Guodong Weng1, Sulaiman Sheriff2, Claus Kiefer1, Irena Zubak3, Andrew A Maudsley2, and Johannes Slotboom1
1Institute for Diagnostic and Interventional Neuroradiology, Support Center for Advanced Neuroimaging (SCAN), University of Bern, Bern, Switzerland, 2Department of Radiology, University of Miami School of Medicine, Miami, FL, United States, 3Inselspital Bern and University Hospital, Bern, Switzerland
Synopsis
To tackle the chemical shift displacement artifacts (CSDA), B1-inhomoheneity, water and fat suppression as well as specific absorption-rate (SAR) limitation at 7T spectroscopy, an EPSI-variant sequence is developed in this study. The conventional slice-selective refocusing pulse is replaced by a spectral-selective adiabatic 2π-pulse pair. The results show a 77% reduction of CSDA, homogeneous refocusing map, water suppression factor ≥ 1000 within an acceptable SAR both in vitro and in vivo measurement. The scan time for the whole brain is within 8 minutes. This new EPSI-variant sequence shows its high application potential clinical routine.
INTRODUCTION
At high magnetic field (≥7T) (i.) B1-inhomogeneity, (ii.) water and lipid signal-suppression
and associated artifacts, (iii.)
limitations on in-vivo specific
absorption-rate (SAR), and (vi.)
chemical-shift displacement artifact (CSDA) are four major factors that impose
restrictions on application, interpretation, and quantification of the
EPSI-data.1 However, the
use of adiabatic (refocusing) pulses allow minimization of the effect of B1+-inhomogeneity 2 and, additionally, the SAR and bandwidth of these type of
RF-pulses can be independently controlled by selection of appropriate RF-pulse
parameters, and these pulses have therefore a clear advance over amplitude-modulated RF-pulse
schemes.3 Finally, CSDA can be minimized by avoiding the use of spatial
selective RF-pulses. In order to tackle all 4 points mentioned at the same
time, we here propose to replace the slice selective (adiabatic) refocusing
pulse (pair) in the EPSI pulse sequence 4 by a spectral-selective adiabatic-complex-secant-hyperbolic 2π-pulse pair 5 in our adapted
EPSI-pulse scheme.METHODS
The changes we made in the EPSI-sequence.4 The
slice-selective 180-degree refocusing pulse (Figure 1a) is replaced by two
identical chemical shift-selective 180-degree secant hyperbolic adiabatic
pulses $$$B_1^+(t) = Ω_0 \cdot sech(βt)^{1+\mu i}$$$ (Figure 1b).
Specifically, the bandwidth (BW) of adiabatic pulses is set to 0.8 kHz (i.e.
2.7 ppm), and the carrier frequency was set at 3.0 ppm. The BW of the
non-adiabatic excitation pulse is 5.5 kHz (18.5 ppm). In vitro tests were carried
out in a spherical brain metabolite phantom (GE), and in vivo tests on one healthy volunteer and
one patient (glioma). All measurements were performed on a 7T MAGNETOM Terra MR-scanner
(Siemens, Germany) in clinical-mode using the 1Tx head coil.RESULTS AND DISCUSSION
The original EPSI-implementation 4 used a
slice-selective Mao refocusing-pulse which is 1.25 kHz (limited by maximum RF-amplitude).
The CSDA of the Mao-refocusing is 23.7%. For the non-selective adiabatic pulses,
the CSDA is only determined by the excitation-pulse and is 5.4%. Therefore, the overall CSDA-error is reduced by approximately 1 - 5.4/23.7 = 77%. Due to the total absence of in-plane
CSDA in case of the proposed chemical shift selective adiabatic refocusing, the amplitude variation of the spectral patterns of metabolites over the total excited 3D-volume
is much smaller.
