Camille D.E. Van Speybroeck1, Wouter M. Teeuwisse1, Tom O'Reilly1, Paul M. Arnold2, and Andrew G. Webb1
1G.J. Gorter Center for High Field MRI, Leiden University Medical Center, Leiden, Netherlands, 2Neurosurgery, Carle Foundation Hospital, Urbana, IL, United States
Synopsis
Very little
quantitative data exist on safety aspects of medical implants in low field MRI
systems. Protocols for clinical scanners must be adjusted due to differences in
bore size and B0 direction. We present measured and simulated data
on magnetically induced displacement forces, image artifacts and SAR on a 50 mT
portable MRI system. Results show that some implants must be considered MR
unsafe even at 50 mT, that image artifacts are less compared to 3 T even with
much lower gradient strengths, and SAR limits can be breached if using very
short pulses and inter-echo times in TSE sequences.
Introduction
Recent advances in low-field
portable MRI systems enable new applications for MRI as well as operation in
venues other than mainline hospitals1,2. Potential advantages of
low-field MRI include reduced patient contraindications due to implanted
medical devices, more benign image artifacts and lower specific absorption
ratio (SAR). However, there are currently no quantitative assessments of these
aspects, nor comparisons to standard clinical systems. Due to the smaller size
and axial B0 configuration in Halbach arrays, application of ASTM
test protocols is not straightforward. We provide the first quantitative
results for magnetically induced forces, image artifacts and SAR from medical
devices with different geometries and susceptibilities on a 50 mT Halbach-based
system3, incorporating suitable adaptations of ASTM protocols.Methods
Magnetically induced displacement forces (Fm): For implanted
medical devices, assessment is described by ASTM F2052-154, which
measures the angle of deflection α to relate the values of Fm and Fg, the
gravitational force. Rather than measurement solely axially for a Halbach array
we assess the spatial gradient in two directions (x and z). The measurement was
made in the region of highest spatial gradient for a septal occluder (Amplatzer),
an Iliac stent (CardiacDevices), an endoscopic clip (Resolution, Boston
Scientific) and a glucose sensor (Freestyle Libre, Abbott). The custom-built
apparatus is able to rotate (Figure 1a).
Image artifacts arising from implanted medical devices:
ASTM F2119-075 quantifies image artifacts produced
by implants. Individual implants, with a reference nylon rod, are placed in a
9cm diameter container filled with CuSO4 solution to shorten T1.
Data were acquired in every combination of frequency- and phase-encoding with
respect to B0 using a TSE sequence. Images were acquired on the 50 mT
system with a bandwidth of 160Hz/pixel (the maximal bandwidth) and a Philips 3 T
system for comparison. Two TSE scans were acquired on the 3 T with a bandwidth
of 160 and 880Hz/pixel respectively. Also, gradient echo (GE) scans were acquired
at 160Hz/pixel with a TE of 4.5ms at 3 T and 16ms at 50 mT (minimal TE of the
50 mT scanner) with and without the implant.
Specific absorption rate (SAR): Although several studies simply state that SAR is not problematic at
very low field, there has been little quantitative assessment, therefore simulations
were performed to determine under what conditions SAR limits might be reached,
even at 2.15MHz. Since all medical implants are much smaller than λ/2, the same SAR is expected with
and without the implant, and is given by6:
$$\mathrm{SAR}=6.81\times 10^{-19}\frac{\nu^2 R^2}{\rho s \mathrm{TR}}\sum_i \frac{N_i}{\tau_i}\left(\frac{2\theta_i}{\pi}\right)^2$$
where Niθi pulses (duration of τi)
are applied, R is the radius of the imaged body part, here
0.092m for the brain, ρ its resistivity (2.17Ωm),
and s its specific gravity (1.07)6,7. Since
TSE sequences are most commonly used at low field, and produce highest SAR, we
simulated the SAR for different pulse durations, echo train lengths and TR/TE
values.Results
Displacement: Table 1 shows the
results of the deflection angle measurements. The stent and occluder showed no
displacement, while the clip was directly drawn to the edge of the bore. The glucose
sensor displacement was measurable between z = -60mm and z = 60mm: at larger z
the sensor was drawn against the magnet. At z = ±60mm Fm = 11.1mN. As demonstrated by the
simulation data (Figure 1b), the spatial gradient of B0 does not change
much outside the measurement area, and so the sensor can be considered MR safe with
respect to the displacement force at 50 mT.
