Weiman Jiang1, Fan Yang1, and Kun Wang1
1GE Healthcare, Beijing, China
Synopsis
This work proposed a novel SAR
characterization method based on an equivalent circuit model and the circuit’s
frequency response analysis. Comparing to well recognized pulse energy method
defined in NEMA MS 8, this method doesn’t need the flux loop fixed on transmit
coil. Meanwhile, it can monitor SAR accurately. The Root
Mean Square (RMS) error and maximum error of the novel method relative to the
method in NEMA MS 8 are 4.96% and 9.47% respectively. This method could avoid
design complexity of integrating the flux loop and has potential to be easily
realized in most MRI scanners.
Introduction
Thermal injury risk introduced by RF field is one of the most
concerned bio-effects in MRI scan.
To protect patient from RF heating, whole body SAR is the critical
value representing the thermal energy absorbed by patients. Especially for the
medically diverse patients, such as pregnant women and patients with implant,
whole body SAR is required appropriate controlment in MR environment. Currently,
commercial MRI scanners usually have relatively large error on SAR monitor,
which results in the clinicians and implants manufactories don’t have enough
confidence with the scanner reported SAR. Accurate monitor of individualized
patient SAR needs extra flux loop installed near body coil according to pulse
energy method defined in NEMA MS 8.1 It increases the system design
complexity of integrating the flux loop. In this work, we proposed a novel SAR
characterization method based on an equivalent circuit model and the circuit’s
frequency response analysis. The feasibility and reliability of this method were
verified in five volunteers imaged in MRI. Without the flux loop, the monitored
whole body SAR shows good accuracy comparing to that monitored by pulse energy
method with the flux loop.Methods
For SAR characterization
by pulse energy method with the flux loop, the total RF power transmitted into
the coil terminal was measured with directional couplers connecting
oscilloscope or power meter, as shown in Figure 1A. Two flux loops connecting
oscilloscope or power meter were used to measure the coil absorbed power.
Subject absorbed power is equal to the difference between the total power and
coil absorbed power. Then subject SAR can be measured.
In this work, we used an equivalent parallel
circuit model to demonstrate the power distribution principle on the coil and
subject.2 Figure 1B shows the
experiment setup basing on this model. The flux loop was not needed. Figure 2 illustrates the equivalent circuit model of the coil with and without
subject in the body coil. The circuit consists of the
power source part and the coil part. For
the coil loaded with subject, the RF power absorbed by the coil and subject are
dealt with the power deposited on two parallel resistors (Figure 2B). Once the
total power absorbed by the coil, coil resistance, and subject resistance are
known, the coil and subject absorbed power can be calculated with
$$$P_{coil}=\frac{R_{es}}{R_{ec}+R_{es} }∙P_{total}$$$ (1)
$$$P_{subject}=\frac{R_{ec}}{R_{ec}+R_{es} }∙P_{total}$$$ (2)
Where $$$P_{total}$$$ is the total
power absorbed by the coil and subject, $$$P_{coil}$$$ and $$$P_{subject}$$$ are the coil and subject absorbed power, and $$$R_{ec}$$$ and $$$R_{es}$$$ are the coil
and subject resistance.
In the experiment setup (Figure1B), $$$P_{total}$$$ is equal to the difference between forward and
reflected power, which can be measured with the directional couplers. $$$R_{ec}$$$ and $$$R_{es}$$$ can be fitted
with circuit’s frequency response by measuring the power reflection rate at
coil terminal for the empty and loaded coil:3
$$$|S11|=|\frac{V_r}{V_f} |=|\frac{√P_r }{√P_f }|=\frac{|R_{ec}-50|}{|R_{ec}+50|} ∙|\frac{cos(atan(\frac{50}{R_{ec}+50 }Q_{ec} (\frac{ω}{ω_0} -\frac{ω_0}{ω})))}{cos(atan(\frac{50}{R_{ec}-50} Q_ec (\frac{ω}{ω_0} -\frac{ω_0}{ω})))}|$$$ (3)
$$$|S11|=|\frac{V_r'}{V_f'} |=|\frac{√P_r'}{√P_f' }|=\frac{|R_{el}-50|}{|R_{el}+50|} ∙|\frac{cos(atan(\frac{50}{R_{el}+50 }Q_{el} (\frac{ω}{ω_0'} -\frac{ω_0'}{ω})))}{cos(atan(\frac{50}{R_{el}-50} Q_el (\frac{ω}{ω_0'} -\frac{ω_0'}{ω})))}|$$$ (4)
Where $$$P_f$$$ and $$$P_f'$$$ are the measured forward power, and $$$P_r$$$ and $$$P_r'$$$ are measured reflected power, at coil terminal for the empty and loaded coil.
With known $$$P_{total}$$$, $$$R_{ec}$$$, and $$$R_{es}$$$, subject absorbed power can be calculated with
Equation 2. Subject SAR then can be obtained. According to the circuit theory,
the equivalent circuit model is applicable for SAR monitor in a relatively wide
RF frequency range near the coil center frequency, in which the power deposited
on capacitor and inductor could be ignored.
To verify
this method and compare it with pulse energy method, we applied both two methods
to five healthy volunteers imaged in compliance with the Institutional Review
Board requirements using a body coil at 1.5T scanner (GE, SIGNA Artist).
Results
Figure 3
illustrates SAR for the five volunteers landmarked at chest and abdomen, monitored
by pulse energy method with the flux loop and equivalent circuit model without
the flux loop. SAR in the frequency range of 63.76-64.06 MHz were monitored to
verify the applicability of the equivalent circuit model. It is obvious that SAR
monitored by the two methods are in good agreement.
Figure 4
shows the error of volunteer SAR monitored by the equivalent circuit model,
from that monitored by pulse energy method with the flux loop. The error is
defined as
$$$Err=\frac{SAR_{circuit\,model}-SAR_{flux\,loop}}{SAR_{flux\, loop}} $$$ (5)
For the scanned
volunteers, the RMS error and maximum error are 4.96% and 9.47% respectively.
Discussion
This
work proposed a novel accurate method to individualized monitor the whole body
SAR. A parallel RLC circuit model was used to reveal well the power
distribution on the subject and coil during MRI scan. The subject in a coil was
treated as a resistor paralleled with the coil’s equivalent resistor. By
measuring the frequency response of the power reflection rate at the coil
terminal for empty and load coil, the equivalent resistance of the coil and
subject can be fitted. And the fraction of the subject absorbed power in the
total power transmitted into the coil was determined. By measuring the total
power with directional couplers connecting oscilloscope (or power meter), the
subject absorbed power and the subject SAR can be determined easily.Acknowledgements
No acknowledgement found.References
1. National Electrical Manufacturers Association.: NEMA
Standards Publication MS 8 – 2016: Characterization of the Specific Absorption
Rate for Magnetic Resonance Imaging Systems. National Electrical Manufactures
Association, Rosslyn, Virginia, USA (2016).
2.
Smith K.C.A., Alley R.E..
Electrical circuit theory. Cambridge University Press, 1992, page 164-166.
3. Joel
P.D.. Handbook microwave component measurements with advanced VNA techniques.
John Wiley & Sons, Inc., 2020, page 4-11.