Zahra Shams1, Evita C. Wiegers1, Wybe J.M. van der Kemp1, Dennis W.J. Klomp1, and Jannie P. Wijnen1
1Department of radiology, University Medical Center Utrecht, Utrecht, Netherlands
Synopsis
The ratio of phosphomonoesters (PME) and phosphodiesters
(PDE) is potentially an important biomarker of cancer in the body, which can be
measured by 31P MR spectroscopy. Here, we introduce a dual band
pulse of low B1+ amplitude for specific refocusing of these
metabolite signals. This can be an alternative solution to the high power
demanding adiabatic refocusing pulses used for 31P MRS in the body,
upon using a multi-echo sequence.
Purpose
The
ratio of phosphomonoesters (PME) and phosphodiesters (PDE), involved in
membrane phospholipid metabolism, is potentially an important biomarker in tumor
growth1. This ratio correlates with aggressiveness of many cancers2,3,
and often changes during successful treatment4. Therefore, it can be
a pivotal metric to monitor therapy response. PDE and PME can be measured by 31P
MR spectroscopy (MRS) with increased intrinsic SNR at 7 Tesla. Higher SNR and
insensitivity to B0 shimming can be achieved by utilizing multi-echo
approaches such as the AMESING5 sequence with adiabatic pulses. This
is of interest particularly in the body where B0 and B1
homogeneity is poor. However, the adiabatic pulses require significant amount
of power to meet the bandwidth requirement. This subsequently results in too high SAR (specific
absorption rate). An alternative solution would be to use relatively uniform B1+
fields to facilitate the use of non-adiabatic RF pulses6. With the limited
B1+ available (15 μT at max in the body), we designed a low B1+
dual-band pulse which can be used in the body to selectively refocus the spins
in the two frequency bands of interest. Phantom and in-vivo experiments in the
brain are shown as a proof of successful operation for the dual band refocusing
pulse at < 15 μT.Methods
We designed
a dual-band Shinnar-LeRoux (SLR) refocusing RF pulse with MATPULSE7
with a duration of 7 ms, B1+ of 14.8 μT and two narrow refocusing
bands with FWHM = 166 Hz. The two refocusing bands were intended to hit the PDE
and PME at 3.0 - 3.5 ppm and 6.2 - 6.8 ppm, respectively. We generated a single
band RF pulse, and simulated a normal refocusing pulse used in spectroscopy from
the Philips pulse archive to compare the required B1+
amplitudes and relative SAR levels to our dual-band pulse. We
calculated the relative SAR of the RF pulses according to $$$\textit{SAR}_{\textit{rel}}\propto \int_0^{T_p} \! B_1^{+2}(t) \, \mathrm{d}t$$$, where Tp is the duration of the RF pulse B1+(t).
MRS
measurements were carried out on a 7 Tesla MR scanner (Philips, Best, NL)
equipped with a double tuned 1H/31P birdcage head coil. We acquired spectra from
a healthy volunteer as well as a phantom enclosing two small spheres. One of
the spheres contained PMEs phosphocholine (PC) and phosphoethanolamine (PE),
and PDE glycerophosphocholine (GPC), in similar concentrations (50 mM); and the
other sphere contained 200 mM inorganic phosphate (Pi). We performed spin echo
(SE) experiments with the dual-band pulse as refocusing pulse; sequence
parameters were: TE/TR 9.98 (ms)/6000
(ms), spectral width 6400 Hz, samples 512, NSA 10, phase cycle 1, no volume
localization. To evaluate the performance of the dual-band pulse, we compared
it to a pulse acquire (PA) sequence with parameters: TE/TR 0.2 (ms) /1000 (ms), flip angle
45ᵒ, spectral width 5000 Hz, samples 512, number of averages (NSA) 4, phase
cycles 2. A 2nd order B0 shimming (either whole phantom or whole brain) was performed prior to
those experiments.Results
The pulse shape and the frequency profile of the designed
dual-band pulse are shown in Figure 1. A B1+ of around 36 (μT)
would be needed to cover
the ppm range including PDEs and PMEs for a complete refocusing with a single
band pulse. Even more power will be needed (~39 μT) if a shorter pulse of 7 (ms) is
desired. Using a broad-band profile to cover the ppm range of ~10 ppm, including
the PME up to the γ-ATP signal would require a B1 amplitude of above 40 μT while still not
covering the entire 31P spectrum. Relative SAR values are much lower
for the dual band pulse (Table 1).
Correct
calibration of the frequency offset of the dual-band pulse is important as shown
in Figure 2A. The third spectrum (Freq. offset 3) shows the fully refocused PDE
and PME signals in the phantom. However, the difference in
signal amplitude between the spin echo and pulse acquire measurement cannot be
judged due to different echo time, NSAs, flip angle and TR. Refocused signals
of PDEs and PMEs from spin echo sequence in vivo in the brain can be seen in Figure
3. Discussion and Conclusion
We introduced a
low B1+ (< 15 μT) dual-band pulse for selective
refocusing of spin groups of PME and PDE; which are two characteristic
metabolites in cancer biology. We showed 31P MRS results in a
phantom and in the brain to verify the pulse shape and performance. Compared to
a single band pulse with hardly sufficient bandwidth for PME and PDE refocusing,
the SAR is reduced by almost a factor of 3. Because of its low power demand, the
low B1+ amplitude dual-band 31P pulse has the
potential to be implemented in multi-echo sequences. This pulse possibly paves
way for efficient multi-echo MRSI of PME and PDE in the body, which will be
investigated in future work.Acknowledgements
We acknowledge funding
form Eurostars Project E!12449 IMAGINE!.References
[1] Glunde K, Bhujwalla ZM, et al., Choline
metabolism in malignant transformation. Nat. Rev. Cancer, 2011; 11: 835–848.
[2] Podo F. Tumour phospholipid metabolism.
NMR Biomed. 1999; 12: 413–439.
[3] van der Kemp WJM, Boer VO, et al., Increase
in SNR for 31P MR spectroscopy by combining polarization transfer with a direct
detection sequence. Magn. Reson. Med. 2012; 68: 353–357.
[4] van der Kemp WJ, Stehouwer BL, et al., Detection
of alterations in membrane metabolism during neoadjuvant chemotherapy in
patients with breast cancer using phosphorus magnetic resonance spectroscopy at
7 Tesla. Springerplus. 2014;3:634.
[5] van der Kemp WJM, Boer VO, et al., Adiabatic
Multi‐Echo P Spectroscopic ImagiNG (AMESING) at 7 tesla for measuring
transverse relaxation times and regaining sensitivity in tissues with short T2*
values. NMR Biomed. 2013; 26:1299–1307.
[6] van Houtum Q, Welting D, et al., Low
SAR 31P (multi-echo) spectroscopic imaging using an integrated whole-body
transmit coil at 7T. NMR in Biomed. 2019;
[7] Matson, G.B. An integrated program for amplitude-modulated
RF pulse generation and re-mapping with shaped gradients. Magn. Reson. Imaging. 1994; 12:1205-1225,