Figure 2 shows the water signal integration maps using
both EPSI-variants. Compared to using Mao-refocusing (Figure 2c), the water map
of adiabatic pulses (Figure 2d) exciting the water is in good correspondence
with the water-only excitation reference maps (Figure a-b): it indicates that
despite the B1-inhomogeneity of the 1Tx-RF-head coil, signal losses due
RF-inhomogeneity are almost absent by using the chemical selective adiabatic
2π-pulses. The histograms of the water amplitude over the volume are shown
in Figure 2a'-d', and indicate that these 2π-pulses lead to significantly more homogeneous refocusing. The reason for this is that above a certain minimal RF-amplitude these
pulses become B1+-insensitive and their flip-angle become nearly independent of
the B1+-amplitude. Since the proposed 2π-pulses have a relatively small BW, they
operate properly at lower amplitude-threshold and this result in lower SAR than using
spatial selective refocusing without the CSDA-drawback.
Three
adiabatic 2π-pulses with different BWs were applied to investigate their
performance on a spherical phantom (Figure 3). Figure 3c shows nearly perfect
water suppression is achieved with water signal suppression factor of ≥1000
compared to Figure 3a. In addition, due to the symmetry of BW around 3 ppm, equivalent
suppression performance is expected for the lipid region (0.9 to 1.3 ppm).
Figure
4 shows water and creatine (both 3.7 and 3.0 ppm) integration maps as well as
water reference maps using the Mao (Figure 4a'-d') and adiabatic (Figure 4a-d)
pulses respectively. See the figure legend for further details.
In vivo studies of healthy brain tissue in two
subjects (Figure 5b-c) show a high degree of agreement with the in vitro
measurement (Figure 5a). The spectrum within the tumour region (Figure 5d)
shows a clearly different pattern than the normal tissue of the same subject
(Figure 5c). Furthermore, it is noteworthy that only k-space regridding and the
Fourier transformation were performed during pre-and post-processing, and no
water removal and baseline correction was used. Therefore, there are still
various possibilities to improve the final results by post-processing, such as
B0 and eddy current correction (ECC). The measurement time is about 8 minutes
without parallel imaging techniques, which means that even a shorter scan time can be
achieved by using GRAPPA or SENSE.CONCLUSION
Single slab whole-brain EPSI allows the use of
chemical selective small BW adiabatic 2π refocusing pulses, and is only
possible to perform at high magnetic field strength (≥7T). It offers a way to
tackle the B1+ inhomogeneity problem, SAR limitation and the CSDA simultaneously.
Additionally, the proposed pulse sequence has excellent homogeneous water (fat)
suppression. It has also been demonstrated that the proposed adiabatic
2π-pulses guarantee a uniform refocusing over both the selected chemical range
and spatial volume. In vitro and in vivo studies confirmed the expected
performance and capability for potential application clinical routine.Acknowledgements
This work was supported by the Swiss National Science
Foundation grant number 182569. Additionally, this project has received funding
from the European Union’s Horizon 2020 research and innovation programme under
grant agreement No 813120. Initial development of the EPSI sequence was supported by NIH grant R01EB016064.References
- Maudsley AA, Andronesi OC,
Barker PB, Bizzi A, Bogner W, Henning A, et al. Advanced magnetic resonance
spectroscopic neuroimaging: Experts’ consensus recommendations. NMR Biomed
[Internet]. 2020 Apr 29 [cited 2020 Dec 7]; Available from:
https://onlinelibrary.wiley.com/doi/abs/10.1002/nbm.4309
- Slotboom J, Mehlkopf A., Bovée
WMM. A single-shot localization pulse sequence suited for coils with
inhomogeneous RF fields using adiabatic slice-selective RF pulses. J Magn Reson
[Internet]. 1991 Nov [cited 2017 Sep 1];95(2):396–404. Available from:
http://linkinghub.elsevier.com/retrieve/pii/002223649190229M
- Park JY, Garwood M. Spin-echo
MRI using π/2 and π hyperbolic secant pulses. Magn Reson Med [Internet]. 2009
[cited 2020 Dec 7];61(1):175–87. Available from:
/pmc/articles/PMC2688743/?report=abstract
- A. Ebel and A. A. Maudsley, “Improved spectral quality for 3D MR spectroscopic imaging using a high spatial resolution acquisition strategy,” Magn. Reson. Imaging, vol. 21, no. 2, pp. 113–120, 2003.
- Conolly S, Nishimura D,
Macovski A. A selective adiabatic spin-echo pulse. J Magn Reson. 1989 Jun
15;83(2):324–34.