Artifacts: Figure 2 shows the worst case
orientations of three implants. The image artifacts for all implants are
smaller at 50 mT than at 3 T, even with the larger 3 T bandwidth. Figure 3 shows
a comparison of GE and TSE scans at 50 mT for the septal occluder.
SAR: Using a 1kW amplifier at 50 mT pulse
durations of ~100μs are applied, covering the typical linewidth over the brain of ~2-4kHz. Figure 4a shows that even
for ETLs of 64 with inter-echo spacings of a few ms the SAR is below the head
limit8. For the situation for magnets with ten times lower homogeneity,
simulations were performed using a pulse duration of 10μs. In this case Figure 4b shows that SAR limits are
reached for inter-echo times of a few ms for an ETL of only 8.Discussion
This study
describes the first quantitative assessments of medical device compatibility,
image quality and SAR for a low field portable MRI system. The results show
that displacement forces are less than for a clinical system, but ferromagnetic
implants may still be classified as MR unsafe. Despite the much lower gradient
strengths available on the 50 mT system, image artifacts for medical devices
are smaller than those at 3 T using maximum gradient strength. Quite severe
artifacts are still present for ferromagnetic implants. Finally, results show
that SAR is well below limits for even very long echo train length TSE
sequences for normal operating conditions. However, if pulse durations have to
be very short in strongly inhomogeneous magnetic fields, and inter-echo times
very short to minimize diffusive signal attenuation in such fields, then SAR
limits can be breached.Acknowledgements
This
work is supported by the following grants: Horizon 2020 ERC Advanced NOMA-MRI
670629, Simon Stevin Meester Award.References
1. Wald L L,
McDaniel P C, Witzel T, et al. Low-Cost and Portable MRI. J Magn Reson Imaging.
2019;52(3):686-696. 10.1002/jmri.26942.
2. Cooley C Z,
McDaniel P C, Stockmann J P, et al. A Portable Brain MRI Scanner for
Underserved Settings and Point-Of-Care Imaging. Pre-print. 2020.
3. O'Reilly T, Teeuwisse W M, de Gans
D, et al. In vivo
three-dimensional brain and extremity MRI at 50 mT using a permanent magnet
Halbach array. Magn Reson Med. 2021;85(1):495–505. 10.1002/mrm.28396.
4. ASTM
International. Standard Test Method for Measurement of Magnetically Induced
Displacement Force on Medical Devices in the Magnetic Resonance Environment, F2052
– 15. West Conshohocken, PA, United States: ASTM International; 2015.
5. ASTM
International. Standard Test Method for Evaluation of MR Image Artifacts from
Passive Implants, F2119 – 07. West Conshohocken, PA, United States: ASTM
International; 2013.
6. Bottomley P
A, Edelstein W A. Power deposition in whole-body NMR imaging. Med Phys.
1981;8(4):510-512.
7. Grandolfo M, Vecchia P, Gandhi O P. Magnetic
Resonance Imaging: Calculation of Rates of Energy Absorption by a Human-Torso
Model. Bioelectromagnetics. 1990;11(2):117-128.
8. International
Electrotechnical Commission. Medical Electrical Equipment - Part 2-33:
Particular Requirements for the Basic Safety and Essential Performance of
Magnetic Resonance Equipment for Medical Diagnosis, IEC
60601-2-33:2010+AMD1:2013+AMD2:2015 CSV. International Electrotechnical
Commission; 2015.