Background
The MR environment presents a number of potential hazards to
patients with cardiac implantable electronic device systems. Tissue damage due to RF heating is one such
hazard for which tests must be performed on a complete system consisting of an
IPG (implantable pulse generator) and one or more leads. Practicalities limit the number of
permutations that are tested.
This leads to a lack of interchangeability between IPGs and
leads which has numerous downsides for patients, including potentially
requiring replacement of functional leads to maintain MR-conditional labeling
and limiting therapy options. We
measured the IPG characteristics relevant to the RF heating hazard. This, in combination with similar testing of MR-Conditional
leads, should contribute to an understanding of cases in which interchanging
IPGs and leads may be safe.
The MR environment has a strong RF field present at the
Hydrogen precession frequency of 63.87MHz in a 1.5T MRI scanner. A system consisting of an IPG and a lead
creates an electrical loop along the lead, through the IPG (from the lead
connection to the IPG’s enclosure, referred to as the “Can”) and back to the
end of the lead through tissue, as shown in Figure 1. The RF field induces current in this loop,
which in turn causes heating (which is most concentrated at the distal "Tip" electrode). The magnitude of the circulating current, and
therefore the heating, is dependent on the impedance of this loop.
In standard IPG construction, the Tip and Ring terminals of
a lead attach to a connector block in the header. From there, a wire routes through a hermetic
feedthrough and then connects to IPG circuitry inside the can. An EMI filtering capacitor just inside the Can
will provide a low-loss path between the lead connection and the Can. As a first order model, the header block to Can
impedance can be modeled as the series combination of the feedthrough wire
impedance and the capacitor impedance.
Evaluating this model shows that the inductance of the
feedthrough wire dominates, and that impedance magnitudes of around 1-5 Ω
are to be expected, dependent primarily on wire length. We hypothesize that the impedance contributed
by the IPG is a consequence of the standard methods of header construction
employed by nearly all cardiac IPGs, that MR-conditional IPGs are not intentionally
adding impedance at this frequency, and that therefore these IPGs will all
present comparable impedances.Methods and Results
Impedance measurements of six IPGs with MR-Conditional
labeling were performed using a VNA (vector network analyzer). These test IPGs
were modified to allow direct access to the Tip and Ring connectors by removing
septum seals and, when necessary, machining a hole into the header. In each IPG, the header cavity closest to the Can was
tested as it has the shortest feedthrough wires, resulting in the
worst-case (lowest) impedance.
A primarily 3D-printed test fixture was constructed to
perform these measurements, as shown in Figure 2. The frame holds a probe arm which contains a
contact point for the Can (adjacent to the header) and a contact point which
reaches into the header. These contact
points are assembled onto a small piece of hand-formable coaxial cable, which
is terminated with an SMA connector to attach to the VNA. This construction method was adopted to
minimize test fixture residuals.
The VNA was calibrated at the end of the test cable using a
mechanical open-short-load calibration kit.
The fixture was then installed, and the measurement reference plane was
extended to the IPG connection. The
complex reflection coefficient (S11) was measured and imported into Matlab,
which was used to calculate the complex impedance and to apply an open-short
impedance correction.
Measurement results are presented in Table 1 and
plotted in Figure 3.Discussion and Conclusions
All measured IPGs show an inductive impedance with a magnitude
between 1 and 5.87 Ω. These values vary over
this range primarily due to the feedthrough wire length and are congruent with
the model described above. This shows
that no additional impedance has been intentionally incorporated by the
manufacturers of these devices to
limit RF-induced electrode heating, showing that it is possible to achieve MR-conditional
labeling with standard header construction, and suggesting that in some cases
interchangeability may be possible without changing RF-induced
heating.Acknowledgements
This work was supported
by Impulse Dynamics (USA) Inc.References
No reference found.