Mark BM Hofman1, Carla Kloeze2, Bastiaan J van Nierop3, Sara H Muller4, Joost PA Kuijer1, Kristie Venhorst5, Cristina Lavini6, Mette K Stam7, Carola van Pul8, Wouter Teeuwisse9, Marco JW Gotte10, Peter Kappert11, Evie EM Kolsteren5, and Marinette van der Graaf12
1Radiology and Nuclear Medicine, Amsterdam UMC, Vrije Universiteit Amsterdam, Amsterdam, Netherlands, 2Catharina Hospital, Eindhoven, Netherlands, 3Radiology and Nuclear Medicine, University Medical Centre Utrecht, Utrecht, Netherlands, 4Antoni van Leeuwenhoek Hospital, Amsterdam, Netherlands, 5Knowledge Institute Medical Specialists, Utrecht, Netherlands, 6Radiology and Nuclear Medicine, Amsterdam UMC location AMC, Amsterdam, Netherlands, 7Radiology, LUMC, Leiden, Netherlands, 8Maxima Medical Centre, Eindhoven, Netherlands, 9CJ Gorter Center for High Field MRI, LUMC, Leiden, Netherlands, 10Cardiology, Amsterdam UMC, Vrije Universiteit Amsterdam, Amsterdam, Netherlands, 11Radiology, UMCG, Groningen, Netherlands, 12Imaging, Radboud UMC, Nijmegen, Netherlands
Synopsis
Can
a patient with a prosthetic heart valve, annuloplasty ring or a mitra clip
undergo an MRI examination? The Dutch Association of Medical Specialists (FMS) wrote
a guideline, using searches of guidelines, literature, implant databases, and
incident databases. The recommendation is: scan the patient with a prosthetic
heart valve, annuloplasty ring or mitra clip with an 1.5T or 3T whole body MRI
system with a horizontal closed bore superconducting magnet without further restrictions.
Introduction
There is a wide variety of
types of heart valve prostheses and annuloplasty rings for which there are various
specific conditions for performing an MRI. Many of these manufacturer-set
conditions are conservative such that they may impair patients diagnosis. Additionally,
in some cases the type of heart valve or the safety profile of the heart valve
is unknown. Hospitals vary in their policies regarding screening and use of MRI
in these patients. This is the case for all type of MRI examinations and not
only for MRI of the heart.
Therefore the Dutch
Association of Medical Specialists (called FMS) has written a guideline on the
clinical question: can a patient with a prosthetic heart valve,
annuloplasty ring or mitra clip undergo an MRI examination in a 1.5T or 3T whole
body MRI systems with a horizontal closed bore superconducting magnet?Methods
The guideline has been
developed conforming to the standard within the FMS as described in Guidelines
for Medical Specialists 2.0 1, based on the AGREE II instrument 2.
The guideline, initiated by Society for Medical
Physics of the Netherlands (NVKF), was written by a working group together with participants from the Dutch Society for Medical Imaging and
Radiotherapy (NVMBR), the Radiological Society of the Netherlands (NVvR), the Netherlands
Society of Cardiology (NVvC) and advisors from the Knowledge Institute of the
FMS.
A
series of searches were performed: guideline search, literature search, implant
search on MRI safety.com and Magresource 3, and a search on implant
incident databases of FDA, the Health Inspection Service in the Netherlands,
and the 'Implant' and 'Event' database of the International Consortium of Investigative
Journalists.
In order to come to a
recommendation, in addition to (the quality of) the scientific evidence, other
aspects are important and were taken into account, such as the expertise of the
working group members, cost/benefits, availability of facilities and organization
of healthcare. These aspects were discussed in the Considerations.
The concept guideline was
subjected to commentaries by the involved Dutch (scientific) associations,
agencies and (patient) organizations. The final version was authorized by the end
of 2019 by the FMS.Literature Review
Literature was searched to answer the question “Can a patient with a prosthetic
heart valve or annuloplasty ring undergo MRI examination?”. The conclusion is that during MRI examinations at 3T or below, no
interactions between a prosthetic heart valve and the magnetic fields and radio
frequency waves caused by the MRI scanner have been detected that could be
harmful to the patient 4-14. Patients with degenerative valve
disease are unlikely to have a greater risk of valve loosening as a result of
MRI 15. Up to 1.5T, there is some evidence, although limited, that
MRI does not cause cardiac dysrhythmia in patients with a prosthetic heart valve 8,16.Considerations
In
none of the incident databases reports were found to be relevant for this
guideline, neither we identified reported incidents in the literature. In the
Netherlands many patients with prosthetic heart valves, annuloplasty rings or mitra clips have been scanned over the years at 1.5 and 3T
systems without further limitations, in which the benefit of an MRI exam outweighed
the potential risks.
Risks
assessment for the heart valve implants was performed based on the standard
risks for metallic implants in the MRI:
- Risk of displacement and
rotation of the implant due to the presence of the static magnetic field and
the spatial gradient of this field. Forces on the few slightly ferromagnetic
implants are small compared to physiological forces 14. Limitations
set on spatial gradients by the manufacturers (Figure 1) were given by
limitation of the test situation, and not so much by observed complications.
- Risk of implant heating
due to interaction with RF field.
Manufactures give SAR limitations, see Figure 1. Heating is measured by in vitro tests, which typically report
heating instead of additional heating (Figure 2). Additional heating as
reported in the literature is less (Figure 3). More importantly, these implants
are effectively cooled by the blood flow which is not taken into account by the
risk assessment of the manufacturers 17.
