Negin Behzadian1 and Shiloh Sison2
1Research and Development, Abbott, Sylmar, CA, United States, 2Research and Development, Abbott, Sunnyvale, CA, United States
Synopsis
MRI RF heating can cause thermal damage to biological tissues,
especially in patients with leaded implantable devices, by concentrating power
deposition at the implant-tissue interface. With the emergence of B1+RMS
as a more precise RF exposure metric than SAR, implantable device manufacturers
have begun MR conditional labeling to fixed B1+RMS levels, in
addition to SAR labeling. Our study characterizes the relationship between RF
heating per SAR vs. B1+RMS-based scaling methods in the context of
3T Normal Operating and First Level Controlled Mode MRI exposure.
Introduction
MRI RF heating can cause thermal damage to biological tissues. Assessment
of 3T MRI RF-induced heating for active implantable medical devices (AIMDs) requires
electromagnetic simulations of clinical RF coils to generate probable RF
E-fields that can induce currents along the length of implants, leading to
concentrated power deposition and local temperature increment at the
implant-tissue interface. The simulated coil excitations must be consistent
with the MR conditional labeling, ranging across Normal Operating Mode,
First-Level Controlled Operating Mode, fixed whole-body Specific Absorption
Rate (wbSAR) or fixed B1+RMS levels. B1+RMS has been recently
adopted by AIMD manufacturers as an additional RF exposure metric for MR conditional
labeling, given that it is neither estimated nor patient-dependent like SAR. Little
has been published however comparing expected RF heating due to SAR vs. B1+RMS
levels, which has generated questions when both SAR and B1+RMS
levels are specified in MR conditional labeling. Our investigation, therefore
aims to characterize equivalence between wbSAR-scaled and B1+RMS-scaled
RF heating predictions in the context of 3T Normal Operating Mode and First
Level Controlled Mode MRI exposure respectively, with a cardiovascular
implantable electronic device (CIED).Methods
We evaluated two different cardiac pacing leads for In Vivo RF MC-2 heating
limited by both 3T Normal Operating Mode and First Level Controlled Mode when
connected to a CIED. In-vitro
temperature rise values were derived for RF exposures scaled to a range of
wbSAR values across 0-4.0 W/kg with a maximum B1+RMS scaling of 7 uT,
as well as for fixed B1+RMS values across 0.5-5.5 uT with
a maximum wbSAR scaling of 2.0 W/kg vs. 4.0 W/kg for Normal and First Level
Operating Mode respectively. We identified wbSAR values at which heating
predictions were equivalent to specific B1+RMS level, namely 1.3 and
2.8 uT, values which AIMD manufacturers currently use in their MR conditional
labeling, as well as B1+RMS values at which heating prediction were
equivalent to specific wbSAR levels of 2.0 and 4.0W/kg, maximum limits imposed
by Normal and First Level Controlled Operating Mode respectively1-3.Results
The normalized in-vitro temperature rise was quantified for
multiple 3T Normal Operating Mode wbSAR-scaled (black) and B1+RMS-scaled
(red) RF heating simulations for two different cardiac leads (Figure 1). The
same is presented for the constraint of First-Level Controlled Mode in (Figure
2).
The (wbSAR, B1+RMS) levels at which the temperature
rises are equivalent under Normal Operating Mode, are plotted in Figure 3,
approximately identical across the two leads: (0.3, 1.3), (1.5, 2.8), (2.0, 4.5).
The (wbSAR, B1+RMS) levels at which the temperature
rises are equivalent under First Level Controlled Operating Mode, are plotted
in Figure 4, approximately identical across the two leads: (0.3, 1.3), (1.7,
2.8), (2.0, 3.0), (4.0, 5.6).Discussion
Our study shows that B1+RMS labeling currently used
by AIMD manufacturers such as 2.8 uT 1,2, is equivalent to a wbSAR
limit of 1.5W/kg under Normal Operating Mode, and 1.7W/kg under First-Level
Controlled Mode. Additionally, a wbSAR limit of 2W/kg is equivalent to 4.5uT
under Normal Operating Mode and 4W/kg is equivalent to 5.6uT under First Level
Controlled Mode. Conclusion
Under Normal Operating Mode, B1+RMS MR conditional
labeling at 3T adopted by manufacturers today is found to have an equivalent
wbSAR, in terms of RF heating generated, of less than 2W/kg. Furthermore, using
B1+RMS conditional labeling at levels up to 4.5 uT is equivalent to
wbSAR up to 2W/kg.
Under First Level Controlled Operating Mode, B1+RMS MR
conditional labeling at 3T adopted by manufacturers today is found to have an
equivalent wbSAR, in terms of RF heating generated, of less than 4W/kg.
Furthermore, using B1+RMS conditional labeling at levels up to 5.6
uT is equivalent to wbSAR up to 4W/kg.
It is expected that because of the margin added to the displayed
wbSAR in addition to the limits imposed by short-duration SAR, wbSAR is more
restrictive than alternative B1+RMS. However if that wbSAR margin were to
decrease then typical 3T B1+RMS labeling would restrict sequences
above 1.5W/kg and 1.7W/kg for Normal and First Level Controlled Operating Mode.Acknowledgements
No acknowledgement found.References
- Abbott.
MRI Procedure Information for Abbott Medical MR Conditional Deep Brain
Stimulation Systems: Clinician’s Manual, 2020. <https://manuals.sjm.com/~/media/manuals/product-manual-pdfs/6/3/631c18a6-6b9d-476f-b938-57c07d53ebe6.pdf>
- Medtronic.
AttestaTM MRI SureScanTM /SpheraTM MRI
SureScanTM Pacing Systems: MRI procedural information for SureScanTM
pacemakers and SureScanTM leads, 2017. <http://manuals.medtronic.com/content/dam/emanuals/crdm/CONTRIB_256428.pdf>
- Boston
Scientific. MRI Technical Guide: IMAGEREADYTM MR Conditional
Defibrillation System, 2019.
<http://irm-compatibilite.com/documents/boston/specific_conditions_boston_defibrillation.pdf>