Jos J.M. Westenberg1, Hans C van Assen1, Pieter J van den Boogaard1, Jelle J Goeman1, Hicham Saaid2, Jason Voorneveld3, Johan Bosch3, Sasa Kenjeres4, Tom Claessens2, Pankaj Garg5, Marc Kouwenhoven6, and Hildo J Lamb1
1Leiden University Medical Center, Leiden, Netherlands, 2Ghent University, Ghent, Belgium, 3Erasmus Medical Center, Rotterdam, Netherlands, 4University of Technology Delft, Delft, Netherlands, 5Norwich University Hospital, Norwich, United Kingdom, 6Philips Healthcare, Best, Netherlands
Synopsis
Echo Planar Imaging (EPI) is associated with
inaccurate velocity quantitation in 4D flow MRI. The systematic errors depend on
the orientation of readout and blip phase encoding gradient with respect to the
flow direction. This study evaluates EPI-related errors in flow rate and
velocity quantitation for in vivo intracardiac 4D flow MRI in a phantom and in
ten healthy volunteers. Errors in median flow rate and velocity remain below
10% for normal in vivo intracardiac flow with EPI factor 5. The error is
smallest when readout and blip phase encoding gradients are both perpendicular
to the main flow.
Introduction
Echo Planar Imaging (EPI)1 using moderate EPI factors was applied for
intracardiac blood flow evaluation with 4D flow MRI to enable scanning times
below 10 minutes2,3. However, even for moderate EPI factors, systematic
quantitative inaccuracies are introduced with EPI4,5. The oscillatory
readout gradients create velocity-dependent phase shifts and the unipolar blip phase
encoding gradients create motion-induced phase shifts with a modulated point
spread function as a result affecting spatial resolution. Dillinger et al.5
recently investigated the limitations of EPI using in silico simulations and in
vitro experiments under high-flow conditions. In vivo intracardiac 4D flow was
not evaluated. The aim of the current study is to evaluate EPI-related errors
in flow rate and velocity quantitation for in vivo intracardiac EPI 4D flow MRI
compared to non-EPI gradient echo (GRE) 4D flow MRI.Methods
In vitro experiments on a pulsatile left
ventricle (LV) phantom (Figure 1, setup described by Saaid et al.6) and
in vivo experiments on ten healthy volunteers (Figure 2, 2/8 female/male, mean
age ± standard deviation 37±12 years; informed consent and local Medical
Ethical Committee approval was obtained) were performed on 3T MRI (Philips Healthcare).
MRI scan parameters are displayed in the table in Figure 3. Whole-heart 4D flow
MRI was obtained using standard non-segmented gradient echo (4DGRE) without EPI
and with EPI factor 5 (4DEPI). 4DEPI was repeated three times, each with different
orientations of readout and blip phase encoding gradient with respect to the
main flow direction (Figures 1 and 2): 4DEPI1_LA refers to EPI-acquisition in
long-axis orientation with readout gradient aligned parallel to the main flow
direction, 4DEPI2_LA refers to EPI-acquisition in long-axis orientation with blip
phase encoding gradient aligned parallel to the main flow direction and 4DEPI3_SA
refers to EPI-acquisition in short-axis orientation with both readout and blip
phase encoding gradients perpendicular to the main flow direction. 4D flow MRI
acquisitions were repeated on a phantom filled with static fluid to allow phase
offset correction by voxel-wise velocity subtraction7.
From each 4D dataset, median LV velocities were obtained from a static cylinder-shaped
control volume (in vitro 6,559 measurement voxels, in vivo 1,693±25 measurement
voxels) positioned below the mitral valve (MV). Spatial median velocities were compared
at peak systole, peak early diastole and over the entire cardiac cycle (i.e., spatio-temporal
median). In vivo MV and aortic valve (AoV) net forward volumes (NFVs) were assessed
by automated retrospective valve tracking8 using CAAS MR Solutions (Pie
Medical Imaging). The difference between MV and AoV NFV determined the
consistency in valvular flow assessment. Tracking was performed on cine
two-dimensional (2D) left 2-chamber and 4-chamber views and coronal and
sagittal views of the LV outflow tract using steady-state free-precession with
TE/TR 1.5/2.9 ms, field-of-view 350 mm, 45° flip angle, acquisition resolution
2.0×1.7×8.0 mm3, retrospective gating with 30 phases reconstructed and
end-expiration breath holding.
