Rieke Lisa Meister1, Shuo Zhang2, Michael Groth1, Julian Jürgens1, Christoph Katemann2, Jan-Hendrik Buhk3, and Jochen Herrmann1
1Department of Diagnostic and Interventional Radiology and Nuclear Medicine, Section of Pediatric Radiology, University Medical Center Hamburg-Eppendorf, Hamburg, Germany, 2Philips, Hamburg, Germany, 3Department of Diagnostic and Interventional Neuroradiology, University Medical Center Hamburg-Eppendorf, Hamburg, Germany
Synopsis
High-resolution MRI
plays an important role in neuro-oncological exams in children. However,
high-quality imaging remains challenging due to long scan times. In this study,
we evaluate the compressed SENSE technique that employs compressed sensing with
coil sensitivity information in a dedicated pediatric neuro-oncological scan
protocol in children with brain tumor. The resulted image quality and scan
duration are compared with the conventional techniques, showing promise for a wide
adaption in the clinical routine practice.
Introduction
High-resolution magnetic
resonance imaging (MRI) is essential in studying morphological and
pathophysiological changes related to tumor development and treatment in the brain.
Practical challenges in pediatric neuro-oncological MRI examination remains due
to long imaging time. Compressed sensing is a new and promising technology that
has recently become clinically available to shorten the scan times 1.
The purpose of this work is to investigate the application of compressed
sensing and its potential advantage with comparison to conventional techniques in
children with brain tumor.Methods
Patients with various
neuro-oncological disorders undergoing brain MRI exams were recruited in this
study. All subjects provided written informed consent and the study was
approved by the institutional review board. Image acquisition was done on a 3.0T
clinical whole-body MRI system (Ingenia, Philips, Best, the Netherlands) with a
standard 32-channel head coil. Pediatric neuro-oncological scan protocol
consisted of imaging pulse sequences, including 3D T1-weighted turbo-field-echo
(TFE) pre- and post-contrast, T2-weighted turbo-spin-echo (TSE), and fluid
attenuated inversion recovery (FLAIR). Conventional T1-TFE and FLAIR used parallel
imaging with sensitivity encoding (SENSE). The corresponding protocol during
the follow-up study was equipped with the recently introduced compressed SENSE
(C-SENSE) technique, which employed compressed sensing with coil sensitivity
information 2. While the main imaging parameters (Table 1) were kept comparable between
the C-SENSE and standard studies, C-SENSE FLAIR was adjusted additionally for
best fluid suppression. Images were assessed for each patient by two experienced
pediatric radiologists who were blinded for type of study and clinical
information in consensus reading. Image analysis was based on visual inspection
according to general quality, artifacts, diagnostic confidence and
visualization of the anatomical structures in designated regions of interest.
In addition, scan duration of the individual sequences was recorded. A p-value of less than 0.05 was considered
statistical significance.Results and Discussion
Twenty-two patients (age 2.3 to
18.8 years; median age 10.37 years) were included. For T1-TFE and FLAIR,
C-SENSE allowed for an improved spatial resolution for approximately 40% and
14%, respectively, and a reduction in scan duration (18% and 32%) at the same
time. For 2D T2-TSE, C-SENSE was employed with the same spatial resolution but
for 41% reduced scan duration. Total scan duration was reduced on average from
14:45 min to 10:44 min by applying C-SENSE, with a reduction of approximately
27%. A comparison between scans with and without C-SENSE was shown in Table 2. Single-shot diffusion-weighted
imaging (DWI) based on echo-planer imaging (EPI) was performed as part of the
scan protocol but not included in the evaluation because the pulse sequence was
not compatible with C-SENSE at the time of this study.
Overall image quality was found comparable between
studies with and without C-SENSE, while it was rated higher for all C-SENSE
sequences (13.2 ± 1.0 with C-SENSE vs 11.9 ± 0.7 without C-SENSE on average). Typical
examples were demonstrated in Figure 1.
Overall artifacts were rated less in all sequences with C-SENSE (lower score
for 3.2 ± 0.9 with C-SENSE vs 4.4 ± 1.0 without C-SENSE on average), which was
significant for FLAIR (2.7 with C-SENSE vs 5.4 without C-SENSE, p <0.05). Details of the qualitative
rating scores were summarized in Figure 2.
No loss of sharpness or contrast was seen in the C-SENSE scans. It is
noteworthy that liquor pulsation artifacts were significantly reduced in
C-SENSE FLAIR images owing to incoherent undersampling and additional parameter
adjustments, which improved visualization of the adjacent anatomical
structures. Artifacts related to C-SENSE were found to have different
manifestations in T1 images, but were deemed no impairment for diagnosis.Conclusion
Pediatric brain MRI in
neuro-oncological studies benefited from C-SENSE in shortening the scan
duration while maintaining or even improving image quality. The advantage for
daily clinical routine has been demonstrated with promise in shortened procedural
time, with additional potentials in protocol standardization, sedation
reduction and patient compliance improvement in children. Further studies in a
larger cohort are warranted to investigate its clinical performance.Acknowledgements
No acknowledgement found.References
-
Lustig M, et al. Magn Reson Med (2007) 58:1182.
-
Sartoretti E et al. Plos One (2019) 14:e0214887.