Finya Ketelsen1,2, Kevin Kröninger2, and Gregor Schaefers1,3
1MRI-STaR - Magnetic Resonance Institute for Safety, Technology and Research GmbH, Gelsenkirchen, Germany, 2TU Dortmund University, Dortmund, Germany, 3MR:comp GmbH, Testing Services for MR Safety & Compatibility, Gelsenkirchen, Germany
Synopsis
To
determine the RF-induced heating of implants, it is necessary, that the RF
field source used for testing fulfills several requirements. This study
introduced the validation of a new developed linear exposure system for testing.
The results for E-Field distributions from numerical simulations are validated experimentally
by measuring these E-field distributions. Furthermore, the temperature rise at
the 100mm titanium rod is compared with the target values form ISO/TS 10974
Annex I.
The system fulfills the requirements and is suitable for testing active and
passive implants for RF-induced heating according to ASTM F2182 and ISO/TS
10974.
Introduction
One
of the safety issues for patients with implants during an MRI examination is RF-induced
heating. For testing these implants, an appropriate RF-source, producing an
electromagnetic field, is necessary. Typically, a MR scanner or a stand-alone RF
birdcage coil system is used.
This study introduces an alternative RF field source and its validation for
testing all kinds of implants for RF-induced heating.Methods
Requirements to a RF-field source are described in ISO/TS 109471 and ASTM-F21822 for active and passive implants. Incident field variation
over the entire implant volume should be less than ±1dB. Ideal exposure
configuration provides nearly uniform tangential electric field incident along
the test object1.
Numerical simulations with Ansys HFSS were used to develop a Linear Exposure
System (LES). This system consists of two independent
channels and is attached to the walls of an ASTM phantom (420mm x 90mm x 650mm).
Erms- and Etan,z-field (in z direction) distributions were
simulated for the xz-plane at central height (45mm) with 10mm x 5mm spatial resolution
and 20mm distance to walls. This was done for two different phases 0° and 180°
between channels.
After construction, the simulated values were measured experimentally with an Erms
probe (SAR Probe, EX3DV4, SPEAG). Input power 45dBm, phantom dimensions and
electrical properties (σ=0.47 S/m) of saline solution were identical to the
simulations.
In a second measurement a 100mm titanium rod (SAIMD-21,2)
is centrally placed in the phantom parallel to the z-axis. At this location, the
field is homogeneous and due to large distance from the wall, reflections
effects are avoided3.
Four fiber optic temperature probes P1-P4 were used to measure the temperature
rise (cf. Fig. 1). P1 and P2 were placed at one axial rod tip and P3 inside the
hole on the same side. P4 was placed axial at the opposite tip. Seven 360s test
runs were performed. After each test run P4 was moved distally away from the
rod (0.5 to 5mm). Incident field was calculated from the temperature rise of an
additional test run, where the rod was removed but the temperature probes
remained in positions.
Temperature rise at each probe and position is normalized to the square of
incident E-field. These values are compared to the target values for the
SAIMD-2 in Annex I2 from an ideal plane wave simulation.Results
Simulated
Erms-Field distributions are shown in Fig. 2a/3a, Etan,z-field
distributions in Fig. 2b/3b and deviations in Fig. 2c/3c for 0°/180°,
respectively.
Erms-field distribution varies over phase relations. At 0° the Erms-field
is homogeneous over a large area in the phantom middle. Etan,z-field
distribution shows the same behavior. Deviation between Erms- and Etan,z-field
for 0° is less than 1% for almost the entire plane. Thus, Erms-field
equals Etan,z-field. E-field variation between the 80V/m and 100V/m
isolines is inside ±1dB.
Erms-field distribution for 180° varies substantially from 0°. The
distribution is homogeneous along elliptical lines and decreases towards the phantom
middle. Etan,z-distribution is homogeneous along z-axis along the
entire phantom. The deviation between Erms and Etan,z
varies over the plane. There is a larger area, where deviation is less 1%, but it
increases in x-direction.
The
measured Erms-field distributions are shown in Fig 4a/c and the deviation
between simulation and measurement is shown in Figure 4b/d for both phase
relations 0°/180°.
The measured Erms-field distributions are similar to the simulated
distributions for both phases. Fig. 4b shows a deviation for 0° of ±5% over the entire plane except the corners. For
180°, Fig. 4d shows a larger deviation, which is still mostly inside ±5%. In corners and at the isocenter, deviation
increases to ±10%.
For
temperature measurements the incident E-field is 162.51±5.5 V/m. Temperature rise at P4 is divided by
the square of incident field at each position. These values and target values
exported from Annex I1 are shown in Fig. 5.
The
target value for the rod hole is 0.621 mK/(V/m)2 and the measured value
at P3 0.629±0.008
mK/(V/m)2. For P1/P2 the measured value is 0.652±0.012 mK/(V/m)2 and target value at
0.5mm is 0.66 mK/(V/m)2.Discussion
The
results in Fig. 2 and Fig. 3 show, that E-field distribution varies with phase.
The Erms-values are equal to Etan,z-values over most parts
of the phantom. The measurements confirm and validate the simulation results
for the Erms-field distribution within deviation of ±5% and therefore also Etan,z-distribution.
In addition, the results in Fig. 5 show that the new exposure system fulfills
the requirements for testing implants according to 1,2.Conclusion
This
study shows that the new RF-field source is able to produce different uniform
tangential E-field distributions over a large area. The simulated E-field
distributions are validated by measurements and the temperature rise at the rod
reaches the target values. Due to almost no field variation, it is suitable for
implants with small diameter and long wires as well as orthopedic implants with
large 3D volumes.Acknowledgements
No acknowledgement found.References
1. Technical specification ISO/TS 10974:2018,
“Assessment of the safety of magnetic resonance imaging for patients with an
active implantable medical device,” The International Organization for
Standardization, 2018.
2. ASTM F2182-19e2, “Standard Test Method
for Measurement of Radio Frequency Induced Heating On or Near Passive Implants
During Magnetic Resonance Imaging”, ASTM international, 2019.
3. M. Kozlov, M. Angelone, “Effect of Multiple
Scattering on Heating Induced by Radio Frequency Energy”, IEEE T ELECTROMAGN C,
Vol 62, No.5, October 2020.