Paul Han1, Thibault Marin1, Yanis Djebra1,2, Georges El Fakhri1, Jinsong Ouyang1, and Chao Ma1
1Gordon Center for Medical Imaging, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, Boston, MA, United States, 2LTCI, Télécom Paris, Institut Polytechnique de Paris, Paris, France
Synopsis
Renal perfusion imaging with multi-delay arterial spin labeling (ASL) can
provide multi-parametric information along with more robust quantification of
renal blood flow. However, renal perfusion imaging with ASL is challenging
especially due to respiratory motion. This work presents a subspace-based fast
MR method for free-breathing multi-delay ASL imaging of the kidney. The
feasibility of the proposed method is shown using in vivo data obtained from a
healthy volunteer on a 3T MR scanner.
Introduction
Renal perfusion imaging with multi-delay arterial
spin labeling (ASL) can provide multi-parametric information along with
more robust quantification of renal blood flow1,2. However, renal
perfusion imaging with ASL is challenging especially due to respiratory motion.
Methods utilizing multiple breath-holds3-6 or respiratory triggered
acquisitions7-9 have been proposed to resolve respiratory motion in multi-delay
ASL imaging of the kidney. However, multiple breath-hold scans are difficult to
perform in patient populations and respiratory triggering methods work at the
expense of scan time. In this work, we present a subspace-based fast MR method for
free-breathing multi-delay ASL imaging of the kidney. In vivo experiment was performed to test the
feasibility of the proposed method for free-breathing multi-delay ASL imaging
of the kidney.Methods
The imaging function of ASL signal $$$\rho(x,t)$$$
is modeled as PS functions10-13:
$$\rho(x,t) = \sum_{l=1}^{L} U_{l}(x)V_{l}(t) \;\;\;\;\;\;\;\;\;\;\;\; (1)$$
where $$$U_{l}(\cdot)$$$ and $$$V_{l}(\cdot)$$$
each denotes the spatial and temporal basis functions, respectively, $$$x$$$ denotes
spatial dimension, $$$t$$$ denotes clock time, and $$$L$$$ denotes model order.
The proposed model indicates that the spatio-temporal imaging function resides
in a low-dimensional subspace and thus can be recovered even from an under-sampled
data as long as the temporal basis functions $$$V_{l}(t)$$$ are estimated and sufficient
number of measurements (e.g., greater than model order $$$L$$$) are sampled. $$$V_{l}(t)$$$
can be estimated from training data (D1) acquired with high sampling rate at a
limited number of locations near the $$$k$$$-space center (Fig.1a). All other $$$k$$$-space
locations are sparsely-sampled randomly across the entire $$$(k,t)$$$-space for
the imaging dataset (D2) to ensure sufficient number of measurements are
sampled at each location for image reconstruction (Fig. 1b).
Once the
temporal basis functions $$${\hat V}_{l}(t)$$$ are determined, the spatial
basis functions $$$U_{l}(x)$$$ can be estimated by fitting the model in Eq.1 to
the sampled $$$k$$$-space data:
$$\underset{U_l(x)}{\operatorname{argmin}}||d(k,t)-A\{\rho(x,t)\}||_2^2
+ R(\{\rho(x,f)\}) \;\;\; s.t. \;\; \rho(x,t) = \sum_{l=1}^{L} U_{l}(x){\hat
V}_{l}(t) \;\;\;\;\;\;\;\;\;\;\;\; (2) $$
where $$$d(k,t)$$$ denotes the sparsely-sampled $$$k$$$-space
measurement of ASL imaging data, $$$A$$$ denotes the transformation from image-space
to k-space, and $$$R(\cdot)$$$ denotes the regularization penalty. In this
work, we have defined $$$A=\Omega
F_s$$$, where $$$\Omega$$$ denotes the sampling matrix, and $$$F_s$$$ denotes
the spatial Fourier transform matrix. Results and Discussion
One healthy volunteer was imaged using a
whole-body 3-T MR scanner (TIM Trio, Siemens Healthcare, Erlangen, Germany)
under a study protocol approved by our local Institutional Review Board. Imaging
was performed using a scheme with repeated acquisition of ASL preparation pulse
followed by a continuous series of data acquisition to acquire images at multiple
PLD times and to maximize data sampling efficiency (Fig.2). Pseudo-continuous
ASL (pCASL)14-15 pulse was applied at an axial labeling
plane 10cm superior to the center of the right kidney with conventional pCASL
labeling parameters16. Data was acquired using spoiled gradient echo readout with the following imaging parameters:
image orientation=coronal, field-of-view (FOV)=340$$$\times$$$340mm, matrix
size=128$$$\times$$$128, slice thickness=10mm,
TR/TE=3.4/1.9ms, flip angle=6°, number of phase-encoding (PE) lines sampled per
ASL pulse=750, number of PE lines sampled per frame=10 (frame rate=34ms; 3
lines for D1 and 7 lines for D2), and total acquisition time=3.3min.
In this work, L1-norm
regularization was used for $$$R(\cdot)$$$ and the reconstruction problem in Eq.2 was
solved using an algorithm based on half-quadratic regularization12. The reconstructed clock time images were binned to
different respiratory motion phases by evaluating the 1D profile at the center
of liver. The binned images were further grouped into control/label images at
different PLD times by correlating clock time with PLD times. The
perfusion-weighted images were generated for each PLD by normalizing the
difference images between label and control scans at the same respiratory
motion phase with the control images from the earliest PLD time (i.e., least
affected by RF pulses from readout) as M0.
