Layla Tabea Riemann1, Christoph Stefan Aigner1, Ralf Mekle2, Sebastian Schmitter1, Bernd Ittermann1, and Ariane Fillmer1
1Physikalisch-Technische Bundesanstalt (PTB), Braunschweig und Berlin, Germany, 2Center for Stroke Research Berlin, Charité Universitätsmedizin, Berlin, Germany
Synopsis
This
work assesses the test-retest repeatability and reproducibility of spectral
shapes and neurochemical profiles of the SPECIAL 1H MR spectroscopy
sequence employing three different adiabatic pulses in-vivo at 7T: the standard
hyperbolic secant pulse and two gradient-modulated pulses, namely a GOIA and a
WURST pulse. Nine healthy volunteers were scanned four times each, with all
three SPECIAL variants to establish three different repeatability or
reproducibility measures and to evaluate the limits in the precision of the
resulting metabolite quantification.
Introduction
In single voxel
spectroscopy, the use of short TEs is favorable to minimize effects due to T2/T2*
relaxation and J-coupling1. One way to achieve this is to use the SPin
ECho, full Intensity Acquired Localized (SPECIAL)2,3 sequence.
To assure reproducible results at 7T, the impact of B0 and B1
inhomogeneities should be minimized4. This is particularly important for the
adiabatic inversion pulse used in SPECIAL, as inaccuracies in the subtraction
will lead to inaccuracies in the localization of the VOI.
This work aims to
determine and compare the test-retest repeatability and the reproducibility of
neurochemical profiles using the SPECIAL sequence at 7T for three different
adiabatic inversion pulses, namely the originally implemented hyperbolic secant
(HS)5 pulse, a gradient-offset independent
adiabaticity (GOIA)6 pulse, and a wideband, uniform rate, smooth
truncation (WURST)7 pulse. Methods
Single volume spectra
from the posterior cingulate cortex (PCC) of nine healthy volunteers were
acquired using a SPECIAL sequence with HS, GOIA, and WURST pulses on a 7T
Scanner (Magnetom 7T, Siemens Healthineers, Erlangen, Germany) utilizing a
1Tx/32Rx-channel head coil (Nova Medical Inc., Wilmington, MA, USA). All measurements
were approved by the local ethics board. Fig.1 shows the sequence diagram (Fig.1a), scan scheme
(Fig.1b), and voxel positioning (Fig.1c). Tab.1 summarizes the
pulse and sequence parameters.
Each volunteer was
scanned four times with three sequence variants: twice on the first day, repeating
all measurements of the first session after repositioning; and twice one week
after when repeated measurements with the same sequence variant were acquired
without repositioning in-between. In this way, the intra-session repeatability
R0 and the reproducibilities after repositioning, i.e. a few
minutes later, R1,M, and one week later, R1,W, were
assessed. The order of acquisition for the sequence
variants was cyclically permuted for the different volunteers to avoid bias
from scan time elapsed before the respective sequence variant was performed.
Spectra were post-processed with MATLAB (The
Mathworks, Natick,
MA, USA), including the summation of the even and uneven averages to
obtain full localization, weighted and phase-corrected coil combination,
frequency correction, and averaging. The post-processed
spectra were quantified with LCModel8. Metabolite concentrations were corrected for
relaxation and individual tissue composition using SPM129.
To compare the
spectral shape of two measurements and obtain R0,R1,M, and R1,W, the spectral difference is calculated in
the frequency domain over the relevant frequency range (4.2 to 0.8 ppm). To
generate one point for subject $$$i$$$ in the Bland-Altman plots, the integral
of the difference of the absolute between the compared spectra $$$BA_{i,y}$$$ was
plotted over the integral of the absolute averaged spectrum $$$BA_{i,x}$$$:
$$BA_{i,y} = \int\limits_{f_0}^{f} | x(f')_{i,1\_1} | - | x(f')_{i,1\_2} | \ df',$$
$$BA_{i,x} = \int\limits_{f_0}^{f} |\ \overline{x(f')_i} | \ df'.$$
$$$| x(f')_{i,1\_1} |$$$ and $$$| x(f')_{i,1\_2} |$$$ indicate the absolute of the real part of the
spectral shapes at frequency $$$f’$$$ for two sessions, while $$$|\ \overline{x(f')_i} |$$$ indicates the absolute mean out of both sessions.
Metabolite
concentrations quantified with mean Cramér-Rao lower bounds (CRLBs)≥20% and single values of CRLBs>50%
were discarded10. To compare results between the different sequence variants,
concentrations and CRLBs were averaged over
all subjects and sessions. Furthermore, coefficients of variance (CVs) were determined for all valid quantification
results for each volunteer over the four sessions and then averaged over all
subjects11. Statistical
significance was assessed with a Wilcoxon signed-rank sum test
with a significance level of p<0.05. Results and Discussion
Spectral
quality achieved with all three pulses was of comparable and high quality (Fig.2a),while the Bland-Altman plots (Fig.2b) indicate more repeatable results (R0) when
using adiabatic GOIA and WURST pulses. For the reproducibility scenarios
(R1,M, and R1,W), all three pulses perform on a similar
level. As expected, repeatability/reproducibility was improved, the shorter the
time in-between measurements, i.e. R0<R1,M<R1,W.
