Hui Liu1, Gaofeng Shi1, Qinglei Shi2, Weishuai Wang3, Jiangyang Pan1, and Yang Li1
1Fourth Hospital of Hebei Medical University, shijiazhuang, China, 2MR Scientific Marketing, Siemens Healthcare, beijing, China, 3CS, Services,, Siemens Healthcare, jinan, China
Synopsis
The
sequence that continuously acquired Golden-angle RAdial Sparse Parallel
acquisition employing compressed sensing reconstruction (“GRASP”) can acquire
high spatial and high temporal resolution as well as motion robustness to DCE
MRI in liver imaging. However, there are still some artifacts in abdominal
imaging, especially in the early arterial phase. In this study, we proposed an
optimization scheme which can significantly improve the image quality both
in plain and all enhanced phases, which
may have important value in the study of abdominal disease using GRASP based DCE
in future.
Purpose
Through optimization of acquiring parameters, to improve image quality
and diagnostic performance of dynamic contrast-enhanced (DCE) magnetic
resonance imaging (MRI) of abdomen with a high spatiotemporal resolution,
continuously acquired Golden-angle RAdial Sparse Parallel acquisition employing
compressed sensing reconstruction (“GRASP”).Methods
Seventeen patients (10men, 7 women; mean age 30.7±5.67,) underwent DCE magnetic
resonance imaging (MRI) of abdomen with the GRASP sequence at a 3T MR scanner (MAGNETOM
Vida, Siemens Healthcare, Erlangen, Germany). The scanning parameters were
optimized with the principle that the number of radial views in each single
layer is as large as possible. The adjusted parameters are as follows: (1)
extending the total scanning time reasonably; (2) increasing the slice
thickness appropriately; (3) reducing the number of slices as far as possible;
(4) reducing the oversampling as far as possible; (5) shorting TR time as much
as possible. In our experience, in order
to ensure image quality, the number of radial views should be larger than 2500
(Table 1). All the acquired images were divided into two groups (before vs
after optimization). Without awareness of the scanning parameters, two
radiologists (5 and 8 years’ experiences in interpreting gynecological cancers) independently assessed
the image quality of these images about aspects of image noise in left and
right liver lobe, the severity of radial artifact, the degree of image sharpness
and the whole quality with a 5-point scale. In addition, regions of interest
(ROIs) were manually drawn to measure the signal intensity of the liver (SIliverr),
the spleen (SIspleen) and the standard
deviation (SD) of the background signal intensity. The contrast-to noise ratio
(CNR) was defined by the following formula: CNR=(SIliverr-SIspleen)/SD
and the signal noise ratio (SNR) was calculated by the following formula: SNR
=SI/ SD. The CNR, SNR and subjective
image quality scores were compared using independent sample t test and
Mann-Whitney U test. A value of P<0.05 was considered as
statistically significant. Results
After optimization,
the signal-to-noise ratio (SNR) of left lobe and right lobe of the liver are
significantly higher than that before optimization no matter in plain scan,
early arterial phase and late arterial phase (all P<0.05) (Table 2). For
contrast-to-noise ratio (CNR), both the left and right lobes of the liver are
significantly higher in plain scan and late arterial phase (all P<0.05)
(Table 3). In subjective image quality
evaluation, scores about image noise, radial artifact, sharpness and whole
image quality are significantly higher after optimization (all P<0.05)
(Table 4).Discussion
Dynamic
contrast-enhanced MRI (DCE-MRI) has been used to non-invasively assess the
microcirculatory perfusion and vascular permeability of the tumor[1],which
has shown promising results in evaluating treatment response to chemoradiotherapy
in liver cancer[2]. In literatures’
report, the diagnostic performance of DCE-MRI derived parameters may be
influenced by the temporal and spatial resolution of images. So, a high image
quality is very important in improving the performance of DCE imaging[3]. In
the abdomen, the motion artifacts caused by respiratory movement and
physiological peristalsis hinder the application of this technique in the
abdomen. A recently described acquisition technique combining a continuously
acquired radial k-space trajectory with golden-angle sampling and sparse
parallel reconstruction employing compressed sensing offers simultaneous high
spatial and high temporal resolution as well as motion robustness to DCE
MRI[4].However, there are still some artifacts in abdominal imaging, especially
in the early arterial phase. Therefore, the purpose of this study is to improve
the image quality of early and whole contrast-enhancement phases by optimizing
the acquiring parameters. Through study, we have found that our optimizing
scheme significantly improved the image quality both in plain and all contrast-enhanced phases
(Figure 1), which may have important value in the study of abdominal disease
using DCE in future.Conclusion
The optimized scheme
in this experiment significantly improves the image quality of liver dynamic
contrast enhanced (DCE) with GRASP sequence, no matter in plain or contrast-enhanced
phase.Acknowledgements
No acknowledgement found.References
[1] Verma S, Turkbey B, Muradyan N, Rajesh A, Cornud F, Haider MA,
Choyke PL, Harisinghani M. Overview of dynamic contrast-enhanced MRI in
prostate cancer diagnosis and management. AJR Am J Roentgenol. 2012
Jun;198(6):1277-88. doi: 10.2214/AJR.12.8510. PMID: 22623539; PMCID:
PMC6309691.
[2] Chen BB, Shih TT. DCE-MRI in
hepatocellular carcinoma-clinical and therapeutic image biomarker. World J
Gastroenterol. 2014 Mar 28;20(12):3125-34. doi: 10.3748/wjg.v20.i12.3125. PMID:
24695624; PMCID: PMC3964384.
[3] Hao W, Zhao B, Wang G, Wang C, Liu H.
Influence of scan duration on the estimation of pharmacokinetic parameters for
breast lesions: a study based on CAIPIRINHA-Dixon-TWIST-VIBE technique. Eur
Radiol. 2015 Apr;25(4):1162-71. doi: 10.1007/s00330-014-3451-z. Epub 2014 Oct
17. PMID: 25323603.
[4]. Rosenkrantz AB, Geppert C, Grimm R,
et al. Dynamic contrast-enhanced MRI of the prostate with high spatiotemporal
resolution using compressed sensing, parallel imaging, and continuous
golden-angle radial sampling: preliminary experience. J Magn Reson Imaging.
2015; 41(5):1365–1373. [PubMed: 24833417].