Pavel Seregin1, Georgiy Solomakha1, Egor Kretov2, Oleg Burmistrov1, and Alexey Slobozhanyuk1
1The Department of Physics and Engineering, ITMO University, Saint Petersburg, Russian Federation, 2Max Delbruck Center for Molecular Medicine in the Helmholtz Association, Berlin, Germany
Synopsis
The concept of the wireless power supply for the MRI coils demands high efficiency in a restricted volume. To meet these requirements, we investigated the possibility of harvesting the energy from the B1 field of a 1.5 T MRI scanner using a new volumetric coil design. We performed a series of experiments with the different positions of the proposed coil within the scanners bore and estimated the accepted power levels during the different types of MR pulse sequences.
Main findings
In
this work, we demonstrate a new circularly polarized (CP) volumetric coil
design for energy harvesting in a 1.5 T MRI scanner and evaluated its
performance at various locations along the B0 field axis inside the
bore and during the different MR pulse sequences.Introduction
One
of the most common MRI systems available at clinics is 1.5 T scanners, which
employ a large birdcage-type body coil for the transmit, creating a circularly
polarized (CP) RF magnetic field. The need to use high power during nuclei
excitation gives rise to the idea of taking some of this energy to feed
wireless devices for patients' health monitoring or local coil amplifiers and
detuning circuits1. However, this concept has two significant
limitations inherent in MRI. First, the harvesting circuitry must be compact
since the free space inside the scanner is strictly limited. Second, the
presence of wireless devices should not disturb the imaging process. Numerous
investigators implement small weakly coupled resonant single turn2,3
or multiturn4,5 loops located away from the imaging object to meet
both requirements. This design limits the harvesting system's efficiency since
such flat structures are working with the linearly polarized magnetic fields.
We propose a new design of a compact volumetric harvesting coil that can work
with circular polarization and receive the energy from two CP-polarized RF
magnetic fields. Methods
The
harvesting coil prototype consists of two loops similar to the so-called
"butterfly" design with flat 0-shaped and bent 8-shaped elements. These
elements are assembled on polystyrene foam brick and occupy three of its edges
with dimensions 110x62 mm for the top and 110x37 mm for the sides (Fig. 1a).
Such design has the advantage of geometrically decoupling the coils while keeping
the overall relatively small size. Each coil is connected to the low-impedance
rectifier circuit based on VS-15MQ040-M3 Schottky diodes. The rectifiers'
outputs are combined in series to increase the resulting DC voltage (Fig. 1b).
The output voltage was measured using a 14-bit oscilloscope Owon XDS3202A, and
the harvested power was calculated using this value. In perspective, the
RF harvesting should provide power to the local coil circuitry during each
phase encoding step, so we measured the power output with a single excitation
pulse. For this, we used a manual frequency adjustment procedure. It employs
three separate pulses, but we measured the harvesting output only during the
first one. We perform
several measurements to investigate the potential working region for the
harvesting coil inside the MR scanner. The harvesting coil was located
on top of a body phantom (Fig. 2a). It was moved with the patient table in the
direction of the main static field to
evaluate the harvesting power dependence on the coil positions. We have
performed two sets of experiments on Siemens Avanto and Siemens Espree 1.5 T
systems. Also, the harvested power was measured for various applied imaging
sequences when the coil with the phantom was placed in the magnet's isocenter
(Siemens Espree 1.5 T).Results
The
body coil size varies from one MRI machine to another. The results of the
experiments with the harvesting coil position are demonstrating this difference
in Figure 2b. The potential area for the harvesting volumetric coil placement
is limited by the region -180:180 mm for Siemens Espree and -200:200 mm for
Siemens Avanto. In this region, the coil demonstrates its maximum performance
when located on the top of the body phantom. The type of MR pulse sequence used
also influences the amount of harvested energy. Since many parameters can be
adjusted in each sequence, we have selected the most relevant ones for
consideration in Figure 3. It was observed that the harvesting voltage for FL2D
sequence was small because of the low power examination (low flip angle (FA))
and only one pulse used for excitation. But for some
pulse sequences, such as the T2 TSE, harvesting worked better due to the high-power
examination (high flip angle), many pulses in the excitation phase, and low TR. Discussion and Conclusion
The
proposed CP harvesting coil improves RF harvesting efficiency by receiving CP
RF magnetic field during the transmit. We have demonstrated that the proposed
element's RF harvesting efficiency depends on the position within the body coil
(and its dimensions) and the applied pulse sequence. The average duration of
one B1 excitation pulse was about 100 us, which allows us to receive more 10W
peak power with a 110x67x32 mm volume butterfly coil. However, the estimated
average power at each step of the phase encoding cycle will be about 100-500
mW. It is possible to increase total harvested power by combining additional
harvesting structures or by the increase of its size.Acknowledgements
This work was supported by the Russian Science Foundation (Project 19-75-10104).References
1.
Nohava, L., Ginefri, J.C., Willoquet, G.,
Laistler, E., & Frass-Kriegl, R. (2020). Perspectives in Wireless Radio
Frequency Coil Development for Magnetic Resonance Imaging Frontiers in Physics, 8, 11.
2.
Kelly Byron, Fraser Robb, Shreyas Vasanawala, John Pauly, and Greig Scott.
"Harvesting Power Wirelessly from MRI Scanners”, Proceedings of the 27th
Annual Meeting of ISMRM, 2019.
3.
J. H¨offlin et al., “Energy harvesting with a figure-8 coil - towards energy autonomous
MRI detection,” in Proc. Exp. Nucl. Magn. Resonance Conf., 2016.
4.
Höfflin, Jens, et al. "Energy Harvesting towards autonomous MRI
detection." Proceedings of the 21th Annual Meeting of ISMRM, 2013.
5.
Venkateswaran, Madhav, et al. "Wireless Power Harvesting During MRI."
2020 42nd Annual International Conference of the IEEE Engineering in Medicine
& Biology Society (EMBC). IEEE, 2020.