Meryem Khalfallah1, Gwenaël Pagé1, Sabrina Doblas1, Bernard Van-Beers1,2, and Philippe Garteiser1
1Center of research on inflammation (UMR1149), Inserm - Université de Paris, Paris, France, 2Beaujon hospital radiology departement, APHP, Paris, France
Synopsis
Temporal diffusion spectroscopy using
oscillating encoding gradients enables assessing tissue microstructure at a
user-selected, arbitrary spatial scale. It has been demonstrated that this
method is feasible with OGSE-EPI sequence. To study its robustness an OGSE-EPI sequence
was implemented and repeatability and signal to noise ratio were assessed with
two radiofrequency hardware setups and with or without respiratory gating. The
effect of OGSE frequency was assessed for the pure diffusion coefficient (D)
and for the apparent diffusion coefficient (ADC). Using a 4-channel surface
coil and respiratory gating when measuring ADC at 100 Hz, the optimal
repeatability coefficient (28%) was obtained.
Introduction
Diffusion-weighted
imaging (DWI) probes the diffusion properties of water, which are modulated by
the microstructure of biological tissues1,2. Conventionally, ADC is obtained with pulsed gradient spin echo
sequences, and lacks specificity due to the integrated influences of several
spatial scales. In temporal diffusion spectroscopy, oscillating encoding
gradients are used, which enables to assess tissue microstructure at a user-selected,
arbitrary spatial scale. Typically, scales smaller than the diameter of a
single cell are achievable1,3. Oscillating gradient spin echo (OGSE) methods have been
implemented with spin echo4, fast spin echo5 or echoplanar imaging (EPI)6. OGSE has been used for estimating hepatocyte size7, nuclear size changes after anticancer therapy5,8,9, and hepatocellular nodule characteristics in liver cirrhosis4.
Although measurement of hepatocyte cell size has
been demonstrated to be feasible with OGSE-EPI, studies of the robustness of
this method in mice are lacking. Here, an OGSE-EPI
sequence was implemented and the repeatability and signal to noise ratio (SNR) were
assessed with two radiofrequency hardware setups and with or without respiratory
gating. The effect of OGSE frequency was assessed for the pure diffusion
coefficient (D) and for the apparent diffusion coefficient (ADC).
Methods
Sequence
An
OGSE sequence was programmed with an EPI readout, by inserting oscillating
gradients symmetrically on either side of the refocusing pulse. The
double-sinusoid waveform proposed by Does2,
was implemented.
Sequence parameters
All
experiments were carried out on a 7T, 300mT/m Bruker system equipped with
either a volume coil or a combination of a volume transmit coil and an actively
decoupled, 4-channel surface coil array for reception. Acquisitions were
performed consecutively with and without respiratory gating. The desired
b-values (0, 200, 300 and 400 s/mm² for the ADC and 200, 300 and 400 s/mm² for
the D) were obtained by varying the gradients amplitude at a constant echo time,
which on our system was minimally 52.8 ms. Other sequence parameters were: TR =
5 s, matrix size = 96 x 64, FOV = 40 x 40
mm, spatial resolution = 0.42 x 0.63 mm2, slice thickness = 1mm, NEPI
shots = 8, Number of averages = 2, with spectral-selective fat suppression.
Two frequencies were tested (100 and 150 Hz). Scan time per b-value was 1 min 20
s and 3 min 40 s without and with respiratory gating, respectively.
Image analysis
ADC and D
values were retrieved by a linear regression of the log of the signal versus
b-values (monoexponential fit). Signal to noise ratio at b = 400 s/mm² was estimated
with the standard deviation of the subtraction of the two available image
averages as noise estimation.
