Hannah Kurdila1, Tayeb Zaidi1, Ting Zhang1, Subha Maruvada1, and Sunder Rajan1
1Food and Drug Administration, Silver Spring, MD, United States
Synopsis
The purpose of this study was to determine the expected
sound exposure to the neonate during neonatal brain protocols. To accomplish this, 7 neonatal brain protocols were recorded
on 7 different MRI machines across 4 vendors. Neonatal protocol sound levels straddled
existing notions of risk, exceeding sound levels known to cause non-auditory
stress responses in neonates but not exceeding the IEC MRI hearing safety
limit. These results indicate that these sound levels could be risky for the
neonate, but that further work is required to clarify this.
Introduction
There is concern that
neonates’ hearing could be harmed due to the noise produced during magnetic
resonance imaging (MRI) procedures, but knowledge gaps preclude making a safety
assessment. As
a first step in making a neonatal MRI hearing safety assessment, a
sound level survey of MRI scanners was conducted to quantify the expected sound
exposure during neonatal brain protocols. Methods
Neonatal brain protocols were
recorded using a sound level meter (B&K 2250, Bruel and Kjaer, Nærum, Denmark).
The protocols recorded on scanners A, C and D were protocols designed by
facilities that regularly scan neonates. The protocols recorded on the
remaining scanners (based at facilities that don’t scan neonates) were made to
emulate the scanner A protocol (Figure 1).
The sound level meter was
calibrated using a sound calibrator (B&K 4231, Bruel and Kjaer, Nærum,
Denmark) prior to each recording session. The sound level meter was fit with a
1/2’’ pressure field microphone (B&K 4192, Bruel and Kjaer, Nærum, Denmark)
and a preamplifier (B&K ZC-0032, Bruel and Kjaer, Nærum, Denmark), mounted
on a phantom provided by each facility and then placed in the isocenter of the
bore. The microphone and preamplifier were attached to a 10-meter-long cable
(B&K AO-0697-D-100, Bruel and Kjaer, Nærum, Denmark), which connected to
the sound level meter that was out of reach of the magnetic field.Results
We measured the sound pressure
levels of 7 neonatal brain protocols on 7 different MRI machines (Figure 2)
across 4 manufacturers. The average scan level A-weighted equivalent value (LAEQ)
was 96.88 dBA and the range was 80.8 – 109.3 dBA.
Our data also enables
within-scanner and between-scanner comparisons. The protocol recorded on the
neonatal scanner was found to be significantly quieter than the other scanners using
the Mann-Whitney U-Test (P = 4.44e-05, f =.0811, Cohen d = 1.68). Further, the
LAEQ ranges of the scans from each protocol vary between 8.49 – 11.9 dBA, which
may indicate that scan type can, at most, alter the sound pressure level by
about 10 dBA. In addition, the range of neonatal brain protocol average LAEQs
across the scanners is 14.87 dBA.
The sound level results are
summarized in Figure 3.Discussion
The recorded neonatal protocol
sound pressure levels exceeded sound
pressure levels that have been associated with negative non-auditory outcomes
in infants including changes in heart and respiratory rates, transcutaneous
oxygen tension, and intracranial pressure1,2, indicating that
neonates likely experience non-auditory reactions to loud sounds during
neonatal brain protocols (Figure 4). However, the degree to
which these non-auditory outcomes are dangerous is unknown 3.
The recorded neonatal protocol
sound pressure levels did not exceed the dosage prescribed by the IEC
60601-2-33 MRI hearing safety limit 4 (Figure 5), though a
comparison to the IEC 60601-2-33 MRI hearing safety limit may not be
appropriate for neonates.
Previous
sound level studies report adult
scan LAEQs between 84 dBA and 112.5 dBA over the past 31 years 5-12. Though adult and neonatal MRI scans are different, the scan LAEQs
from our neonatal sound level survey are in-line with the adult scan LAEQs from
the literature. This suggests that adult scans may not be louder than neonatal
scans.
