Jonathan Cuthbertson1,2, Trong-Kha Truong1,2, Jasmine Chen1,2, Fraser Robb3, Allen W. Song1,2, and Dean Darnell1,2
1Medical Physics Graduate Program, Duke University, Durham, NC, United States, 2Brain Imaging Analysis Center, Duke University, Durham, NC, United States, 3GE Healthcare, Aurora, OH, United States
Synopsis
The integrated RF/wireless coil
design allows for simultaneous MR image acquisition and wireless data transfer
with the same coil element in order to reduce the number of wired connections
in the scanner. Here, we use this coil design to wirelessly transmit the
scanner trigger signal to perform the Q-spoiling required for MR imaging.
Proof-of-concept experiments in a phantom showed that this coil design was able
to accurately and reliably transmit the scanner trigger from the adjacent
console room to a WiFi-enabled module in the scanner bore, while having minimal
impact on the image SNR or wireless performance.
Introduction
Integrating wireless data
transfer in an MRI scanner would allow for the reduction of wired connections
between RF coil arrays and the scanner electronics, which take up space within
the scanner bore and add to the cost and complexity of the system1,2,3,4. To enable wireless MR data transfer with
an RF coil array, control signals (e.g., the Q-spoiling scanner trigger to RF-isolate
the preamplifier during the scanner transmit cycle) must be transferred from
the scanner to the array according to the pulse sequence timing. Recent bench-top
measurements of multiple antenna designs to transmit the control signals suggest
that a reflected biquad coil loop design could potentially provide the best
wireless connection for Q-spoiling, but would require modifications to the
scanner5,6. In contrast, a novel integrated RF/wireless (iRFW) coil
design was previously developed, in which RF currents at the Larmor frequency
(e.g., 127 MHz for 3T scanners) and in a wireless communication band (e.g., 2.4
GHz for Wi-Fi) flow on the same coil for simultaneous MR image acquisition and
wireless data transfer, respectively, without requiring any scanner
modifications or additional antenna systems within the scanner bore1.
Previously, the iRFW coil design has been used for simultaneous imaging and
wireless control of multiple peripheral systems (e.g., wireless localized B0
shimming2,3, wireless respiratory tracking3, and wireless
physiological motion monitoring with an ultrasound sensor7). In this
work, the iRFW coil design is further developed to wirelessly perform the Q-spoiling
required for MR imaging. Specifically, the iRFW coil design and an onboard
Wi-Fi transceiver module are used to wirelessly receive control signals from
the MRI scanner trigger to detune the receive coil element by activating a PIN
diode during the transmit cycle. Methods
First, a 10-cm diameter RF coil
element was modified into an iRFW coil element by adding custom 127-MHz (Fig.
1, red) and
2.4-GHz (Fig. 1, orange) band-stop filters between the coil element and
the Wi-Fi transceiver module and preamplifier, respectively, to prevent RF
losses incurred to both of them in their frequency bands1. Next, the
coil element was further modified for wireless Q-spoiling by placing a PIN
diode between the 2.4-GHz band-stop filter and the preamplifier (Fig. 1,
blue). This PIN diode is wirelessly activated by applying a DC voltage from the
RF-isolated Wi-Fi module during the scanner transmit cycle. Specifically, wireless
Q-spoiling is performed by sending a wireless command initiated by the scanner
trigger from a computer in the console room to the on-board Wi-Fi module
attached to the coil, which then applies a DC voltage to the PIN diode to
detune the coil element for the duration of the transmit cycle.
Two proof-of-concept experiments
were performed to verify that wireless Q-spoiling with the iRFW coil did not
degrade its image quality or wireless performance. First, a time series of 76 gradient-echo
EPI images (as used in fMRI experiments) was acquired with the coil on a water
phantom using either wired or wireless Q-spoiling, which was achieved by applying
an activation voltage to the PIN diode with either the conventional wired
connections from the scanner or the Wi-Fi module GPIO, respectively. The SNR of
the mean image and the temporal SNR (TSNR) of the image time series were then
calculated for both Q-spoiling methods to evaluate their image quality. During
this experiment, the PIN diode activation and scanner trigger voltages were
measured simultaneously with an oscilloscope to verify that the Wi-Fi module
could provide a similar signal as the scanner for Q-spoiling. Second, the Wi-Fi
radiated power pattern was measured in a fully anechoic chamber with and
without an activation voltage applied to the PIN diode to verify that the
wireless performance would not be degraded during the scanner transmit cycle.Results
The iRFW coil was able to
reliably apply an activation voltage to the PIN diode for wireless Q-spoiling (Fig.
2, blue) that was similar to the voltage applied by the scanner during
conventional wired Q-spoiling (Fig. 2, purple). Additionally, the SNR
and TSNR of the coil averaged over the phantom were similar for wired
Q-spoiling (75.3 and 53.1) and wireless Q-spoiling (75.0 and 54.4) (Fig. 3). Finally, wireless Q-spoiling with
the iRFW coil showed no impact on the Wi-Fi center frequency 3D (Fig. 4A)
or principal plane (Fig. 4B) radiated power gain patterns, which
determine its wireless performance. Further,
the radiated power of the coil across the entire Wi-Fi frequency band changed
by <1% when a voltage was applied to the diode for Q-spoiling relative to
the same unbiased diode measurement.Discussion and Conclusion
This work demonstrates that the
integrated RF/wireless coil design can reliably perform wireless Q-spoiling for
MR image acquisition without degrading image quality relative to the
conventional Q-spoiling method. Additionally, the wireless performance of the
coil is maintained throughout the MRI scan, which is critical to protect the MR
preamplifiers during the scanner transmit cycle. Expanding upon these results,
we envision that the wirelessly transmitted signals that correspond to the scanner
trigger could possibly be used to activate on-board RF switching to enable wireless
power harvesting or to improve the syncing of fMRI stimuli and subject responses
with the timing of the MR image acquisition. Acknowledgements
This work was in part supported by GE Healthcare and grants R01 NS075017
and R01 EB028644 from the National Institutes of Health.References
- Darnell
D et al. Integrated radio-frequency/wireless coil design for simultaneous MR
Image acquisition and wireless communication. Magn. Reson. Med. 2019;81:2176-83
- Cuthbertson
J et al. A 4-Channel iPRES-W Coil Array for Simultaneous MR Image Acquisition
and Wirelessly-Controlled Localized B0 Shimming of the Spinal Cord.
Proceedings of the ISMRM, May 2019, Montreal; pg. no. 1489
- Cuthbertson
J et al. Dual-Stream iPRES-W Head Coil Array for MR Imaging, Wireless
Respiratory Tracking, and Wireless Localized B0 Shimming.
Proceedings of the ISMRM, August 2020, Virtual Conference; pg. no. 1262
- Aggarwal
K et al. A Millimeter-Wave Digital Link for Wireless MRI. IEEE Trans Med Imaging
2017;36(2);574-83
- Lu JY et al. Wireless Q-spoiling of Receive
Coils at 1.5T MRI. Proceedings of the ISMRM, April 2017, Honolulu; pg. no. 4297
- Lu, Jy et al.
Antenna Design for Wireless Clock Syncing and Q-spoiling in MRI.
Proceedings of the ISMRM , June 2018, Paris; pg. no. 28
- Willey D et al. Integrated RF/Wireless Coil and
Ultrasound-Based Sensors to Enable Wireless Physiological Motion Monitoring in
MRI. Proceedings of the ISMRM, August 2020, Virtual Conference; pg. no. 1282