The 4D flow datasets were visually inspected for the presence of artifacts. Relative
difference and Pearson correlation (r) between MV and AoV NFV were determined between
sequences and compared using paired t-tests. Comparison of median velocities
between sequences was done after logarithmic transformation to account for skewness
of the distribution, as follows: logarithmic transformation of the median
velocity was performed first per cardiac phase, median differences were
calculated over the pump or cardiac cycle, respectively, and back
transformation resulted in ratios of median velocities between sequences. In
vivo, mean ratios and standard deviation between sequences were determined and
compared using paired t-tests.Results
The
table in Figure 4 shows median peak velocity magnitudes and interquartile
ranges obtained from the control volume inside the LV phantom. Median peak velocity
was ≤ 5.5%
lower for 4DEPI compared to 4DGRE. Ratios of median velocity comparing 4DEPI with
4DGRE show highest similarity (ratio 0.981) for 4DEPI3_SA with both readout and
blip phase encoding gradients perpendicular to the main flow direction and lowest
similarity (ratio 0.945) for 4DEPI1_LA with
readout gradient aligned parallel with the main flow direction.
In vivo, one 4DEPI1_LA dataset was rejected due to presence of unresolved artifacts. The
table in Figure 5 shows in vivo results. None of the acquisitions showed
statistically significant differences between MV and AoV NFV and all showed
strong correlations (r≥0.89, p<0.01). Mean NFV inconsistency was largest
(6.4±8.5%) for
4DEPI3_SA with both readout and blip phase encoding gradients perpendicular to the
main flow direction. The difference in median LV velocity was largest (9%) for
4DEPI1_LA with readout gradient parallel to the main flow direction.Conclusion
Flow
in the direction of the EPI readout or blip phase encoding gradient results in
inaccurate velocity and flow quantitation for EPI-accelerated 4D flow MRI. Errors
in median flow rate and median velocity and the inconsistency in MV and AoV NFV
all remained below 10% for normal in vivo intracardiac flow with EPI factor 5.
The error in flow rate was smallest (<2%) when readout and blip phase
encoding gradients are both perpendicular to the main flow direction.Acknowledgements
Contract
grant sponsor: Dutch Heart Foundation; Contract grant number: CVON2017-08-RADAR (for second author)References
1. DeLaPaz RL. Echo-planar Imaging.
Radiographics 1994; 14:1045-58.
2. Kozerke S, Hasenkam JM, Pedersen EM,
Boesiger P. Visualization of flow patterns distal to aortic valve prostheses in
humans using a fast approach for cine 3D velocity mapping. J Magn Reson Imaging
2001; 13(5):690-698.
3. Westenberg
JJ, Roes SD, Ajmone Marsan N, et al. Mitral valve and tricuspid valve blood flow: accurate quantification
with 3D velocity-encoded MR imaging with retrospective valve tracking.
Radiology. 2008; 249(3):792-800.
4. Duerk JL, Simonetti OP. Theoretical
aspects of motion sensitivity and compensation in echo-planar imaging. J Magn
Reson Imaging. 1991; 1:643–650.
5. Dillinger H, Walheim J, Kozerke S.
On the limitations of echo planar 4D flow MRI. Magn Reson Med. 2020; 84(4):1806-1816.
6. Saaid
H, Voorneveld J, Schinkel C, et al. Tomographic PIV in a model of the left ventricle: 3D flow past
biological and mechanical heart valves. J Biomech. 2019; 90:40-49.
7. Hofman
MBM, Rodenburg MJA, Markenroth Bloch K, et al. In-vivo validation of interpolation-based phase
offset correction in cardiovascular magnetic resonance flow quantification: a
multi-vendor, multi-center study. J Cardiovasc Magn Reson. 2019; 21(1):30.
8. Kamphuis VP, Roest AAW, Ajmone Marsan N, et
al. Automated cardiac valve
tracking for flow quantification with four-dimensional flow MRI. Radiology
2019; 290(1):70-78.