Hongjun An1, Dongmyung Shin1, Hyeong-Geol Shin1, Woojin Jung1, and Jongho Lee1
1Department of Electrical and computer Engineering, Seoul National University, Seoul, Korea, Republic of
Synopsis
In
MRI, scan parameters are carefully selected for desired results. We propose a
new optimization method, AutoSON, that determines optimal scan parameters for
flexibly-selected objectives on given tissue properties such as distribution
and noises. AutoSON optimizes not only scan parameters but also a neural
estimator, which extracts the desired information from MR signals (e.g., quantification
mapping). The method successfully optimized the flip angles in DESPOT1 for T1
mapping and classification of white matter and gray matter. The last example
does not have a simple analytic equation, and therefore, demonstrates a potential
utility of the method.
Introduction
In
MRI, scan parameters (e.g., TR, FA) are selected for a desired objective by
considering properties of interests (e.g., T1, noises). Previously,
a few optimization methods have been utilized to select the scan parameters1.
However, these methods typically require assumptions such as an unbiased network
based on noisy measurement, limiting their applications. Recently, deep-learning
methods have been proposed to design MRI sequences2, 3, which focus
on the development of new sequences, not optimizing sequence parameters of an
existing sequence. In this study, we proposed a new automation method that
jointly optimizes both sequence parameters and a neural network that extracts
information for an objective (AutoSON). This new method allows flexible
optimization to any objective (e.g. T1 mapping, classification of tissues) that
can be modeled as a loss function in the network.Methods
[AutoSON]
AutoSON jointly optimized scan parameters and a neural network (Fig. 1). The MR
simulator generated signals corresponding to the scan parameters from a dataset
that consists of tissue properties. The neural network utilized the
simulated-signals as inputs in order to extract information of a desired
objective (e.g., T1 mapping, tissue classification). The loss function of the
neural network was back-propagated to the network parameters and sequence
parameters, jointly optimizing the network and the sequence. The network structure
was a multi-layer perceptron4. The training dataset was constructed for
the objective (see Experiment 1-3).
[Experiment 1: T1 mapping using DESPOT1]
To
verify the performance of AutoSON, we utilized DESPOT15, which has a
simple analytic signal equation, as a test sequence to optimize scan parameters
for T1 mapping6. In DESPOT1, two data with two FAs (α1, α2)
are utilized. AutoSON optimized these two flip angles to provide high accuracy
in T1 estimation for noise corrupted signals. The input for the network was
magnitude signals from the DESPOT1 equation, and the output was a T1 value.
Training signals are generated using the following parameters: T1=159
ms to 1695ms, T2=50ms, α1, α1=5°-40°,
M0=0.5-1.0,
TR/TE=10/5ms, and noise SD=2e-5. L2 loss was utilized. The optimization results
were compared with the analytic solutions for T1=1645ms.
[Experiment 2: Tissue classification using DESPOT1]
To
demonstrate flexibility in optimization, the FAs for DESPOT1 were re-optimized
to classify white matter and gray matter. The network input was the same but
the output was modified to generate gray matter signal as 0 and white matter
signal as 1. Two datasets, T1=1288±67ms and T2=76±5 ms
for gray matter, and T1=917±32ms and T2=67±4ms for white
matter, were generated with M0=0.5-1.0, noise SD=1e-3 and initial =25°,20°.
The cross-entropy loss was utilized.
[Experiment3: T2 mapping using Partially spoiled SSFP]
To
demonstrate that AutoSON can optimize a sequence with no simple analytic form,
we optimized phase-based T2 mapping7. This method uses two
partially spoiled SSFP images with different RF phase increment. AutoSON optimized 3 parameters: FA, RF phase increment, and TR. The input of the network
was the real and imaginary parts of the ratio of two complex signals. The
output was a T2 value. For the training, the dataset was set to be results of
M0, T1, and T2 of in-vivo data8. Large T2 (>200 ms; CSF) were excluded.
The initial parameters were chosen from Wang et al. (FA=18°, phase increment=2°,
TR=10ms). L2 loss was utilized. For MR signal simulator, a numerical simulation
using Bloch equation was performed for 500 repetitions for steady-state. The
100 spins were induced a 2π phase dispersion in each repetition. Gaussian
noises (SD:2e-4) were added to real and imaginary axes.
