Mubashir Hussain1 and Gregor Schaefers1,2
1Magnetic Resonance Institute for Safety, Technology and Research GmbH, Gelsenkirchen, Germany, 2MR:comp GmbH, Testing Services for MR Safety & Compatibility, Gelsenkirchen, Germany
Synopsis
This work focus on estimating the RF-induced deposited power at the
multi-electrode array of cortical implant (CorTec GmbH, Germany) by means of
lead electromagnetic models (LEM). To develop the LEM for such a complex
structure, we computed the transfer function for each of the electrode and
generated a set of artificial tangential electric fields for validation of the LEM. The transfer
functions and the tangential electric fields along the leads are then used to develop
the LEM which can estimate the deposited power at the electrode array for an arbitrary
lead pathway.
Introduction
Cortical implants significantly improves the quality of the life of the patients with neurological disorder by restoring or replacing certain circuitry of the brain which no longer functions properly. These implants, however, may cause thermal hotspots in the brain during an MRI examination and, therefore, require a thorough investigation concerning MRI safety and compatibility.
To
assess the RF-induced heating at a hotspot, the medical implants are often
simulated to observe their electromagnetic behavior under high frequency radio
waves. However, a complete 3D electromagnetic simulation of a sophisticated
implant with long leads is often too time consuming and require extensive
computing resources. In such cases, the transfer function approach along with
the lead electromagnetic model (LEM) is employed to estimate the magnitude of
the electric field around the electrodes. Further, the LEMs provide a
possibility to estimate the radio frequency (RF) induced heating of an implant
for an arbitrary trajectory of the implant’s geometry. Therefore, the objective
of this work is to compute the transfer functions and validate the lead
electromagnetic models for the cortical implant developed by CorTec (CorTec GmbH,
Germany) with 1x4 multi-electrode arrays. The developed LEM would be able to
predict the net incident electric field around electrodes for any potential
lead pathway.
The lead
electromagnetic models as described in ISO/TS 109741 comprise of a transfer function of the implant,
tangential electric field along the lead pathway and a constant. The LEM is
given by
$$~~~~~~~~~~P_{lem} = A \left| \int_0^l S(l) E_{tan}(l)dl \right|^2. ~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~(1)$$
Here, $$$S(l)$$$ is the transfer function, $$$E_{tan}(l)$$$ is the tangential
electric field along the lead pathway, $$$A$$$ is the calibration factor, and $$$l$$$ is the length of the
lead. Moreover, the RF-induced net
incident power in a tissue due to the presence of a medical implant can be
computed using direct simulations as2
$$~~~~~~P = \int^V \sigma \left( \left| E_{lead}(v) \right|^2 - \left| E(v) \right|^2 \right) dv, ~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~(2)$$
where, $$$E_{lead}(v)$$$ and $$$E(v)$$$ are the electric fields with and without the
lead, respectively, $$$\sigma$$$ is the conductivity of the surrounding medium,
and $$$V$$$ is the hotspot integrating volume (HSIV). Methods
The CorTec's cortical implant considered in this
study consists of a 45cm long lead which is composed of four copper wires. Each
wire has a diameter of 0.15mm without insulation and 0.17mm with insulation and
is connected to a zig-zag structured electrode (Fig. 1). All four wires are
bound together with a large silicone insulation of 1mm diameter and the whole
lead is further covered with silicone insulation of dimeter 1.4mm. The transfer functions of all four leads are
computed by using reciprocity method3. Further, the transfer
functions are computed using ANSYS HFSS (ANSYS Electromagnetics 2020 R1) at
resonance frequency of 64MHz with leads being inserted in gel medium with
physical properties defined as per ASTM standard4.
To
validate the LEM, an artificial set of tangential electric fields are generated
by four power sources located on one side of the gel box. All other walls of
the box are taken to be the perfect matching layer boundaries. A set of 40
artificial electric fields are generated by randomly choosing magnitude and
phase of the power sources (Fig. 2). Once an electric field is generated, its
tangential component ($$$E_{tan}(l)$$$)
is computed along the lead pathway. This tangential field along the lead and
the transfer function are then substituted in Eq. (1) to compute the LEM
response without the constant A as,
$$~~~~~x_{i} = \left| \int_0^l S(l) E_{tan,i}(l)dl \right|^2, i \in \{1,2, \cdots, 40 \} ~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~ (3)$$
Further, the incident net power in the HSIV is
calculated directly from electromagnetic simulations using Eq. (2) which is then used in linear
regression along with Eq. (3) to calibrate the constant A. Results and Discussion
We computed the transfer function and validated the
LEM for all four electrodes. Fig. (3) shows the magnitude and phase of the
transfer functions computed for the electrodes 1 and 4. $$$S(l)$$$ shows a significant dependence of the position of the
electrode with respect to the other electrodes. This means that induced current
in the neighboring wires and electrodes have an effect on the RF response
of the investigated electrode. Further, since the distance among the wires are very small and remains constant along the lead pathway, the transfer functions are computed by estimating the total current flowing in all wires rather than computing current in the single wire whose electrode is excited by the current source5.
The
linear regression fit gives a reasonable match for all the lead electromagnetic
models. Fig. (4) shows the linear fit for the LEM of electrode 1. Some
numerical noise is expected because of the complex structure of the electrodes
and the approximations made in numerical integration. The linear fit, however,
catches the overall trend in the increasing behavior of LEM well.Conclusion
The developed lead electromagnetic model is
validated using linear regression approach and is generic in nature. The model can be used on arbitrary lead
pathway to estimate the RF-induced heating due to the cortical implant with 1x4
multi-electrode array.Acknowledgements
We acknowledge the financial support provided by BMBF
(Federal Ministry of Education and Research, Germany) for this project (Lightbridge,
Grant 16ES0787).References
[1]. ISO/TS 10974: Assessment of the safety of magnetic resonance imaging for patients with an active implantable medical device," International Organization for Standardization, Geneva, Switzerland; 2018.
[2]. M.
Kozlov, M. Horner and W. Kainzs. Modeling radiofrequency responses
of realistic multi‑electrode leads containing helical and straight
wires, Magnetic Resonance Materials in
Physics, Biology and Medicine, 33(3): 421-437, 2020.
[3]. S.
Feng, R. Qiang, W. Kainz, and J. Chen. A technique to evaluate mri-induced
electric fields at the ends of practical implanted lead. IEEE Transactions on
microwave theory and techniques, 63:305–313, 2015.
[4]. ASTM
standard F 2182-2011a, 2011. ASTM 2182 Standard test method for measurement of
radio frequency induced heating near passive implants during magnetic resonance
imaging. ASTM Int. 1–14
[5]. J.
Kabil, J. Felblinger, P. Vuissoz and A. Missoffe. Coupled transfer function
model for the evaluation of implanted cables safety in MRI, Magnetic Resonance
in Medicine, 00:1-9, 2020