Michiru Kajiwara1, Tomoyuki Haishi2,3, and Yasuhiko Terada1
1Institute of Applied Physics, University of Tsukuba, Tsukuba, Japan, 2Department of Radiological Sciences, School of Health Sciences at Narita, International University of Health and Welfar, Narita, Japan, 3MRTechnology Inc., Tsukuba, Japan
Synopsis
23Na MRI provides novel information
on energy metabolism, and there are growing expectations for its application in
clinical MRI. However, implementation of 23Na MRI for existing 1H
MRI systems requires expensive dedicated transceivers and RF coils for sodium
imaging. Here, we have developed an add-on crossband repeater system that can
acquire 23Na MRI signals simply by inserting it into the magnet’s
bore of an existing 1H MRI, and proved this concept by phantom and in vivo mouse experiments. This add-on
platform is applicable to other 1H MRI systems and will enhance the
feasibility of 23Na MRI in clinical practice.
INTRODUCTION
Sodium MRI provides novel information on energy metabolism, and has the potential to elucidate physiological
processes [1] in kidneys, brains, heart, knees, skins, and etc. However, clinical
sodium MR imaging remains expensive because of high cost of broadband
multi-nuclear transmit/receive spectrometer and dedicated RF coils for sodium
imaging. These high costs and poor availability are the large impediment for
clinical application of sodium MRI. In this study, we developed a novel add-on 23Na
MRI platform (crossband repeater, CBR) consisting of a crossband convertor and receive/transmit
coils, that could be built into an existing 1H MRI scanner without
the need for modifying the scanner configuration. This study presents an
initial proof-of-concept for the add-on 23Na MRI CBR to a 1.5T 1H
scanner, followed by its application to in
vivo imaging of live mouse. The concept of the proposed system is applicable
to other 1H scanners, and would expand the opportunity for
researchers to assess clinical applications of sodium MRI.METHOD
Concept of
add-on 23Na MRI crossband repeater
We focused on the CBR technology
used in wireless communication, and applied it to transmit signals between the 1H
and 23Na resonance frequencies. The overview and diagram of the
add-on 23Na MRI CBR operated at 1.5T are shown in Figs. 1 and 2. The
64MHz excitation pulses transmitted from the 1.5T 1H-MRI system are
received by a 1H pick-up coil in the gantry, downconverted to 17MHz 23Na
excitation pulse through frequency mixer I, amplified by an RF transmitter, and
transmitted to a 23Na RF coil to excite a sample and acquire NMR
signals. The acquired 17MHz 23Na signal is amplified, upconverted to
64 MHz through frequency mixer II, and transmitted back to the 1H-MRI
by a 1H transmit coil. In this system, the MRI signal received by
the 23Na coil is passed to the 1H receiver coil by
electromagnetic coupling; The 1H-MRI system operates as if it is
receiving the 1H signal, but what is actually acquired is 23Na
signal modulated to 64MHz.
Apparatus
The 1H-MRI
system consisted of a 1.5T/280mm horizontal-bore superconducting magnet
(JASTEC, Japan), gradient coil, transmit/receive RF coil (birdcage coil, 8-elements,
110mm in diameter, 130mm in length), and MRI console. The add-on 23Na
platform consisted of a crossband convertor (DSTechnology, Saitama, Japan) 23Na
transmit/receive coil (17 MHz, birdcage coil, 8-elements, 42mm in diameter, 120mm
in length), 1H pick-up receive/transmit coils (64MHz, loop coils, 50mm
in diameter), RF amplifier (500kHz - 150MHz, 2kW, BTO2000-AlphaSA, TOMCO,
Australia), RF shield between a sample and 1H transmit coil, and
TX/RX switch and 23Na preamplifier (17MHz). All the coils were
homebuilt.
Bench test of
control signals
We measured the fidelity
and time response of the control signals using an oscilloscope (2GHz, ViewGo
II, IWATSU, Japan).
Sensitivity
measurement
To evaluate the
sensitivity of the 23Na signals, the noise and FID signals from
phantoms (saturated saline solution and purified water) were measured in the
following three cases: (A) using a conventional 23Na transceiver (as
a reference) (Fig. 2(b)), (B) using the CBR (Fig. 2(c)), and (C) using the CBR with
direct, bypass electrical connection from the crossband convertor to the 1H-MRI
transceiver using a BNC cable (dot-dashed line in Fig. 2(c)), instead of
connecting via electromagnetic coupling between the 1H transmit coil
of the CBR and the receive coil of the 1H system. The experiment C
was performed to evaluate the signal loss in electromagnetic coupling between
the two 1H RF coils.
In vivo imaging
A
mouse under gas anesthesia (C57BL/6(M), 6 weeks) was imaged using 3D gradient
echo sequences. The sequence parameters for 1H imaging were; matrix
size = 128×64×32, repetition time (TR) / echo time (TE) = 100ms/15ms, and scan
time = 13min 40sec. Those of 23Na imaging were; matrix size =
64×32×32, TR/TE = 40ms/2.8ms, number of excitations = 80, compressed sensing
with acceleration factor of 3, and scan time = 18min 12sec.
RESULT
The bench tests (Fig. 3)
of the RF excitation pulses show that the output signals tracked the input
signals with high fidelity and short rise/fall times.
From the sensitivity
measurement of the MRI signals (Fig. 4), the total signal loss in the CBR
system (B) was 21.1dB (equals to a summation of 9.1dB loss in the crossband convertor
and 12.0dB loss in the electromagnetic coupling), compared with the
conventional 23Na system (A). The signal-to-noise ratios (SNRs) were
37.9dB (A), 34.1dB (B), and 35.2dB (C).
Figure 5 shows the
results of 1H/23Na imaging. The 23Na MRI’s SNR
in the bladder were 21.7dB, comparable to that
of the reference saline solution phantom (18.6dB). DISCUSSSION
The control signals generated
in the CBR had high fidelity, allowing the acquisition of 23Na
signals in synchronization with the existing 1H system. Despite the
large signal loss compared with the conventional 23Na system, the
CBR system provided the 1H/23Na images with the high SNR comparable
to the conventional system. The performance of the system was
verified by live mouse imaging.CONCLUSION
We have developed the prototype of
the add-on 23Na platform using the CBR, that can realize 23Na-MRI
by simply attaching it to an existing 1H-MRI system.Acknowledgements
No acknowledgement found.References
[1] Hu R,
Kleimaier D, Malzacher M, Hoesl MAU, Paschke NK, Schad LR. X‐nuclei imaging:
current state, technical challenges, and future directions. J Magn Reson
Imaging. 2020;51:355‐376.