Evaluation
was performed for the scan parameters of AutoSON vs. Wang et al. by using the T2
mapping method from Wang et al. for a fair comparison of the sequence parameter
optimization. The T2 mapping results of in-vivo data (another subject data from ref.8) were compared using normalized-root-mean-squared
error (NRMSE).
Results
Figure
2 shows the optimization results of DESPOT1. When started with various initial
points, all results converged to FAs of α1=2.6±0.0° and α2=16.5±0.2°.
These results are close to the optimum FAs from the analytic equation (α1=2.6°,
α2=15.1°).
For
the task of classifying gray and white matter, the results are illustrated in
Figure 3. The optimized FAs (α1=3.2°, α2=18.6°) were
located in-between the optimal angles of white matter (α1=3.5°, α2=20.2°)
and gray matter (α1=3.0°, α2=17.1°), showing reasonable
results. As the optimization progresses, the separation of the two signals
becomes clear (Figure 3b).
The
T2 mapping results are summarized in Figure 4. The optimized scan
parameters (FA=74.2°, phase increment=2.4°, TR=6.7ms) yielded reduced NRMSE
than the original parameters (NRMSE from 8.6±13.5% to 2.6±3.3%). When the
signals are plotted for the two scan parameters (original vs AutoSON), the
newly optimized results show a signal separation depending on T2, demonstrating
successful optimization (Figure 5).Discussion and Conclusion
In this work, a new sequence
optimization method that jointly optimizes scan parameters and a neural network
was proposed. Our results demonstrates feasibility and some of the results (e.g. T2 mapping) may not be feasible (FA too large). Still, the proposed method is flexible in its optimization target. The
optimization is performed as long as the target can be formulated as a loss
function. Additionally, the method does not require any assumptions (e.g. unbiased estimator)
and analytic equation, and therefore flexible. Acknowledgements
This work was supported
by the National Research Foundation of Korea (NRF) grant funded by the Korea
government (MSIT) (No. NRF-2018R1A2B3008445).References
[1] Teixeira, R. P. A., Malik, S. J., & Hajnal, J. V. (2018). Joint system relaxometry (JSR) and Cramer‐Rao lower bound optimization of sequence parameters: a framework for enhanced precision of DESPOT T1 and T2 estimation. Magnetic resonance in medicine, 79(1), 234-245.
[2] Loktyushin, A., Herz, K., Dang, N., Glang, F., Deshmane, A., Doerfler, A., ... & Zaiss, M. (2020). MRzero--Fully automated invention of MRI sequences using supervised learning. arXiv preprint arXiv:2002.04265.
[3] Zhu, B., Liu, J. Z., Koonjoo, N., Rosen, B., & Rosen, M. (2018). AUTOmated pulse SEQuence generation (AUTOSEQ) using Bayesian reinforcement learning in an MRI physics simulation environment. In Proc. 26th Annu. Meeting ISMRM.
[4] Lee, J., Lee, D., Choi, J. Y., Shin, D., Shin, H.-G., & Lee, J. (2020). Artificial neural network for myelin water imaging. Magnetic Resonance in Medicine, 83(5), 1875-1883.
[5] Deoni, S. C., Rutt, B. K., & Peters, T. M. (2003). Rapid combined T1 and T2 mapping using gradient recalled acquisition in the steady state. Magnetic Resonance in Medicine: An Official Journal of the International Society for Magnetic Resonance in Medicine, 49(3), 515-526.
[6] Schabel, M. C., & Morrell, G. R. (2008). Uncertainty in T1 mapping using the variable flip angle method with two flip angles. Physics in Medicine & Biology, 54(1), N1.
[7] Wang, X., Hernando, D., & Reeder, S. B. (2020). Phase‐based T2 mapping with gradient echo imaging. Magnetic Resonance in Medicine, 84(2), 609-619.
[8] Ji, S., Oh. S., Shin, H.-G., Lee, J., (2020). Quad-contrast imaging with quantitative relaxation maps for clinical neuro-evaluation, In Proc. 28th Annu. Meeting ISMRM.