Repeatability study
In
all tested conditions, the MRI acquisitions were repeated twice during a single
anesthesia in a group of five healthy mice (after appropriate ethical
authorization) Repeatability was estimated with Bland-Altman tests and coefficients
of repeatability (standard deviation (%) * 1.9610) were calculated.Results
Repeatability coefficient and SNR values were
better (i.e smaller repeatability coefficient and higher SNR) when using the 4-channel
surface coil than when using the volume coil in 7 of 8 tested combinations
(Table 1). Volume coil images had larger sensitivity area while surface coil
images had higher signal magnitudes (figure 1). Repeatability coefficient and SNR
values were better when using respiratory gating in 6 of 8 and 8 of 8 tested
combinations respectively. Visually, respiratory gating had an important effect
on image quality (figure 1). SNR values were better at 150 Hz in 8 of 8 tested
combinations. Also when measuring ADC (with four b-values), we obtained better
repeatability coefficient and SNR values than when measuring D (three b-values)
in 6 of 8 tested combinations (figure 2). The optimal repeatability coefficient
was 28% obtained using the 4-channel surface coil and respiratory gating when
measuring ADC at 100 Hz.Discussion
Gains in repeatability were achieved by
increasing the image SNR using a 4-channel surface coil. Despite the slight TR variability
introduced by respiratory gating, it was found that a better repeatability was
achieved than without respiratory gating, where the TR is strictly defined by
the MR sequence, but where animal breathing perturbs the acquisitions. In Jiang
et al.[7], acquisitions were performed with respiratory
gating. A better repeatability was obtained for ADC. This could be explained by
the greater number of b-values or the introduction of high signal b0 images to
the ADC fit.Conclusion
The OGSE
sequence with EPI readout is applicable at 7T in vivo in mice. The proposed
acquisition conditions using 4-channel parallel coil, respiratory gating and 4
b-values enable to optimize the repeatability of the measurement.Acknowledgements
This work was funded in part by the Agence Nationale de la Recherche Investissements d’avenir programme ANR-17- RHUS-0009 (Quid-NASH).
Authors wish to thank the Inserm UMS34 platform FRIM (Federation of Research on imaging and multimodality).
This work was funded in part by the Agence Nationale de la Recherche AAPG programme ANR-20-CE19-0005-01 (STEDI-NASH).
References
1. Xiaoyu
Jiang, e.a., In vivo magnetic resonance
imaging of treatment-induced apoptosis. Scientific Reports, 2019.
2. Mark D. Does, e.a., Oscillating Gradient Measurements of Water
Diffusion in Normal and Globally Ischemic Rat Brain. Magnetic Resonance in
Medicine 2003.
3. Junzhong Xu, e.a., Quantitative Characterization of Tissue
Microstructure withTemporal Diffusion Spectroscopy. J Magn Reson Imaging,
2009.
4. Mathilde Wagner, e.a., Comparison of Pulsed and Oscillating
Gradient Diffusion-Weighted MRI for Characterizing Hepatocellular Nodules in
Liver Cirrhosis: ex vivo Study in a Rat Model. J. MAGN. RESON. IMAGING,
2019.
5. Junzhong Xu, e.a., Characterizing Tumor Response to
Chemotherapy at Various Length Scales Using Temporal Diffusion Spectroscopy.
PlosOne, 2012.
6. Xiaoyu Jiang, e.a., In vivo imaging of cancer cell size and
cellularity using temporal diffusion spectroscopy. Magnetic Resonance in
Medicine, 2017.
7. Xiaoyu Jiang, e.a., Mapping hepatocyte size in vivo using
temporal diffusion spectroscopy MRI. Magnetic Resonance in Medicine, 2020.
8. Xiaoyu Jiang, e.a., Quantification of Cell Size Using Temporal
Diffusion Spectroscopy. Magnetic Resonance in Medicine, 2016.
9. Colvin, e.a., Earlier detection of tumor treatment response using magnetic resonance
diffusion imaging with oscillating gradients. Magn Reson Imaging, 2011.
10. Anneloes E. Bohte, e.a., MR Elastography of the Liver: Defining
Thresholds for Detecting Viscoelastic Changes. Radiology, 2013.