Joao Filipe Fernandes1, Harminder Gill1, Julio Sotelo2,3,4, Shu Wang1, Alessandro Faraci1, Cristian Montalba5, Jesus Urbina6, Ronak Rajani1, David A. Nordsletten1,7, Kawal Rhode1, Sergio Uribe6,8,9, and Pablo Lamata1
1School of Biomedical Engineering and Imaging Sciences, King’s College, London, United Kingdom, 2School of Biomedical Engineering, Universidad de Valparaiso, Valparaiso, Chile, 3Biomedical Imaging Center, Pontificia Universidad Católica de Chile, Santiago, Chile, 4Millennium Nucleus for Cardiovascular Magnetic Resonance, ANID - Millennium Science Initiative Program, Santiago, Chile, 5Biomedical Imaging Center, Pontificia Universidad Catolica de Chile, Santiago, Chile, 6Radiology Department, School of Medicine, Pontificia Universidad Catolica de Chile, Santiago, Chile, 7Departments Biomedical Engineering and Cardiac Surgery University of Michigan, Ann Arbor, MI, United States, 8Biomedical Imaging Center, Pontificia Universidad Catolica de Chile, Santiago, Chile, 9Millennium Nucleus for Cardiovascular Magnetic Resonance, Santiago, Chile
Synopsis
Aortic valve (AV) conditions cause extra burden to the heart and frequently lead to clinicalintervention. In the present study we set a 4D-flow-MRI framework to evaluate in-vitro personalizedcompliant 3D-printed AV. We evaluated a healthy and three diseased AV under rest to stresspulsatile flow conditions. The results obtained provide further evidence that trans-valvular non-invasive pressure drop is estimated more accurately accounting for full velocity profile, via thesimplified advective work-energy relative pressure (SAW), than accounting solely for the maximalvelocity as it is clinically stablished. Both the methodology and the findings can potentially improveclinical decision-making.
Introduction
The Aortic valve (AV) is the cardiac
valve that has to cope with the biggest amount of stress, and therefore is the
one more frequently developing pathologies1.
The AV conditions typically increase with ageing, resultant of a calcification
of the leaflets (rheumatic AV) or annulus resulting in a narrowed AV area (stenotic valve). The AV can
also be congenitally bicuspid. All these conditions generate an extra load to
the heart. Ultimately, the AV can be
repaired/replaced with clinical intervention, with the associated inherent
risks. The main clinical-decision parameter to access the AV condition is the
trans-valvular pressure drop (∆P), which is assessed non-invasively by simplified Bernoulli equation (SB=4*v2) based on the peak velocity
across Av or invasively as the peak-to-peak (PtP) ΔP between LV and
ascending aorta. With the continuous development of 4D-flow MRI, it becomes
possible to acquire the flow across AV in 3D and over time with good
resolution. Therefore, it is possible to account for a full velocity profile across
the vena-contracta in the estimation ΔP via the simplified advective
work-energy relative pressure (SAW = 1/Q (A), where Q is flow and A is the
advective energy rate)2.
Moreover, typical cardiovascular MRI scanning include 4D flow sequences can take
for up to 45 minutes. For the elderly population, typical in AS is prevalent
this can preclude investigation. Therefore, in vitro phantom models which can
closely replicate valvular and aortic physiology and function can allow for
intervention planning, valve assessment and potentially evaluate novel imaging
techniques without the need for in vivo subject participation. The authors
describe a compliant, functional valve phantom within a compliant aortic
phantom under pulsatile flow to aiming to provide a realistic model of several
AV conditions and compare invasive catheterisation against 4D-flow-MRI velocity
ΔP.Methods
The fabrication of compliant valves
mimicking healthy AV and three AV conditions (rheumatic, stenotic and bicuspid)
was based on the segmentation of computer tomography image of a patient-specific
AV. Then the valve mould was meshed and 3D-printed. De-gassed silicone 0030 was
injected into the mould and cured for 2 days. Each valve was inserted in the
the aortic root of a compliant aorta phantom immersed in Agar gel and connected to
a hydraulic circuit filled with a blood-mimicking fluid and activated by an MRI-compatible
flow pump (Figure 1)3.
