Ruvini Navaratna1,2, Daiki Tamada2, Gregory Simchick2, Diego Hernando1,2, and Scott B Reeder1,2,3,4,5
1Medical Physics, University of Wisconsin - Madison, Madison, WI, United States, 2Radiology, University of Wisconsin - Madison, Madison, WI, United States, 3Biomedical Engineering, University of Wisconsin - Madison, Madison, WI, United States, 4Medicine, University of Wisconsin - Madison, Madison, WI, United States, 5Emergency Medicine, University of Wisconsin - Madison, Madison, WI, United States
Synopsis
MR-based
relaxation parameters such as T2 are sensitive to iron content in the liver. Unfortunately, current
T2 mapping techniques often suffer from long acquisition times. A recently
introduced phase-based T2 mapping technique shows promise for rapid liver T2
quantification within a single breath-hold. However, its ability to quantify
short T2 in liver iron overload is unknown. In this work, we present a modified
3D phase-based T2 mapping method for quantifying short T2 values encountered in
liver iron overload. Our results demonstrate preliminary feasibility of
phase-based RF modulated GRE to map short T2 values rapidly and accurately.
Introduction
MR-based relaxation
parameters such as T2 are increasingly used for liver iron overload assessment
because of their good correlation with liver iron concentration (LIC).1–4 Unfortunately, current T2 mapping methods
suffer from lengthy acquisition times. A novel phase-based T2 mapping method,
in which T2 is encoded into the phase (Figure 1a) of a 3D gradient echo
(GRE) signal, has emerged as a promising technique to rapidly quantify
whole-liver T2 within a breath-hold.5 However, the ability of this technique to
quantify short T2 levels seen in iron overload patients is unknown. Therefore,
the purpose of this work is to modify the phase-based T2 mapping method for
short T2 quantification.Theory
The following three modifications are proposed for phase-based short T2
quantification:
- Short TR acquisition: To facilitate the buildup of coherent transverse magnetization in order
to measure short T2 values, we propose the use of a TR on the order of the
shortest expected iron overload T2, about 3-5ms at 3T (St. Pierre et al1 and $$$R2(3T)\approx1.5\times R2(1.5T)$$$6).
-
"Average
first, fit second”: To
avoid negative phase (and therefore non-physical T2 estimates) due to noise (Figure 1b),
we propose that the signal phase is averaged within an ROI, then mapped to the
corresponding T2 value.
-
T1-corrected reconstruction: Although previous work has demonstrated insensitivity to T1 (required
for the lookup table) at long T1,5 to minimize inaccuracies in our T2
estimate due to unknown, shorter T1, we propose a T1-corrected T2 estimation: $$\begin{equation}T2_{corrected}=T2_{PB}(\widehat{T1})\tag{1}\end{equation}$$ where: $$\begin{equation}\widehat{T1}=\underset{T1}{\operatorname{argmin}}|T2_{PB}(T1)-T2_{fit}(T1)|\tag{2}\end{equation}$$ and $$$T2_{PB}$$$
is the T2 estimated using the phase-based lookup
table, and
$$$T2_{fit}$$$
is an a priori equation relating T2 and T1 (Figure 3b).
Methods
Simulations: Bloch equation simulations of the phase-based
T2 encoding method5 were performed using 1000 isochromats with flip angle = 18°, RF phase increment = $$$\pm$$$2°, TE = 0ms, and T1 = 500ms. The signal
phase as a function of T2 (0-10ms) was plotted for various TRs (3, 6, 9ms) for
noiseless simulations. In addition, the signal phase distributions were plotted
at various T2 (10, 50ms) at TR = 10ms for simulations with added complex noise
to the signal (SNR$$$\approx$$$20) to demonstrate the
effects of noise on phase estimation.
To validate the choice
of flip angle = 18° and RF phase increment = $$$\pm$$$2° under noise and B1
inhomogeneity, 1000 isochromats were simulated with TE = 0ms, TR = 10ms, T2 =
10ms, and T1 = 500ms. T2 standard deviation and absolute T2 bias were
calculated for various flip
angle (5-40°) and phase increment (1-5°) combinations in simulations with noise (SNR$$$\approx$$$20) or with a B1 calibration coefficient7 of $$$\beta=1.2$$$, respectively.
Phantom Acquisition: Eleven agarose gel (2% w/v) vials with
varying MnCl2 concentration (0.3 – 3.7 mM) were constructed to achieve short T2
values, nominally 3 – 30ms (Figure 3a), and short T1 values, nominally
30 – 300ms. Vials were imaged on a 3T MR system (Signa Premier, GE Healthcare,
Waukesha, WI) using a standard head coil (AIR coil, 48 channel, GE Healthcare). Reference T2 and
T1 values were found using a single-echo spin-echo (SE) or inversion-recovered fast spin-echo (FSE-IR)
sequence, respectively. Acquisition parameters are shown in Table
1.
