David Prutchi1, Jason Meyers1, and Ramez Shehada2
1Impulse Dynamics (USA) Inc., Marlton, NJ, United States, 2Medical Technology Laboratories, La Mirada, CA, United States
Synopsis
RF-induced heating of an active implantable medical device
(AIMD) composed of a pulse generator (IPG) and leads depends on the transmission
line impedance of the lead and its proximal-end termination by the impedance of
the IPG. We demonstrate that the RF impedance of IPGs
is minimal relative to that of the leads, which dominates the overall impedance
of the implantable system. Accordingly, mixed hybrid systems composed of
MR-Conditional leads and any MR-conditional IPG are expected to have a comparable
overall impedance and consequently produce the same RF-induced heating as their
corresponding original systems specified by the manufacturers.
Introduction
MR-conditional active implantable medical devices (AIMDs)
composed of an implantable pulse generator (IPG) and leads are deemed to be MRI-safe only when used in specific IPG-leads combinations that have been jointly
tested for RF-induced heating by their manufacturer. This is because the RF-induced heating at the
lead electrodes depends on the transmission line impedance characteristics of
the lead and its proximal-end termination by the impedance of the IPG. The RF impedance characteristics of the
IPG-leads combination determines how induced RF currents get transmitted to the
tip of the lead to cause heating at the electrode-tissue interface. Accordingly, mixing MR-conditional IPGs and
MR-conditional leads from different manufacturers do not guarantee the MR-conditional
safety of their combination as an implantable system. The current regulatory stance is that an
implanted system comprised of the IPG of one manufacturer does not fulfill the
requirements of MR-conditional labeling if used with the leads from a different
manufacturer. In this study we measured
and compared the RF impedances of various MR-conditional IPGs and leads from
different manufacturers to determine the impact of mixing these components on
the overall RF impedance and consequently assess the RF safety of the hybrid
implantable system.Methods
The RF input impedance for several MR-conditional cardiac leads
and IPGs were measured and compared to evaluate their relative contribution to
the overall RF impedance of potential hybrid systems that can be formed from mixed
combination these leads and IPGs. The
RF-induced heating safety of the aforementioned hybrid systems was assessed by
comparing their RF impedances to that of corresponding original systems
specified by the manufacturers.Results and Discussion
Table 1 and Figure 1 compare the contribution of the input
impedance of various MR-conditional IPGs against the total lead/IPG impedance
to 63.87 MHz RF currents at the Tip (distal) electrode when paired with lead models that
are MR-conditional labeled for whole-body MRI at 1.5T.
For these comparisons, lead impedance data which
incorporates the effect of the implantable device’s enclosure in gel slurry1
was used since it represents the worst-case condition. As shown, the lead is overwhelmingly
responsible for limiting 63.87 MHz RF currents from flowing through the Tip
electrode in an implanted system. The
IPG’s contribution to the limitation of RF currents is relatively small (range
of 1.03 to 8.04%, mean 2.72%).
Table 2 presents the worst-case IPG
contributions to the total lead/IPG impedance to 63.87 MHz RF currents at the
lead’s Tip electrode with different IPG/lead combinations. Whenever two IPGs were evaluated for the same
leads, the lowest IPG input impedance was used to be representative of the
worst-case condition.
The minimum contribution (0.98%) to total impedance happens
when the Medtronic CapsureFix 5076 52 cm lead is connected to the Biotronik
Iperia IPG. However, the contribution of
the Medtronic Visia IPG with the same lead in a configuration labeled as
MR-conditional is 1.03%, which is a difference of barely -4.85%. This is probably negligible when compared to
possible changes in lead impedance as a function of water content, medium
conductivity, aging, etc.Conclusions
For all the IPGs and leads studied, the IPG input impedance
is relatively negligible compared to the lead’s impedance and hence can be
considered as a short circuit at the 1.5 T MRI frequency of 63.87 MHz. Because the input
impedances of the IPGs are all relatively low, a system composed of any of the
studied IPGs and any of the transvenous leads approved as MR-conditional will
have very similar RF impedance to a system composed of an MR-conditional
labeled IPG and the same lead, and therefore both systems are expected to
exhibit similar RF-induced electrode heating behavior.Acknowledgements
This work was supported by Impulse Dynamics (USA) Inc.References
- Meyers J, Prutchi D, Shehada R, Input Impedance
Comparison of MR-Conditional Cardiac Implantable Pulse Generators at the 1.5T MR
Frequency of 63.87 MHz , Submitted to ISMRM 2021
- Prutchi D, Meyers J, Shehada R, Impedance of
MR-Conditional Pacemaker Leads at the 1.5T MR Frequency of 63.87 MHz, Submitted
to ISMRM 2021