For all different
inversion pulses and all selected metabolites, similar concentrations and
inter-subject variances were observed (Fig.3a). Compared to the HS, slightly
higher concentrations for GSH and tCr are obtained with GOIA and WURST pulses.
The effect is statistically significant but small. Concentrations and CRLBs are
both comparable with literature values for this region11. Out of a total of 468 metabolite concentration
values (9 volunteers x 4 sessions x 13 quantified
metabolites), the number of measurements that are not
included as they exceed the CRLBs limits was 18 for HS-SPECIAL, 7 for GOIA-SPECIAL, and 4 for WURST-SPECIAL. The CRLBs are higher for
most of the metabolites measured with HS-SPECIAL (Fig.3b), again a small but
statistically significant effect. In addition,
HS-SPECIAL showed the highest averaged intra-subject CV for most of the
metabolites, except for GABA and Lac (Fig.3c). These results might be explained
by the reduced chemical shift displacement error and sharper slice profiles due
to the increased RF bandwidth of the gradient-modulated pulses.Conclusion
In conclusion, the gradient-modulated
pulses that were used in the SPECIAL sequence as adiabatic inversion pulses tend
to result in decreased intra-subject variation and CRLBs, as well as in
improved short-term repeatability compared to the commonly used HS pulse. Thus,
the use of GOIA and WURST pulses with SPECIAL yields a more robust metabolite quantification than
the HS pulse. However, after one week, intra-individual physiological
variations seem to become the dominating factor outshining the performance
differences between the different pulses.Acknowledgements
This
project has received funding from the EMPIR program co-financed by the
Participating States and from the European Union’s Horizon 2020 research and innovation program.
This paper reflects only the author's view and EURAMET is not responsible for
any use that may be made of the information it contains. Furthermore this project received funding from grant number IT7/8-1 of the DFG.References
1.
Zhong K, Ernst T. Localized in vivo
human 1H MRS at very short echo times. Magn Reson Med.
2004;52(4):898-901. doi:10.1002/mrm.20201
2.
Mlynárik V, Gambarota G, Frenkel
H, Gruetter R. Localized short-echo-time proton MR spectroscopy with full
signal-intensity acquisition. Magn Reson Med. 2006;56(5):965-970.
doi:10.1002/mrm.21043
3.
Mekle R, Mlynárik V, Gambarota G,
Hergt M, Krueger G, Gruetter R. MR spectroscopy of the human brain with
enhanced signal intensity at ultrashort echo times on a clinical platform at 3T
and 7T. Magn Reson Med. 2009;61(6):1279-1285. doi:10.1002/mrm.21961
4.
Ladd ME, Bachert P, Meyerspeer M,
et al. Pros and cons of ultra-high-field MRI/MRS for human application. Prog
Nucl Magn Reson Spectrosc. 2018;109:1-50. doi:10.1016/j.pnmrs.2018.06.001
5.
Bernstein. Handbook of MRI
Pulse Sequences. Elsevier; 2004. doi:10.1016/B978-0-12-092861-3.X5000-6
6.
Tannús A, Garwood M. Adiabatic
Pulses. NMR Biomed. 1997;10(2-3):423-434. doi:DOI:
10.1002/(sici)1099-1492(199712)10:8<423::aid-nbm488>3.0.co;2-x
7.
Andronesi OC, Ramadan S, Ratai
EM, Jennings D, Mountford CE, Sorensen AG. Spectroscopic Imaging with Improved
Gradient Modulated Constant Adiabadicity Pulses on High-Field Clinical
Scanners. J Magn Reson. 2011;23(1):1-7.
doi:10.1161/CIRCULATIONAHA.110.956839
8.
Provencher SW. Estimation of
metabolite concentrations from localized in vivo proton NMR spectra. Magn
Reson Med. 1993;30(6):672-679. doi:10.1002/mrm.1910300604
9.
Friston K, Ashburner J, Kiebel S,
Nichols T, Penny W, eds. Statistical Parametric Mapping. 1st ed. Elsevier;
2007. doi:10.1016/B978-0-12-372560-8.X5000-1
10.
Kreis R. The trouble with quality
filtering based on relative Cramér-Rao lower bounds. Magn Reson Med.
2016;75(1):15-18. doi:10.1002/mrm.25568
11.
Terpstra M, Cheong I, Lyu T, et al.
Test-retest reproducibility of neurochemical profiles with short-echo,
single-voxel MR spectroscopy at 3T and 7T. Magn Reson Med.
2016;76(4):1083-1091. doi:10.1002/mrm.26022