In Vivo Acquisition: One healthy volunteer participated in an
in vivo feasibility study (under IRB approval with informed consent). Liver imaging
was performed on a 3T MR system (Signa Premier, GE Healthcare, Waukesha, WI)
using a posterior and anterior
receive array coil (AIR Coil, GE Healthcare) for the abdomen. The reference T2
value in the volunteer’s liver was found using single-voxel, multi-TE STEAM-MRS8 on a 20 x 20 x 20mm3 volume. Acquisition parameters are shown in Table
1.
Reconstruction/Analysis: The phase-based T2 estimate was found by selecting an ROI, averaging
the signal phase, then estimating T2 using Eqs. 1-2. $$$T2_{fit}$$$ in phantoms was found from
the reference T2 vs. T1 linear regression. For the in vivo study, this is $$$T2_{fit}(ms)=0.07\cdot T1(ms)-44$$$ (from a recent
iron overload prospective study9 - manuscript in preparation).Results
Simulations: Simulations demonstrate that shorter TRs
improve short T2 estimation and averaging signal phase before estimation of T2
is necessary when T2 is short (Figure 1). Optimization
simulations demonstrate low T2 standard deviation due to noise and low bias due
to B1 inhomogeneity using a flip angle of 18° and RF phase increment
of $$$\pm$$$2° (Figure 2).
Phantom
Experiments: $$$T2_{fit}$$$ in phantoms was found to
be $$$T2_{fit}(ms)=0.09\cdot T1(ms)+0.22$$$. Results confirm that T2 estimation is improved using the T1-corrected reconstruction (Figure 3).
In Vivo
Feasibility Study: Images
from the volunteer study demonstrate good agreement between the modified phase-based
T2 (26ms) and reference STEAM-MRS T2 (28ms) (Figure 4).Discussion and Conclusions
In this work, we have
proposed three modifications to a phase-based T2 mapping method: the use of
reduced TRs, “average first, fit second,” and a T1-corrected reconstruction. The feasibility of this modified T2 mapping method was tested
using simulations, phantom experiments, and in vivo studies.
In summary, this study
demonstrates that the proposed modifications to the phase-based T2 mapping
method has potential to quantify T2 in even the most severe iron overloaded
patients within a breath-hold. Studies in patients with known or suspected iron
overload are planned.Acknowledgements
The authors wish
to acknowledge support from the NIH (R01-DK100651, UL1-TR00427, R01-DK117354),
as well as GE Healthcare who provides research support to the University of
Wisconsin. Finally, Dr. Reeder is a Romnes Faculty Fellow, and has received an
award provided by the University of Wisconsin-Madison Office of the Vice
Chancellor for Research and Graduate Education with funding from the Wisconsin
Alumni Research Foundation.References
-
St
Pierre TG, Clark PR, Chua-anusorn W, et al. Noninvasive measurement and imaging
of liver iron concentrations using proton magnetic resonance. Blood.
2005;105(2):855-861. doi:10.1182/blood-2004-01-0177
- Hernando D, Kramer JH, Reeder SB.
Multipeak Fat-Corrected Complex R2* Relaxometry: Theory, Optimization, and
Clinical Validation. Magn Reson Med. 2013;70(5):1319-1331. doi:10.1002/mrm.24593
- Reeder SB, Bice EK, Yu H, Hernando D,
Pineda AR. On the Performance of T2∗
Correction Methods for Quantification of Hepatic Fat Content. Magn Reson Med. 2012;67(2):389-404.
doi:10.1002/mrm.23016
- Wood JC, Enriquez C, Ghugre N, et al.
MRI R2 and R2* mapping accurately estimates hepatic iron concentration in
transfusion-dependent thalassemia and sickle cell disease patients. Blood.
2005;106(4):1460-1465. doi:10.1182/blood-2004-10-3982
- Wang X, Hernando D, Reeder SB.
Phase-based T2 mapping with gradient echo imaging. Magn Reson Med.
2020;84(2):609-619. doi:10.1002/mrm.28138
- Hernando D, Wang C, Mattison R, Yokoo T,
Reeder S. Spectroscopy-Based R2 Relaxometry for Liver Iron Quantification at
1.5T and 3.0T. In Proceedings of the Annual Meeting of ISMRM. No.
1993.; 2017.
- Roberts NT, Hinshaw LA, Colgan TJ, Ii T,
Hernando D, Reeder SB. B0 and B1 inhomogeneities in the liver at 1.5 T and 3.0
T. Magn Reson Med. doi:https://doi.org/10.1002/mrm.28549
- Hamilton G, Middleton MS, Hooker JC, et
al. In vivo breath-hold 1H MRS simultaneous estimation of liver proton density
fat fraction, and T1 and T2 of water and fat, with a multi-TR, multi-TE
sequence. J Magn Reson Imaging. 2015;42(6):1538-1543.
doi:10.1002/jmri.24946
- Hernando D, Zhao R, Taviani V, et al.
Repeatability and reproducibility of confounder-corrected R2* as a biomarker of
liver iron concentration: interim results from a multi-center, multivendor
study at 1.5T and 3T. In Proceedings of the Annual Meeting of ISMRM.
No. 1020; 2019.