Kyoko Fujimoto1, Tayeb A Zaidi1, Dave Lampman2, Josha W Guag1, Shawn Etheridge2, Hideta Habara3, and Sunder S Rajan1
1Center for Devices and Radiological Health, US Food and Drug Administration, Silver Spring, MD, United States, 2Hitachi Healthcare Americas, Twinsburg, OH, United States, 3Healthcare Business Unit, Hitachi, Ltd., Tokyo, Japan
Synopsis
Radio-Frequency (RF)
induced thermal injury is a common adverse event reported to US Food and Drug
Administration. RF-induced
heating risk depends on the interaction of implantable medical devices with
electric field distribution in the MRI systems. A study showed that RF-induced
heating of a neuromodulation device was much lower in the planar system
compared to the cylindrical system. In this study, the hip and knee implants
were studied in an anatomical human model at a 1.2T planar system and a 1.5T
cylindrical system.
INTRODUCTION
Radio-Frequency (RF)
induced thermal injury is a common adverse event reported to US Food and Drug
Administration1. RF-induced heating risk
depends on the interaction of implantable devices with electric field (E-field)
in the MRI systems. Therefore, the RF safety risk of implants may differ when
different coil designs are used. A recent study showed a significantly lower RF
heating due to a neuromodulation device in the planar system2. In
this study, the hip and knee implantable devices were examined using validated
computational modeling. Specific absorption rate (SAR) results were compared in
an anatomical human model between a 1.5T cylindrical system (64MHz) and a 1.2T planar
system (49.5MHz). METHODS: Modeling Validation
The MITS
1.5 64MHz birdcage coil (ZMT, Zurich, Switzerland) and the open-bore OASIS 1.2T
MRI scanner (Hitachi Ltd) were used for measuring E-field and magnetic field
(H-field) to validate the electromagnetic modeling setups. ASTM phantom was prepared
with saline with a conductivity of 0.27S/m. Fifteen data points of E-field and
magnetic field (H-field) were collected in the ASTM phantom using a hand-held
measurement system with an E-field probe and an H-field probe (EASY4MRI, SPEAG,
Switzerland) in both systems. The probe performance was confirmed using a
transverse electromagnetic cell (Model 8802, Narda, USA). The corresponding
models were prepared based on the physical coils3,4, and the coil
and phantom setups are shown in Fig.1.
The
simulated values were scaled by the ratio of the average H-field values between
simulation and measurement.
$$H'_{S}=H_{S}(\frac{\sum H_{M}}{\sum H_{S}})$$
$$E'_{S}=E_{S}(\frac{\sum H_{M}}{\sum H_{S}})$$
HS
and HM are the simulated and measured H-field values, and ES is
the simulated E-field value. The H'S and E'S values were used
to find correlation between the measured and simulated E- and H- field values.METHODS: Modeling with Implants
The
generic models of hip and knee implantable devices were used along with the
AustinMan body5. The effective length of the hip implant was 33.4cm,
and the socket was 3.2cm in diameter. The effective length of the knee implant
was 18.6cm. Both were modeled as titanium. All the modeling was done in the
Sim4Life platform (V5.2.2, Zurich Med Tech, Switzerland). The Huygens’ Approach6
was used to calculate the incident field with an unloaded coil first, then used
to compute fields within the body model with 2mm-isotropic grids. Two scenarios
were simulated for each implant: (a) the imaging landmark at the center of the
body and the hip or knee level and (b)
the imaging landmark at the center of the implant (resulting in a 10cm side
shift of (a)).
The SAR results
were further analyzed in MATLAB (Mathworks, MA, USA). The maximum 1g- and
10g-averaged SAR values (SAR1g and SAR10g) were
calculated by using two normalization methods: whole-body SAR=2W/kg and a slice
of B1+=2µT. The B1+ slice was determined in axial slice
6cm away from the iso-center, so the device is not included.
RESULTS: Measurement
The scaled
measurement results were compared with the simulation results (Fig.2). The
correlation (R2) of the E-fields was 0.86 for the planar coil and
0.98 for the cylindrical coil. The correlation of H-fields was 0.85 for the planar
coil and 0.76 for the cylindrical coil. RESULTS: Modeling
The
maximum SAR1g and SAR10g results are shown in Fig.3. None
of the maximum values were found near the implants. The maximum SAR values were
observed at the body parts close to the coils, i.e. in the arms and legs for
the cylindrical coil and in the back for the planar coil. The slice SAR1g
and SAR10g maps at the implant locations showed that the presence of
the implants resulted in a discontinuity in E-fields (Fig.4). Overall, the SAR
values around the implants were much lower for the planar system. The maximum
intensity projection maps of SAR1g
showed the SAR values in the planar system were up to 75% lower than those in
the cylindrical system (Fig.5).DISCUSSION
The
potential RF heating induced by the two implant models used in this study were
relatively low in both systems; however, the landmark differences revealed the
high SAR exposure concern of the other side of the body when the imaging
landmark is on the implant in the cylindrical system. The results suggest that
the planar system may be more suitable especially for the clinical cases that
patients have knee or hip implants on both sides of the body.
The
maximum SAR values demonstrated with two different normalization methods
revealed that B1+ scaling for cylindrical system showed about 50%
lower SAR with the iso-center at the center of the body and about 15% lower SAR
with the iso-center at the implant. For the planar system, the maximum SAR
normalized with B1+ scaling was up to 60% lower for the knee
landmarks and 40% lower for the hip landmarks, compared to the SAR values
normalized with the 2W/kg limit. The use of B1+=2μT as a safety limit
may allow more patients to be scanned with MRI systems. CONCLUSION
The validated
computational modeling offers a viable approach to compare the potential risk
of RF-induced heating of implants. The 1.2T planar system
demonstrated lower risk of heating of hip and knee implants compared to the
1.5T system. Having different coil designs may improve patient access to MRI
scans. Acknowledgements
DISCLAIMER: The mention
of commercial products, their sources, or their use in connection with material
reported herein is not to be construed as either an actual or implied
endorsement of such products by the Department of Health and Human Services.References
1.
Delfino JG,
Krainak DM, Flesher SA, et al. MRI-related FDA adverse event reports: A 10-yr
review. Medical Physics. 2019;46(12):5562–5571.
2.
Golestanirad L,
Kazemivalipour E, Lampman D, et al. RF heating of deep brain stimulation
implants in open-bore vertical MRI systems: A simulation study with realistic
device configurations. Magn. Reson. Med. 2019;001-9.
3.
S. Suzuki, T.
Shimoda, and T. Taniguchi, “High frequency coil for magnetic resonance imaging
device https://patentscope.wipo.int/search/en/detail.jsf?docId=WO2008108048,”
2009.
4.
Lucano E, Liberti
M, Mendoza GG, et al. Assessing the electromagnetic fields generated by a
radiofrequency mri body coil at 64 mhz: Defeaturing versus accuracy. IEEE
Transactions on Biomedical Engineering, 2016;(63):1591–1601
5.
Massey JW and
Yilmaz AE. AustinMan and AustinWoman: High-Fidelity, anatomical voxel models
developed from the VHP color images. Conf Proc IEEE Eng Med Biol Soc 2016:
3346-3349.
6.
Benkler, S.,
Chavannes, N., Kuster, N. Novel FDTD Huygens source enables highly complex
simulation scenarios on ordinary PCs. Conf Proc IEEE APSURSI, 2009: 1-4.