For mirroring cardiac output from rest to stress, each valve was imagined under
three pulsatile flow rates reaching 150ml/s, 200ml/s and 250ml/s (respectively except
the later in the stenotic AV to avoid overcoming the maximum rated pressure of
the phantom). In each flow condition, the PtP ∆P across the valve was measured
by fluid-filled catheters inserted before and after the valve. Phase-contrast MRI
4D-Flow was acquired for each valve under each flow condition. The aortic
phantom was segmented to define the velocity filled of view. The velocity profiles
were obtained at the vena-contracta, were the basis of SB and SAW computation
over the full pulsatile cycle. Linear regression and Bland-Altman plots between
each methodological non-invasive ΔP estimations and the invasive ΔP measurements
were investigated.Results
In Figure 3 it is summarized the maximal pressure drop
across the different implanted valves, under each flow conditions, for catheterization
measurements and MRI-derived estimations (SB and SAW). The ΔP variation over
time at the vena-contracta for the highest flow rate conditions is presented in
Figure 4. SAW ΔP shows better agreement with invasive pressure recordings than
SB ΔP which is the basis of clinical routine (correlation R2 of 0.808
versus 0.751 respectively, with a linear regression coefficient β of 0.81
versus 1.51) (Figure 5A). Also, the Bland-Altman analysis indicates that the
comparison between SAW ΔP and invasive measurements have better agreement with
lower limits of agreement (Figure 5B&C).Discussion
The generation of the 3D-printed compliant valves and the
respective insertion in the compliant aorta phantom allowed a
comprehensive 4D-flow MRI study of the ΔP estimation based on the flow profiles
very similar to in-vivo. The results show a very good agreement between the ΔP
estimated via SAW in the vena-contracta and the PtP ΔP measured
invasively. Simultaneously, it was further confirmed that
the peak ΔP via SB is an overestimation of the real ΔP and has higher
variability. These differences on the Velocity-based ΔP (SB and SAW) indicate
that accounting for full velocity profile provides a more widespread,
and thus accurate, assessment of ΔP than accounting simply
for peak velocity used clinically. The potential impact of
this study in clinical practise is valuable in 2 directions. Firstly, it
provides extra evidence that the non-invasive estimations of AV ΔP should be
improved. Secondly, it sets a framework to the evaluation of more complex
and/or implantable personalized AVs in a fluid circuit outside of the human body
and therefore risk-free, before any actual intervention, and potentially
improving the respective outcome.Conclusion
The present study provides in-vitro evidence that non-invasive ΔP across aortic valves (healthy
and diseased) is better predicted by considering the full 3D flow profile
at the vena-contracta. the results further prove that peak
velocity across the valve overestimates aortic valve ΔP. Simultaneously, it was set a framework for generating and evaluating personalized compliant valves
which are MRI-compatible.Acknowledgements
PIC project, European Union’s Horizon 2020 Marie Skłodowska-Curie ITN Project under grantagreement No 764738. PL holds a Wellcome Trust Senior Research Fellowship (g.a. 209450/Z/17/Z).References
1. Thaden JJ, Nkomo VT, Enriquez-Sarano M. The global burden of aortic stenosis. Prog Cardiovasc Dis 2014;56:565-71.
2. Donati F, Myerson S,
Bissell MM et al. Beyond Bernoulli: Improving the Accuracy and Precision
of Noninvasive Estimation of Peak Pressure Drops. Circ Cardiovasc
Imaging 2017;10.
3. Urbina J, Sotelo JA,
Springmuller D et al. Realistic aortic phantom to study hemodynamics
using MRI and cardiac catheterization in normal and aortic coarctation
conditions. J Magn Reson Imaging
2016;44:683-97.