The proposed approach successfully
reconstructed artifact-free dynamic images from the highly under-sampled $$$(k,t)$$$-space
data (Fig.3a). The 1D profile evaluated at the center of liver (denoted by red
line in Fig.3a) from the reconstructed images show changes in liver position over
clock time reflecting respiratory motion (Fig.3b). Discontinuity in the liver
position was observed periodically over time (yellow arrowhead in Fig.3b) due
to the application of pCASL pulse with relatively long labeling duration
(Fig.3b). The reconstructed images were binned to different respiratory motion
phases and PLD times as shown in Fig.4. Difference in liver position (denoted
by yellow dashed lines in Fig.4a) was observable in images binned at different
respiratory motion bins (Fig.4a). At fixed respiratory motion phase, lower signal
within the descending aorta (yellow arrow) was observed in the label images at
early PLD times compared to the control images. The perfusion-weighted images
were generated over multiple PLD times using the proposed method, as shown in Fig.5.
The results showed expected signal decay from both the kidney cortex and
medulla regions over PLD time.Conclusion
We present a multi-delay ASL method with subspace-based fast MR for
free-breathing perfusion imaging of the kidney. The feasibility of the proposed
method is shown using in vivo data obtained from a healthy volunteer on a 3T MR
scanner. Further investigation is necessary with increased number of subjects
to validate and accurately assess the performance of the proposed method.Acknowledgements
This work was supported in part by the National
Institutes of Health (P41EB022544, R01CA165221, R01HL137230, R01HL118261,
T32EB013180, and K01EB030045).References
1. Nery, Fabio, Isky Gordon, and David L. Thomas.
"Non-invasive renal perfusion imaging using arterial spin labeling MRI:
challenges and opportunities." Diagnostics 8.1 (2018): 2.
2. Nery, Fabio, et al. "Consensus-based technical
recommendations for clinical translation of renal ASL MRI." Magnetic
Resonance Materials in Physics, Biology and Medicine 33.1 (2020): 141-161.
3. Shimizu, Kazuhiro, et al. "Arterial transit
time-corrected renal blood flow measurement with pulsed continuous arterial
spin labeling MR imaging." Magnetic Resonance in Medical
Sciences (2016): mp-2015.
4. Li, Xiufeng, et al. "Quantitative single breath‐hold renal
arterial spin labeling imaging at 7T." Magnetic resonance in
medicine 79.2 (2018): 815-825.
5. Conlin, Christopher C., et al. "Renal plasma
flow (RPF) measured with multiple-inversion-time arterial spin labeling (ASL)
and tracer kinetic analysis: Validation against a dynamic contrast-enhancement
method." Magnetic resonance imaging 37 (2017): 51-55.
6. Kim, Dong Won, et al. "Measurement of arterial
transit time and renal blood flow using pseudocontinuous ASL MRI with multiple
post‐labeling
delays: Feasibility, reproducibility, and variation." Journal of
Magnetic Resonance Imaging 46.3 (2017): 813-819.
7. Cutajar, Marica, et al. "Repeatability of renal
arterial spin labelling MRI in healthy subjects." Magnetic Resonance
Materials in Physics, Biology and Medicine 25.2 (2012): 145-153.
8. Cox, Eleanor F., et al. "Multiparametric renal
magnetic resonance imaging: validation, interventions, and alterations in
chronic kidney disease." Frontiers in physiology 8 (2017): 696.
9. Cutajar, Marica, et al. "Comparison of ASL and
DCE MRI for the non-invasive measurement of renal blood flow: quantification
and reproducibility." European radiology 24.6 (2014): 1300-1308.
10. Liang ZP. "Spatiotemporal imaging with
partially separable functions." Biomedical Imaging: From Nano to Macro,
2007. ISBI 2007. 4th IEEE International Symposium on. IEEE, 2007.
11. Haldar, Justin P., and Zhi-Pei Liang.
"Spatiotemporal imaging with partially separable functions: A matrix
recovery approach." Biomedical Imaging: From Nano to Macro, 2010 IEEE
International Symposium on. IEEE, 2010.
12. Zhao B et al. "Image reconstruction from highly
undersampled-space data with joint partial separability and sparsity
constraints." IEEE transactions on medical imaging 31.9 (2012): 1809-1820.
13. J. He et al. "Accelerated High-Dimensional MR
Imaging With Sparse Sampling Using Low-Rank Tensors," IEEE transactions on
medical imaging, vol. 35, pp. 2119-29, Sep 2016.
14. Dai, Weiying, et al. "Continuous flow-driven
inversion for arterial spin labeling using pulsed radio frequency and gradient
fields." Magnetic Resonance in Medicine: An Official Journal of the
International Society for Magnetic Resonance in Medicine 60.6 (2008):
1488-1497.
15. Wu, Wen-Chau, et al. "A theoretical and
experimental investigation of the tagging efficiency of pseudocontinuous
arterial spin labeling." Magnetic Resonance in Medicine: An Official
Journal of the International Society for Magnetic Resonance in Medicine 58.5
(2007): 1020-1027.
16. Alsop, David C., et al. "Recommended
implementation of arterial spin-labeled perfusion MRI for clinical
applications: a consensus of the ISMRM perfusion study group and the European
consortium for ASL in dementia." Magnetic resonance in medicine 73.1
(2015): 102-116.