Amol Pednekar1, Deep B. Gandhi2, Hui Wang3, Jean A. Tkach1, Andrew T. Trout1, and Jonathan R. Dillman1
1Department of Radiology, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, United States, 2Imaging Research Center, Department of Radiology, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, United States, 3MR Clinical Science, Philips, Cincinnati, OH, United States
Synopsis
2D
GRE MRE liver images acquired at 4 transverse levels in breath-hold times
>13s per slice is currently standard of care (SC_2D_4BH). Inadequate
breath-holds lead to inaccurate stiffness estimation and/or failed studies. The
combination of wave polarity-inversion motion encoding and compressed-SENSE enables
MRE images to be acquired in less than half the breath-hold time (2D_CS_HBH) e.g.
<7s, with identical spatial resolution and field of view. In 19 participants,
mean liver shear stiffness values estimated with SC_2D_4BH and 2D_CS_HBH correlated
very strongly (ICC>0.96) with a bias of <0.15 kPa (<6%). 2D_CS_HBH MRE is beneficial in participants with
compromised breath-holding capacity.
Introduction
Magnetic
resonance elastography of the liver (MREL) has been shown to aid in the
non-invasive diagnosis and staging of liver fibrosis [1]. Two-dimensional (2D)
gradient-recalled-echo (GRE) MRE employing: 1) a mechanical driver frequency of
60 Hz, 2) motion encoding gradients (MEG) applied along a single direction, and
3) inflow saturation, acquired during breath-hold (BH) is currently the
standard of care (SC) sequence for the assessment of liver fibrosis [2]. Four
transverse MREL images through mid-liver are typically acquired in 4
consecutive breath-holds of ~14 – 18sec (SC_2D_4BH) [3]. Given the motion sensitive nature of MREL,
inadequate breath-holds lead to inaccurate stiffness estimation and/or failed
studies. MREL images can be obtained in less than half the breath-hold time
(2D_CS_HBH) e.g. <7s, with identical spatial resolution and field of view by
setting the repetition time to 1.5 the period of applied mechanical frequency
to encode motion using wave polarity
inversion, plus data under-sampling using compressed Sensitivity Encoding
(C-SENSE) [4].
The
purpose of this study is to compare liver stiffness measurements obtained using
SC_2D_4BH technique to those obtained by the 2D_CS_HBH.Methods
In this prospective HIPAA compliant
IRB approved study, 19 participants (age 36.9±12.6 years, range 20-62 years, 8
males) underwent MREL imaging with written informed consent. All imaging was
performed on a Philips Ingenia 1.5T scanner (Best, The Netherlands), using 28-channel
torso coil. MREL
images were obtained using both the commercially available SC_2D_4BH MREL
sequence (Fig. 1 A) and 2D_CS_HBH MREL sequence with (Fig. 1 B) and without
inflow saturation (Fig. 1 C). Commercially available SENSE (based on regular
undersampling and coil sensitivity information, and spatial solution space
constraint based on prior knowledge of the image extent) and C-SENSE (combines
spatial domain variable-density pseudorandom undersampling of k-space with the
SENSE reconstruction algorithm using iterative reconstruction and sparsity
constraints) were employed [5]. All image reconstructions and stiffness map (using
multimodal direct inversion algorithm with 95% confidence interval mask
overlays [6]) computations were performed inline in real-time available on the
scanner.
All
the MREL data, including magnitude, phase, and wave images as well as shear
stiffness maps, were exported to a post-processing workstation (with IntelliSpace
Portal v10.1; ISP, Philips). Manual liver shear stiffness measurements were
made by a single trained image analyst (>1-year experience), under the
supervision of a board-certified Pediatric Radiologist (>10-years
post-fellowship experience). On four slices from each acquisition, a single
freehand region of interest (ROI) was drawn to encompass as much of the right
hepatic lobe as possible while remaining within the boundaries of the 95%
confidence mask and avoiding the liver capsule, large blood vessels, dilated
bile ducts, and areas of artifact, including hot-spots immediately under the
passive driver and respiratory motion in cases of failed breath holds (Fig. 2).
