David Prutchi1, Jason Meyers1, and Ramez Shehada2
1Impulse Dynamics (USA) Inc., Marlton, NJ, United States, 2Medical Technology Laboratories, La Mirada, CA, United States
Synopsis
This study investigates the change in the RF filtering
characteristics of the leads of active implantable medical devices (AIMDs) as
body fluids seep into the leads during the initial post-implant period. Our findings indicate that the RF
characteristics change dramatically with fluid absorption, making it necessary
to precondition the leads by soaking in isotonic saline solution to simulate
the in-vivo scenario when conducting in-vitro MR safety testing.
Furthermore, leads designed with RF-attenuating lumped inductances must
consider the effect of fluid absorption on changing the peak RF attenuation
frequency.
Introduction
The RF currents induced into the lead of an active implantable
medical device (AIMD) are transmitted through the lead body to reach its
electrodes and cause heating at the tissue interface. Accordingly, the heating of the tissue will
depend on the transmission line characteristics of the lead body and its
ability of conducting the RF currents to the electrodes. The
characteristic impedance refers to the equivalent resistance of a transmission
line, owing to distributed capacitance and inductance as the voltage and
current waves propagate along its length.
Following implantation, conductive body fluids seep into the body of
the lead to replace the air between the lead conductors thereby changing the
characteristic impedance (Zo) of the lead at the RF
frequencies of the MRI environment. In
this study we investigated the post implantation change in Zo
as the body fluids replace the air in-between the lead conductors.Methods and Results
To demonstrate the effect of liquid medium absorption by
pacemaker leads on their transmission-line characteristics, we measured the
open- and short-terminated impedances of a number of current MR-conditional
lead samples before and after soaking them for 10 days in isotonic saline
solution (0.9% NaCl) at 37°C. Results using an Anritsu MS46522A, 8.5GHz bandwidth
vector network analyzer (VNA) are shown in Table
1. The characteristic impedance Z0
of the lead can be calculated from measurements of the open (unterminated) lead
impedance Zopen and the short-circuit-terminated impedance Zshort:
$$Z_0 = \sqrt{Z_{open}*Z_{short}}$$
We also measured the complex impedance of the leads in the
ASTM F2182-191 thorax phantom filled with gel slurry of 1.32 g/l
NaCl and 10g/l partial sodium salt of polyacrylic acid (PAA) with a measured
conductivity of 4 mS/cm at 30°C.
Figure 1 presents the magnitude and phase of the Biotronik
and Abbott leads, while Figure 2 presents those for the Boston Scientific
leads.Discussion
In all pacemaker leads studied, the characteristic impedance
at 63.87 MHz dropped after soaking, with an average of change of 32.99%.
Figure 3 shows the Smith chart for the Biotronik Solia-S 60
lead, which presents a typical behavior.
The impedance changes approximately along a constant X circle (in Z=R+jX),
suggesting a change in the resistive properties of the lead as it is soaked.
Another interesting observation is that in the measurements
and the graphs of Figure 2, The Boston Scientific Ingevity+ leads clearly
present as outliers. This is because the outer coil in these leads is made of
ETFE-insulated filars2, thus increasing the outer coil’s inductance. As the leads soak, their impedance decreases
together with an increase in the frequency at which the impedance peaks. Presumably, the lead is designed to stabilize
such that the peak would occur in the proximity of the 63.87 MHz 1.5 T MRI
frequency. However, the conditions
necessary for this seem prone to be sensitive on lead aging. Conclusion
The results above indicate that it is critical to
precondition the leads by soaking them in an isotonic saline solution to simulate
the in-vivo scenario prior to conducting MR safety testing in-vitro. This step is necessary to avoid misleading
safety results as the diffusing body fluids change the peak impedance of the
lead. Furthermore, the design of MR
Conditional leads with RF-filtering lumped inductances must consider the effect
of post-implantation fluid diffusion in shifting the peak RF attenuation frequency
of the lead.Acknowledgements
This work was supported by Impulse Dynamics (USA) Inc.References
- ASTM F2182-19, Standard Test Method for
Measurement of Radio Frequency Induced Heating On or Near Passive Implants
During Magnetic Resonance Imaging, 2019.
- Boston Scientific, INGEVITY™+ Pacing Lead -
Active Fixation Models: 7840, 7841, 7842 – Specification Sheet, CRM-699008-AA,
2020.