- Risk of vibration or
induction of currents by the oscillating magnetic field gradients applied for the
spatial encoding of the MRI signal.
This is considered insignificant due to the
small surface of these implants.
- Artifact in the MRI
image.
This is typically limited to less than 1 cm
around the implant.
- Risk of forces due to the Lenz effect during rapid
movement of conductive implants in the static magnetic field of the MRI
scanner.
This effect is suggested to limit valve motion 18,19. With
reasons, see whole guideline, we think that this effect is insignificant in
clinical practice.
- Risk of Implant Disruption.
The only dysfunction could be caused by the
above mentioned Lenz effect.
Recommendation
A patient with a prosthetic
heart valve, annuloplasty ring or mitra clip can be scanned without further restrictions
with an 1.5T or 3T whole body MRI system with a horizontal closed bore
superconducting magnet.
The whole guideline can be
found in the FMS guideline database 20. An English
translation will appear early 2021. The guideline will be reviewed in five
years, or earlier when necessary.Acknowledgements
The guideline development was financed by the
Quality Funds for Medical Specialists (Stichting Kwaliteitsgelden Medisch
Specialisten: SKMS). We thank Diana P Gutierrez for secretary support of the
guideline project, and Evert J Vonken and Albert van der Zwan
for participating in the working group. References
1. Adviescommissie Richtlijnen van de Raad Kwaliteit. Medisch
Specialistische Richtlijnen 2.0; 2012. http://richtlijnendatabase.nl/over_deze_site/over_richtlijnontwikkeling.html
2. Brouwers M, Kho ME,
Browman GP, et al. on behalf of the AGREE Next Steps Consortium: AGREE II:
Advancing guideline development, reporting and evaluation in healthcare. CMAJ;
2010; www.agreetrust.org.
3. MagResource, MR:comp GmbH, Gelsenkirchen, Germany; www.MagResource.com
4. Edwards MB, Phil M, Ordidge RJ, et al. Translational and rotational
forces on heart valve prostheses subjected ex vivo to a 4.7-T MR System. J Magn
Reson Imaging 2002;16:653-659.
5. Edwards MB, Phil M, Taylor KM, Shellock FG. Prosthetic heart valves:
Evaluation of magnetic field interactions, heating, and artifacts at 1.5 T. J
Magn Reson Imaging 2000;12:363-369.
6. Edwards MB, Phil M, Ordidge RJ, et al. Assessment of magnetic field (4.7
T) induced forces on prosthetic heart valves and annuloplasty rings. J Magn Reson
Imaging 2005;22:311-317.
7. Pruefer D, Kalden P,
Schreiber W, et al. In vitro investigation of prosthetic heart
valves in magnetic resonance imaging: Evaluation of potential hazards. J Heart
Valve Dis. 2001; 10:410-4.
8. Randall PA, Kohman LJ, Scalzetti EM, et al. Magnetic Resonance Imaging
of prosthetic cardiac valves in vitro and in vivo. Am J Cardiol. 1988;62:973-976.
9. Saeedi M, Thomas A, Shellock FG. Evaluation of MRI issues at 3-Tesla for
a transcatheter aortic valve replacement (TAVR) bioprosthesis. Magn Reson
Imaging 2015;33:497-501.
10. Shellock FG. Prosthetic heart valves and annuloplasty rings: Assessment
of magnetic field interactions, heating, and artifacts at 1.5 Tesla. J Cardiovasc
Magn Reson. 2001;3,317-324.
11. Shellock FG. Biomedical implants and devices: assessment of magnetic
field interactions with a 3.0-Tesla MR system. J Magn Reson Imaging
2002;16:721-32.
12. Shellock FG and Crues JV. High-field-strength MR imaging and metallic
biomedical implants: An ex vivo evaluation of deflection forces. Am J
Roentgenol. 1988;151:389-92.
13. Shellock FG and Morisoli SM. Ex vivo evaluation of ferromagnetism,
heating, and artifacts produced by heart valve prostheses exposed to a 1.5-T MR
system. J Magn Reson Imaging 1994;4:756-758.
14. Soulen RL, Budinger TF, Higgins CB Magnetic resonance imaging of
prosthetic heart valves. Radiology 1985;154:705-707.
15. Edwards MB, Draper ERC, Phil M, et al. Young I.R. Mechanical testing of
human cardiac tissue: some implications for MRI safety. J Cardiovasc Magn
Reson. 2005;7:835–840.
16. Hartnell GG, Spence L, Hughes LA, et al. Safety of MR Imaging in
Patients Who Have Retained Metallic Materials After Cardiac Surgery. Am J
Roentgenol. 1997;168:1157-9.
17. Stijnman RS, van Nierop
BJ, et al. Assessment of the alleviating impact of
perfusion on RF-induced heating due to artificial cardiac valves. Annual
Meeting ISMRM 2019, 4154.
18. Condon B and Hadley DM. Potential MR hazard to patients with metallic heart
valves: the Lenz effect. J Magn Reson Imaging 2000;12:171-176.
19. Golestanirad L, Dlala E, Wright G, et al. Comprehensive analysis of Lenz
effect on the artificial heart valves during magnetic resonance imaging. Prog
Electromagn Research 2012:128,1-17.
20. Guideline database,
Federation Medical Specialists, the Netherlands. https://richtlijnendatabase.nl/richtlijn/gebruik_mri_bij_patienten_met_implantaten/startpagina_-_mri_bij_patienten_met_implantaten.html