Carmen P S Blanken1, Lukas M Gottwald2, Jos J M Westenberg3, Eva S Peper2, Bram F Coolen4, Gustav J Strijkers4, Aart J Nederveen2, R Nils Planken2, and Pim van Ooij2
1Radiology and Nuclear Medicine, Amsterdam UMC, Amsterdam, Netherlands, 2Radiology and Nuclear Medicine, Amsterdam UMC, location AMC, Amsterdam, Netherlands, 3Radiology, Leiden UMC, Leiden, Netherlands, 4Biomedical Engineering and Physics, Amsterdam UMC, location AMC, Amsterdam, Netherlands
Synopsis
We
compare pseudo-spiral undersampled whole-heart 4D flow MRI with CS reconstruction
with a clinically used EPI readout in a cohort of healthy subjects and patients
with valve regurgitation. Our results indicate that pseudo-spiral CS 4D flow
MRI is at least as reliable as EPI-based 4D flow MRI in terms of inter-valve consistency
of blood flow measurements and agreement with 2D MRI-based regurgitant volume
measurement. Doubling the undersampling factor of the CS acquisition results in
<10% deviation of the measurements compared to the original acquisition, suggesting
that CS scan times may be shortened to expedite clinical implementation.
Introduction
4D flow MRI facilitates detailed evaluation of cardiac
hemodynamics in various types of heart
disease. Whole-heart coverage within clinically feasible scan times (<10
min) requires acceleration. EPI readout is a widely used acceleration
technique, but is susceptible to image distortion artifacts at high
acceleration factors due to phase
accumulation effects, and velocity misregistration in high-velocity regions due
to flow displacement effects. A recently introduced pseudo-spiral Cartesian sampling strategy with
random undersampling in time and compressed sensing reconstruction (1-3),
here referred to as CS, is not sensitive
to these artefacts and may allow for even shorter scan times. In this
study, we compare CS and EPI whole-heart 4D flow MRI acquisitions obtained in
healthy subjects and patients with valve regurgitation, based on blood flow
measurements over the heart valves. Additionally, we investigate the effect of
increasing the CS undersampling factor. We hypothesize that CS-based blood flow
measurements are as consistent between valves as EPI-based measurements, and CS
may be accelerated even further.Methods
12 healthy volunteers (aged 25±3y, 7 female) and 7
patients with valve regurgitation (aged 42±17y, 2 female) underwent cardiac MRI
at 3T including whole-heart CS and EPI 4D flow MRI. 4D flow data were acquired during
free-breathing. Data were binned into 30 cardiac phases using retrospective
ECG-gating. Acquired and reconstructed spatial resolutions were 3.0x3.0x3.0 mm3
and 2.8x2.8x3.0mm3 and three-directional VENC was set to 150 cm/s
(volunteers) and 200-300 cm/s (patients). Scan times ranged from 7 to 10
minutes, depending on FH-dimension of the (transversal oblique) FOV and the subject’s
heart rate during the EPI scan. EPI was obtained with an EPI factor of 5 and a
SENSE factor of 2. The CS undersampling factor ranged from 5.5 to 8.2 to keep
the scan times for both scans the same. CS 4D flow scans were reconstructed
offline with ReconFrame (Gyrotools, Zurich, Switzerland) and using the Berkeley
Advanced Reconstruction Toolbox (BART) (4). A sparsifying total variation
transform in time was used with regularization parameters of r = 0.001 and 20
iteration steps.
Semi-automated retrospective tracking of all four
heart valves was performed using 2D cine bSSFP by dedicated software (CAAS MR
Solutions v5.1 – 4D flow, Pie Medical Imaging) with through-plane valve motion
correction and automatic aliasing and phase offset correction (5,6). Regurgitant
volumes (Rvol) measured by EPI and CS 4D flow were compared to regurgitant
volumes measured by 2D techniques (2D flow for the pulmonary (PV) and aortic
valve (AV), and short-axis planimetry-based stroke volume – 2D flow-based
forward flow for the tricuspid (TV) and mitral valve (MV)).
Valve tracking-based blood flow quantification was
also performed on CS reconstructions of only 50%, 25% and 12.5% of the acquired
data.
To
assess consistency of the blood flow measurements, Bland-Altman parameters and coefficients of variation (SDnet
flow differences/meannet flow*100%) were determined.Results
An example of an EPI
and CS streamline visualization in a healthy subject, obtained by means of semi-automated
retrospective valve tracking, is shown in Figure 1. Between valves, blood flow
measurements obtained with CS demonstrated smaller mean differences, limits of
agreement and coefficients of variation than those obtained with EPI (Table 1).
Figure 2 shows EPI and CS streamline visualizations in four different patients in
which valve regurgitation was quantified. Agreement of the 4D flow MRI-based Rvol
measurements with 2D MRI-based measurements was comparable between EPI and CS
(Figure 3). Reconstructing the CS 4D flow acquisitions with decreasing portions
of the acquired data (Figure 4) resulted in increasing measurement differences
in net flow volumes compared to using 100% of the acquired data, but coefficients
of variation stay within 10% when using 50% of the acquired data.Discussion
In this study, we
compared pseudo-spiral undersampled whole-heart 4D flow MRI with CS reconstruction
with a clinically used EPI readout in healthy subjects and patients. Our
results indicate that CS-based blood flow measurements are at least as
consistent between valves as EPI-based measurements, and agreement of regurgitant
volume measurement with 2D MRI-based measurement is comparable between EPI and
CS. The observation that doubling the undersampling factor of the CS
acquisition resulted in <10% deviation of the measurements compared to the
original acquisition suggests that CS scan times may be shortened.Conclusion
Pseudo-spiral
undersampled whole-heart 4D flow MRI with CS reconstruction is a valuable and reliable
technique for blood flow quantification across the heart valves and may be
accelerated further to expedite clinical implementation.Acknowledgements
No acknowledgement found.References
1. Lustig M, et
al., Sparse MRI: The application of compressed sensing for rapid MR imaging.
Magn. Reson. Med. 2007; 58:1182-1195
2. Gottwald LM,
Peper ES, Zhang Q et al. Pseudo Spiral Compressed Sensing for Aortic 4D Flow MRI: a Comparison with k-t
Principal Component Analysis. Proc 27rd Annu Meet ISMRM, Paris, France. 2018
3. Peper ES, Gottwald LM, Zhang Q et al. 30 times
accelerated 4D flow MRI in the carotids using a Pseudo Spiral Cartesian
acquisition and a Total Variation constrained Compressed Sensing
reconstruction. Proc 27rd Annu Meet ISMRM, Paris, France. 2018
4. Uecker M, et al. Berkeley advanced reconstruction
toolbox. Proc. Intl.
Soc. Mag. Reson. Med. 23; 2015
5. Westenberg JJ, Roes SD, Ajmone Marsan N, et al. Mitral valve and
tricuspid valve blood flow: accurate quantification with 3D velocity-encoded MR
imaging with retrospective valve tracking. Radiology. 2008;249:792–800
6. Blanken CPS, Westenberg JJM, Aben JP, et al. Quantification of Mitral Valve Regurgitation from 4D Flow MRI Using
Semiautomated Flow Tracking. Radiology:
Cardiothoracic Imaging 2020 2:5