Additionally, to compare the shear stiffness values in the same tissue across
the techniques, identically sized and positioned ROIs were drawn on both sets
of shear stiffness maps. Overall participant-specific liver shear stiffness was
calculated as the ROI area weighted mean liver shear stiffness of the estimates
extracted from each slice. Two-sided paired t test, intra-class correlation coefficient (ICC), , and Bland-Altman analysis were used to compare and assess
agreement in the ROI sizes and shear stiffness estimates between the sequences.Results
MREL image acquisition and
reconstruction was performed successfully in all the participants. Breath-hold
times were 13.3 ± 0.2 s per slice with SC_2D_4BH MRE, 6.3 ± 0.2 s with
interleaved flow saturation, and 4.9 ± 0.2 s without flow saturation for 2D_CS_HBH
techniques. Descriptive and comparative statistics of the liver stiffness
values and area within 95% confidence mask are presented in Tables (1-2) and
Fig. (3). Area weighted mean liver stiffness estimates for identical tissue
ROI for the SC_2D_4BH (2.17 ± 0.68 kPa,
range 1.60 – 4.53 kPa) and 2D_CS_HBH with flow saturation (2.04 ± 0.66 kPa,
range 1.46 – 4.39 kPa), and without flow saturation (2.03 ± 0.61 kPa, range 1.40
– 4.15 kPa), MRE sequences correlated
very strongly (ICC>0.96).Discussion
Liver shear stiffness values obtained by 2D_CS_HBH, with or
without flow saturation, correlate very strongly (ICC > 0.96) with stiffness
values obtained by SC_2D_4BH,
underestimating stiffness significantly (p<0.03) with bias <0.15 kPa
(<6%). While all the techniques had identical acquired spatial resolution
and coverage, the difference in total ROI area across 4 slices was less than 17
cm2. Liver stiffness values estimated by 2D_CS_HBH with or without
flow saturation were comparable (p>0.07). 2D_CS_HBH
MREL has the potential to markedly reduce breath hold times, allowing liver stiffness mapping in participants who are only able to perform consistent 5 to
6 second breath-holds. Alternatively, in participants who are able to perform consistent
10 to 12 second breath-hold, the 2D_CS_HBH
technique enables either reduced number of
breath-holds, or increased coverage relative to the standard 2D_SC_4BH MREL approach.Conclusion
Liver stiffness values measured in
less than half (<5-7 s) the standard (>13 s) breath-hold time per slice
achieved by the combined use of wave polarity-inversion motion encoding and
C-SENSE (2D_CS_HBH) were comparable to liver stiffness values estimated by the
standard of care MREL sequence. 2D-CS_HBH MREL has potential clinical benefits
in participants with limited and/or compromised breath-holding capabilities.Acknowledgements
No acknowledgement found.References
[1]
Singh S et al. Diagnostic performance of magnetic resonance elastography
in staging liver fibrosis: a systematic review and meta-analysis of individual
participant data. Clinical gastroenterology and hepatology, 2015: 13(3),
440–451.
[2]
Venkatesh SK. et al. Hepatic MR elastography: Clinical performance in a
series of 1377 consecutive examinations. Radiology, Magnetic resonance
elastography of liver: technique, analysis, and clinical applications. J
Magn Reson Imaging. 2013: 37, 544–555.
[3]
Yin, M. et al. Hepatic MR elastography: Clinical performance in a series
of 1377 consecutive examinations. Radiology, 2016: 278, 114–124.
[4]
Wang H. et al. Fusing acceleration and saturation techniques with wave
amplitude labeling of time-shifted zeniths MR elastography. Magn Reson in Med. 2021: 85(3), 1552-1560, doi:
10.1002/mrm.28488.
[5]
Geerts-Ossevoort L. et al. A. Compressed SENSE. Speed done right. Every
time. Philips Field Strength Magazine, 2018: 278, 6619.
[6] Silva AM.
et al. A. Magnetic resonance elastography: Evaluation of new inversion
algorithm and quantitative analysis method. Abdom Imaging, 2015: 40,
810-817.