Fabian Küppers1,2,3, Seong Dae Yun1, and N. Jon Shah1,2,4,5
1Institute of Medicine and Neuroscience 4, Forschungszentrum Juelich GmbH, Jülich, Germany, 2Institute of Medicine and Neuroscience 11, Forschungszentrum Juelich GmbH, Jülich, Germany, 3RWTH Aachen University, Aachen, Germany, 4Department of Neurology, RWTH Aachen University, Aachen, Germany, 5JARA - BRAIN - Translational Medicine, Aachen, Germany
Synopsis
Combining
GE and SE signals enables
quantification of
several pertinent
parameters making
it useful in
various applications. An
10-echo GESE sequence based on EPIK, known
as EPI with keyhole, has
been presented in our earlier work. This work employs an
accelerated sequence implementation
and aims to perform quantification of T2/T2* for both
phantom
and
in
vivo
data. The method was configured to provide improved resolution within
feasible TEs for brain scans, i.e.
1 minute. The
quantified T2/T2*
values were
in good agreement with values from reference methods. Moreover, an
SNR/tSNR analysis proves
the signal and time gain from multi-shot acceleration.
Introduction
Combining
GE and SE signals has been shown to be effective
for the
simultaneous acquisition of T2 and T2* maps1,
enabling applications such as vessel-size imaging2,
CBV3
and OEF4
analysis.
To this ends, several
sequence implementations based on EPI-readouts
have been published4-6.
However,
the sequences are
limited by the relatively
long echo-time
(TE),
low
SNR at later
TEs
and thus, a
reduced
number of echoes and
lower
resolution. In contrast,
EPIK-readouts7,8
have been shown to
outperform standard
EPI-readouts in terms of SNR and slice coverage, and
in our previous
work9
we introduced a novel
10-echo sequence implementation based on EPIK with an
increased matrix-size of
128. The sequence has now
been further improved in
terms of TE
and parameter quantification performance. The improved method was
verified on two self-made carrageenan phantoms and in
vivo data with
GM/WM separation. The
achieved sequence speed-up was
verified by (t)SNR
analysis. Methods
Two
spherical phantoms containing a solid carrageenan-agarose mixture10
were produced to facilitate diffusion-free, homogeneous phantom
measurements with
reduced
susceptibility artefacts.
The
sequence diagram of 10-echo GESE-EPIK is provided in Fig.1. The
sequence acquires
k-space
with a
128x128
matrix-size, yielding to a spatial
resolution
of 1.9×1.9mm2
with 3mm thickness.
Sixteen
EPIK keyhole lines, GRAPPA factor 2 and a
multi-shot
factor (SPARSE) of 14 are implemented, yielding
TE=10,20,37,47,57,67,77,94,104,114ms.
Phantom data sets were acquired with six
slices and TR=1000ms,
while in
vivo
data acquired
at
3T (male;
29 years) contained
20 slices with TR=2800ms,
leading to TA=59s.
T2*
reference
measurements
were
acquired with a conventional multi-echo (64) gradient-echo sequence11
with
TE starting at 2.9ms in
increments of 1.38ms and TR=1200ms for the
phantom
and 2000ms for in
vivo,
yielding TA=2:38min and 4:36min, respectively. T2 reference acquisitions
were
performed using
four
single spin-echoes
with TE=10,35,60,85ms, TR=650ms (TA=4x2:18min) for the
phantom
and 4500ms (TA=4x7:18min) for the
in
vivo
data. Both methods had
the
same resolution and slices as GESE-EPIK.
All
images were
coregistered to the 10th
multi-GE echo and masked for the brain region.
WM/GM ROIs
were obtained with
SPM12. Reference data were
fitted using
non-linear least-squares fit for
mono-exponential decay to
compute T2 and T2*. The GESE-EPIK data can be analysed using
two approaches:
1)
a
fast-fitting
approach fits T2* on the first two GE, extrapolating
the signal S0=S(t=0) and fitting S0
together with both SE for
T2 and
2)
a matching pursuit algorithm12
computes
the best parameter combination based on all 10 echoes to match the
theoretical signal decay, which
requires higher computational power and time.
SNR
analysis was
performed by dividing the mean signal in the phantom or brain region
by the standard
deviation of
noise, which
was
computed from four
square ROIs
in the image
corners
to avoid influence
from structured noise that may propagate from the subject in phase or
frequency
direction.
Correction
for Rician-noise was
applied. For temporal-SNR analysis,
each
voxel’s
mean value over
the acquired number of repetitions given by the multi-shot factor
was
divided by its standard deviation.Results
Echoes
acquired using GESE-EPIK
at
different slice locations are depicted in Fig.2. T2/T2*
maps from
GESE-EPIK and the
reference methods
for both phantoms are
presented in Fig.3
with histograms.
While
T2*-maps
from GESE-EPIK show broader distributions, the mean values of
T2/T2* are
in good agreement with the
reference methods. Fig.4
shows the results
for in vivo
data. The
consistency between
GESE-EPIK and the
reference methods is
shown
with good agreement
for the mean
value and standard deviation. Considering the two different fitting
approaches for GESE-EPIK data, it was observed that both approaches
show differences smaller
than 1ms for the
mean values of T2/T2*. WM
and GM separation
based on SPM masks were successfully
performed and yielded
T2 values of 69.2+-6.5ms and 63.6+-6.4ms and
T2* of
56.2+-9.3ms and 52.0+-9.0ms for GM and WM, respectively.
