Erik Huijing1, Evita Wiegers1, Dennis Klomp1, Fredy Visser1,2, Edwin Versteeg1, Koenraad Rhebergen3, Kim Annink4, Niek van der Aa4, Floris Groenendaal4, Jeroen Dudink4, Thomas Alderliesten4, Maarten Lequin1, Manon Benders4, and Jannie Wijnen1
1Radiology, University Medical Center Utrecht, Utrecht, Netherlands, 2Philips HealthCare, Best, Netherlands, 3Otorhinolaryngology and Head & Neck Surgery, University Medical Center Utrecht, Utrecht, Netherlands, 4Neonatology, University Medical Center Utrecht, Utrecht, Netherlands
Synopsis
We designed a new acoustic hood for neonatal MRI exams at 7 Tesla. The
new design is lightweight and allows easy access to the neonate during the MRI
examination. We were able to optimize the new acoustic hood to a noise
reduction of 10.6dB, exceeding the damping of a prototype with damping of
3.1dB. The new hood was designed to fit a head coil and allows the use of a
ventilation tube.
Introduction
Subjects are often exposed to loud acoustic noise
during an MRI exam. This is a well-known cause of discomfort and could
potentially lead to hearing damage if protection devices (earplugs/earmuffs/headphones)
are not properly used. In neonates, acoustic noise imposes another risk for
waking up an infant while scanning, which will most likely lead to a failed
examination. Previous research (1)(2) has shown that the acoustic hood may
provide additional hearing protection next to headphones, ear muffs or other
noise attenuators.
In this work, we optimized
the design of an acoustic hood, dedicated to neonatal MRI. We
compare this final product with a prototype that had been used over the past
year (fig. 1a).Methods
The requirements for the neonatal acoustic
hood were:
- The NOVA Medical (2TX 32RX) RF
head coil must be placed inside the hood
- Patient ventilation must be
sufficient
- The hood must fit on the
table of a 7T Achieva MRI (Philips, Best, The Netherlands)
- In case of an emergency there
should be easy access to the patient
- The new product should be
easy to handle by a single person
- The new hood should have the
same or more noise attenuation as the prototype
- The hood should not generate
1H signal during imaging
Design
The final product was
designed in SolidWorks (Dassault Systèmes). It was built with TC-2 sound
damping foam (supplied by Easy Noise Control BV) and Polycarbonate arches
(supplied by WSV Kunststoffen).
The final product (fig. 1b) consists
out of four parts: three parts hood and one mattress. The mattress, made of
sound damping foam, was designed to fit exactly on the 7T table. The hood
parts, also made of sound damping foam, contain a plastic plate in the middle. The
two parts at the feet end and middle part of the bed are arch-shaped. The
middle part is designed such that easy access to the neonate is possible
(fig.1c).
The part on the head end of
the table was designed to closely fit around the NOVA head coil and has its
backside closed off as much as possible. It only contains openings for cables
and a ventilation tube.
The presence of
1H
signal originating from the hood material was evaluated using a non-selective
3D gradient-echo acquisition with ultrashort TE (Philips, 7T).
Acoustic tests
The whole setup was tested twice
on noise attenuation. First, the performance of the acoustic hood was evaluated
in an acoustic room (ISO 8253-2 2012) (fig 2a). Four loudspeakers were
placed around the setup with a microphone at the ear-position. The loudspeakers
(Yamaha MSP5) played pink noise and were separately calibrated to a sound level
of 70dB(A). Sound levels were recorded with the microphone (Brüel & Kjær 4189
with ZC-0032 amplifier) and measured with a sound level meter (Brüel & Kjær
2250-S). Three measurements were taken from this setup: left-right speakers on,
front-back speakers on and all speakers on.
Then, a second test was done
inside the 7T MRI to measure the sound level reduction during several scan
protocols (fig. 2b). Measurements were done with a microphone (Behringer
ECM8000), recorded with Audacity and processed in Matlab (MathWorks). The
microphone level had not been calibrated. Therefore, the results inside the MR
scanner show relative damping.
Results
Our final product matches all the requirements. The
material did not generate 1H signal during imaging.
The acoustic test showed that when the left and right
speakers were turned on, the hood damped more than 10dB. When the front and
back speakers were turned on, only 3.5dB damping was achieved, resulting in
damping of 6.2dB when all speakers were turned on (fig. 3a).
The tests inside the 7T MR scanner (fig. 3b) showed that
the acoustic hood damped the acoustic noise during scanning between 9 and 13dB,
with an average of 10.6dB. This is 3.1 dB more damping as compared to the
prototype acoustic hood.Discussion and conclusion
We presented a new acoustic hood for neonatal MRI at 7T that damps the
acoustic noise between 6 and 10dB. The hood is lightweight and easy to handle
by a single person. The new hood achieves an additional 3.1dB damping compared
to the previous prototype hood.Acknowledgements
No acknowledgement found.References
1.
A. Nordell, M. Lundh, S. Horsch, B. Hallberg, U. Aden,
B. Nordell, M. Blennow. The acoustic hood: A patient-independent device improve
acoustic noise protection during neonatal magnetic resonance imaging.
Acta Paediatr. 2009
2. E.
Huijing, K. Rhebergen, P. Stroosnijder, D. Klomp, J. Wijnen, M. Philippens, K.
Annink, S. Mandija.Damping
of acoustic noise for fetal/neonatal MRI and MR-guided radiotherapy. ISMRM 2020