However,
in practice, the acoustic exposure to the neonatal patient may be significantly
louder, as the amount of attenuation from infant hearing protection varies
wildly depending on the options used. The sites we visited that
routinely scanned infants (location 1 and 2) reported using the following
neonatal hearing protection devices (HPDs): neonatal noise attenuators (Minimuffs, Natus, Pleasanton, CA) only,
neonatal noise attenuators and a soft silicone putty (Macks pillow soft
silicone earplugs, McKeon Products, Inc., Warren, MI), and foam earplugs
designed for adults cut-to-infant-size (such as E-A-Rsoft Yellow Neons, E-A-R,
Boulder, CO). The noise reduction
rating (NRR) 13 of Minimuffs is only 7 dBA, Macks silicone earplugs
is 22 dBA and E-A-Rsoft Yellow Neons is 33 dBA, though when cut, this NRR is no
longer dependable. Based on the calculations in an OSHA standard 14
and using these NRRs, Minimuffs alone are expected to offer no attenuation in
practice, the dual-use of Minimuffs and Macks is expected to offer 12.5 dBA of
attenuation and the use of E-A-Rsoft Yellow Neons on adults is expected to
attenuate the exposure by 12.5 dBA.Conclusion
The sound levels achieved
during neonatal protocols likely induce negative non-auditory outcomes in
infants, but it is inconclusive whether they cause negative auditory outcomes
due to the dearth of information surrounding neonatal hearing tolerance. In
response to this gap, we are pursuing a prospective cohort study to evaluate
whether MRI machine sound levels induce a negative auditory outcome in infants.
Without this information, a definitive safety evaluation cannot be made. For
additional future work, we would like to clarify whether neonatal MRI protocol
sound levels are comparably loud to adult MRI protocol sound levels, which the
data suggested. Acknowledgements
We thank Maggie Johnson,
Stanley Fricke, Nathan Macdonald and John Mugler for providing access to the
MRI machines. We would also
like to thank Dr. Jana Delfino for helpful discussions and editing.
References
1. Long JG, Lucey JF, Philip
AGS. Noise and Hypoxemia in the Intensive Care Nursery. Pediatrics.
1980;65(1):143.
2. Anderssen SH, Nicolaisen
RB, Gabrielsen GW. Autonomic response to auditory stimulation. Acta Paediatr.
1993;82(11):913-8.
3. Morris B, Philbin, M.
Kathleen, Bose, Carl. Physiological Effects of Sound on the Neonate. Journal of
Perinatology. 2000;20.
4. IEC. Medical Electrical
equipment - Part 2-33: Particular requirements for the basic safety and
essential performance of magnetic resonance equipment for medical diagnosis.
2010.
5. Hurwitz R, Lane SR, Bell
RA, Brant-Zawadzki MN. Acoustic analysis of gradient-coil noise in MR imaging.
Radiology. 1989;173(2):545-8.
6. McJury PhD M, Shellock PhD
FG. Auditory Noise Associated With MR Procedures: A Review. Journal of Magnetic
Resonance Imaging. 2000;12(1):37-45.
7. McJury MJ. Acoustic noise
levels generated during high field MR imaging. Clin Radiol. 1995;50(5):331-4.
8. Counter SA, Olofsson A,
Grahn HF, Borg E. MRI acoustic noise: Sound pressure and frequency analysis.
Journal of Magnetic Resonance Imaging. 1997;7(3):606-11.
9. Price DL, De
Wilde JP, Papadaki AM, Curran JS, Kitney RI. Investigation of acoustic noise on
15 MRI scanners from 0.2 T to 3 T. J Magn Reson Imaging. 2001;13(2):288-93.
10. Radomskij P,
Schmidt MA, Heron CW, Prasher D. Effect of MRI noise on cochlear function. The
Lancet. 2002;359(9316):1485-6.
11. Hattori Y,
Fukatsu H, Ishigaki T. Measurement and evaluation of the acoustic noise of a 3
Tesla MR scanner. Nagoya J Med Sci. 2007;69(1-2):23-8.
12. McNulty JP,
McNulty S. Acoustic noise in magnetic resonance imaging: An ongoing issue.
Radiography. 2009;15(4):320-6.
13. NIOSH.
Occupational Noise Exposure. 1998 June.
Contract No.: 98-126.
14. OSHA. Methods
for Estimating HPD Attenuation. https://www.osha.gov/SLTC/noisehearingconservation/attenuation.html.