As the TE-range
of GESE-EPIK depends on the
keyhole and SPARSE factors,
the influence of different
SPARSE factors on the SNR and tSNR performance was analysed.
Results for phantom and in
vivo data are
presented in Fig.5,
showing the mean SNR/tSNR for each echo, acquired using
two different implementations with a
SPARSE factor of 7
(TEmax=134ms)
and 14 (TEmax=114ms).Discussion and Conclusion
A fast imaging technique for
simultaneous T2 and T2* quantification within
one minute was shown to provide artefact-free images with increased
resolution and number of
echoes while providing relatively
short TEs.
SNR analysis
underlined the suitability of high SPARSE-factors for TE reduction as
no significant loss
in
SNR or
temporal stability was
observed in either the
phantom or
in
vivo
data. The
phantom and in
vivo results showed
good internal consistency for both relaxation parameters obtained
from GESE-EPIK and reference methods while
showing distinct
differences between
WM and GM ROI.
Compared to literature T2*
values13
of 50.0+-5.6ms
and 55.7+-8.72ms, good
agreement was
observed, while T2 values were
found to be below
the
literature values14
(WM:83.9+-10.6/GM:85.6+-11.1ms). Further
in vivo measurements
are
planned to provide generalised
results and application to T2’ and OEF calculation.
In conclusion, 10-echo GESE-EPIK
is a technique for fast multi-parametric imaging of the human brain,
providing T2 and T2* maps with good agreement to reference methods. A
variety of parameters and applications can be investigated
using the method.Acknowledgements
Hereby,
I would like to show my gratitude to my supervisor Prof. N. J. Shah
and coauthor Dr. S. Yun for their expertise and assistance. Next to
this, we thank our colleagues, the MTAs and the organizational team
from the INM-4 institute for their support.References
1 Oshio K, et al. GRASE (Gradient- and Spin-Echo) Imaging: A
Novel Fast MRI Technique. Magnetic Resonance in Medicine 1991, 20,344-349.
2
Kiselev VG, et al. Vessel Size Imaging in Humans. Magnetic Resonance in Medicine 2005, 53:553–563.
3 Donahue K, et al. Simultaneous Gradient-Echo/Spin-Echo EPI
of Graded Ischemia in Human Skeletal Muscle. JMRI 1998, 8:1106+1113.
4 Yin Y, Zhang Y, and Gao JH. Dynamic
Measurement of Oxygen Extraction Fraction Using a Multiecho
Asymmetric Spin Echo (MASE) Pulse Sequence. Magnetic Resonance in
Medicine 2018, 80:1118–1124.
5 Schmiedeskamp H, et al. Multiple gradient- and spin-echo
EPI acquisition technique with z-shimming to compensate for
susceptibility-induced off-resonance effects. Proc. Intl. Soc.
Mag. Reson. Med. 17, 2009.
6 Schmiedeskamp H, et al. Compensation
of Slice Profile Mismatch in Combined Spin- and Gradient-Echo
Echo-Planar Imaging Pulse Sequences. Magnetic Resonance in Medicine 2012, 67:378–388.
7
Zaitsev M, Zilles K, Shah NJ. Shared k-space echo planar
imaging with keyhole. Magnetic Resonance in Medicine 2001, 45(1):109-17.
8
Shah
NJ, da Silva NA, Yun S. Perfusion
weighted imaging using combined gradient/spin echo EPIK: Brain tumour
applications in hybrid MR-PET. Hum Brain Map 2019.
9 Kueppers F, Yun S, Shah NJ. A
Novel 10-echo GESE multi-parametric sequence based on the EPIK
readout for simultaneous quantification of T2 and T2*. Proc. Intl. Soc. Mag. Reson. Med. 28 (2020) – 3158.
10
Stoecker T and Shah NJ. MP-SAGE: A New MP-RAGE Sequence With
Enhanced SNR and CNR for Brain Imaging Utilizing Square-Spiral Phase
Encoding and Variable Flip Angles. Magnetic Resonance in Medicine 2006, 56:824 – 834.
11
Dierkes T, Neeb H, Shah NJ.Distortion correction in echo-planar imaging and
quantitative T2*
mapping. International
Congress Series 2004, 1265:181-185.
12 Mallat SG and
Zhang Z. Matching Pursuits With Time-Frequency
Dictionaries. IEEE Transactions On Signal Processing Vol 41, No 12, 1993.
13
Oros-Peusquens AM, et al. A Single-Scan, Rapid Whole-Brain
Protocol for Quantitative Water Content Mapping With Neurobiological
Implications. Applied Neuroimaging – Frontiers in Neurology 2019.
14
Oros-Peusquens AM, et al. Magnetic field dependence of the
distribution of NMR relaxation times in the living human brain. Magn Reson Mater Phy 2008, 21:131–147.